Learn · Glossary

# Gas analysis glossary

170 plain definitions of the terms used on analyser data sheets, in sample handling and in hazardous-area work, with a link to the lesson that explains each one.

## 4

**4-20 mA**
An analogue signal standard in which the reading is carried as a current, 4 mA at the bottom of the range and 20 mA at the top. See [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html).
Explained in: [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html), [Wiring, grounding and electrical noise for gas analysers](https://www.sarveshanalytics.com/learn/wiring-grounding-and-electrical-noise.html)

## A

**Absolute pressure**
Pressure measured from a perfect vacuum. Gauge pressure plus atmospheric pressure. Gas laws and partial pressure use absolute pressure.
Explained in: [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html)

**Accuracy**
How close a reading is to the true value. See [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html).
Explained in: [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html), [How to read a gas analyser datasheet](https://www.sarveshanalytics.com/learn/reading-a-gas-analyser-datasheet.html), [What accuracy do you really need from a gas analyser?](https://www.sarveshanalytics.com/learn/what-accuracy-do-you-really-need.html)

**Adsorption**
The sticking of gas molecules to a solid surface. Used in dryers and PSA beds, and a cause of slow moisture readings as water adsorbs on tube walls.

**Air ingress**
Air leaking into a process or a sample line. At ppm oxygen levels even a very small leak is significant: [Trace oxygen measurement and air ingress: ppm done right](https://www.sarveshanalytics.com/learn/trace-oxygen-measurement-and-air-ingress.html).

**Alarm set point**
The value at which an alarm activates. Normally paired with a deadband to avoid chatter: [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html).
Explained in: [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

**Ambient temperature**
The temperature of the air around an instrument, which has limits in the datasheet.
Explained in: [Where to mount a gas analyser: panel, shelter or field](https://www.sarveshanalytics.com/learn/where-to-mount-a-gas-analyser.html)

**Analyser (analyzer)**
An instrument that measures the concentration of a gas or other property of a process sample and outputs a signal such as 4-20 mA.
Explained in: [What is online gas analysis? A plain introduction](https://www.sarveshanalytics.com/learn/what-is-online-gas-analysis.html), [Online vs laboratory vs portable gas analysis](https://www.sarveshanalytics.com/learn/online-vs-laboratory-vs-portable-gas-analysis.html)

**Analyser house (shelter)**
A small building or cabin that holds analysers close to the process, protected from the weather and often heated and ventilated: [Where to mount a gas analyser: panel, shelter or field](https://www.sarveshanalytics.com/learn/where-to-mount-a-gas-analyser.html).
Explained in: [Where to mount a gas analyser: panel, shelter or field](https://www.sarveshanalytics.com/learn/where-to-mount-a-gas-analyser.html)

**Asphyxiant**
A gas, such as nitrogen or argon, that harms by displacing oxygen rather than by being toxic: [Oxygen enrichment and oxygen deficiency hazards](https://www.sarveshanalytics.com/learn/oxygen-enrichment-and-oxygen-deficiency-hazards.html).

**ASU (air separation unit)**
A plant that separates air into oxygen, nitrogen and argon, usually by cryogenic distillation.
Explained in: [How an air separation plant works and what is measured](https://www.sarveshanalytics.com/learn/how-an-air-separation-plant-works.html)

**ATEX**
European directives covering equipment for explosive atmospheres. Marking states group, category and temperature class. Other schemes, such as IECEx, are similar in intent.
Explained in: [Hazardous area classification: zones, gas groups, T-Classes](https://www.sarveshanalytics.com/learn/hazardous-area-classification-zones-gas-groups-temperature-classes.html)

**Audit trail**
A record of actions, such as calibrations and setting changes, with date and who did them, used by quality systems: [Records, audit trails and qualification for regulated plants](https://www.sarveshanalytics.com/learn/records-audit-trails-and-qualification-for-regulated-plants.html).
Explained in: [Records, audit trails and qualification for regulated plants](https://www.sarveshanalytics.com/learn/records-audit-trails-and-qualification-for-regulated-plants.html)

**Auto-calibration**
A function that switches zero and span gas to the analyser and adjusts it automatically, started by a timer, keypad or Modbus command. Optional on most ACCURRA-S models.
Explained in: [Auto-Calibration and validation checks for gas analysers](https://www.sarveshanalytics.com/learn/auto-calibration-and-validation-checks.html), [Total cost of ownership of a gas analyser](https://www.sarveshanalytics.com/learn/total-cost-of-ownership-of-a-gas-analyser.html)

**Autoignition temperature**
The temperature at which a flammable mixture ignites without a spark. Equipment surface temperature class must stay below it.

## B

**Back pressure**
Pressure at the vent or outlet of a sample line, which changes the sample pressure at the sensor.

**Balance gas**
The main gas in a calibration mixture, such as nitrogen, that carries the small amounts of target gas: [Calibration gases: choosing, handling and certificates](https://www.sarveshanalytics.com/learn/calibration-gases-choosing-and-handling.html).
Explained in: [Calibration gases: choosing, handling and certificates](https://www.sarveshanalytics.com/learn/calibration-gases-choosing-and-handling.html)

**Barometric pressure**
The pressure of the atmosphere, which varies with weather and altitude; about 1013.25 hPa at sea level in a standard atmosphere.
Explained in: [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html)

**Binary mixture**
A gas mixture of two components. Thermal conductivity analysers measure a binary mixture directly.
Explained in: [How thermal conductivity gas analysers (TCD) work](https://www.sarveshanalytics.com/learn/how-thermal-conductivity-gas-analysers-work.html), [Measuring hydrogen: purity, mixtures and safety limits](https://www.sarveshanalytics.com/learn/measuring-hydrogen-purity-and-hydrogen-in-gas-mixtures.html), [Steel plant gases: coke oven, blast furnace, converter](https://www.sarveshanalytics.com/learn/steel-plant-gases-explained.html)

**Biomethane**
Biogas upgraded to a high methane content, usable like natural gas.

**Blanketing**
Maintaining a slight positive pressure of inert gas above a liquid or powder so air cannot enter: [Inert gas blanketing in pharma and chemical plants](https://www.sarveshanalytics.com/learn/inert-gas-blanketing-in-pharma-and-chemical-plants.html).
Explained in: [Inert gas blanketing in pharma and chemical plants](https://www.sarveshanalytics.com/learn/inert-gas-blanketing-in-pharma-and-chemical-plants.html)

**Blast furnace gas**
A lean by-product fuel gas of iron making, mostly nitrogen, CO and CO2: [Steel plant gases: coke oven, blast furnace, converter](https://www.sarveshanalytics.com/learn/steel-plant-gases-explained.html).

**Blowback**
A burst of clean air or gas sent back through a probe filter to clear dust, triggered on a timer or when sample flow fails.
Explained in: [How a zirconia oxygen analyser works](https://www.sarveshanalytics.com/learn/how-a-zirconia-oxygen-analyser-works.html), [Filters for gas analysers: dust, mist, membrane](https://www.sarveshanalytics.com/learn/filters-and-particulate-protection.html), [Probes and blowback for hot, dusty and tarry gas](https://www.sarveshanalytics.com/learn/hot-dusty-gas-probes-and-blowback.html)

**Bypass loop (fast loop)**
A line that circulates a large flow past the tapping, from which the analyser draws a small branch, to reduce lag: [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html).
Explained in: [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html)

## C

**Calibration**
Adjusting an analyser against gas of known concentration so its reading is correct.
Explained in: [Why ambient pressure affects oxygen purity readings](https://www.sarveshanalytics.com/learn/why-ambient-pressure-affects-oxygen-purity-readings.html), [What is online gas analysis? A plain introduction](https://www.sarveshanalytics.com/learn/what-is-online-gas-analysis.html), [Online vs laboratory vs portable gas analysis](https://www.sarveshanalytics.com/learn/online-vs-laboratory-vs-portable-gas-analysis.html)

**Calibration check**
Applying a known gas and recording the deviation without adjusting the analyser. Compare calibration, which adjusts: [How gas analyser calibration works: zero and span](https://www.sarveshanalytics.com/learn/how-gas-analyser-calibration-works.html).
Explained in: [How gas analyser calibration works: zero and span](https://www.sarveshanalytics.com/learn/how-gas-analyser-calibration-works.html), [How often should you calibrate a gas analyser?](https://www.sarveshanalytics.com/learn/how-often-to-calibrate-a-gas-analyser.html), [Auto-Calibration and validation checks for gas analysers](https://www.sarveshanalytics.com/learn/auto-calibration-and-validation-checks.html)

**Calibration gas**
A certified gas mixture of known concentration used to calibrate or check an analyser: [Calibration gases: choosing, handling and certificates](https://www.sarveshanalytics.com/learn/calibration-gases-choosing-and-handling.html).
Explained in: [Site utilities checklist for gas analysers](https://www.sarveshanalytics.com/learn/site-utilities-checklist-for-analysers.html), [Calibration gases: choosing, handling and certificates](https://www.sarveshanalytics.com/learn/calibration-gases-choosing-and-handling.html), [What accuracy do you really need from a gas analyser?](https://www.sarveshanalytics.com/learn/what-accuracy-do-you-really-need.html)

**Carbon molecular sieve**
An adsorbent used in nitrogen PSA generators that adsorbs oxygen faster than nitrogen.

**Carbon potential**
The carbon content steel would reach at equilibrium with a furnace atmosphere; the control variable in carburising: [Heat treatment furnace atmospheres explained](https://www.sarveshanalytics.com/learn/heat-treatment-furnace-atmospheres-explained.html).

**Catchpot**
A vessel that removes liquid condensate from a sample gas before it reaches downstream components.

**CBG (compressed biogas)**
Biogas upgraded to a high methane content and compressed for use as fuel.
Explained in: [Measuring methane and biogas composition online](https://www.sarveshanalytics.com/learn/measuring-methane-and-biogas.html), [Biogas, landfill gas and digesters explained](https://www.sarveshanalytics.com/learn/biogas-landfill-gas-and-digesters-explained.html)

**CEMS**
Continuous emission monitoring system: equipment that measures stack gases and reports them at regulatory conditions: [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html).
Explained in: [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html)

**Certificate of analysis**
The document that comes with a calibration gas cylinder, stating the concentration, its uncertainty, the date and the traceability.

**Chilled mirror hygrometer**
An instrument that cools a mirror until dew forms and detects it optically. A reference method for dew point: [Moisture measurement methods compared for dry gas](https://www.sarveshanalytics.com/learn/moisture-measurement-methods-compared.html).
Explained in: [Moisture measurement methods compared for dry gas](https://www.sarveshanalytics.com/learn/moisture-measurement-methods-compared.html)

**Clinker**
The hot, partly fused product of a cement kiln that is ground to make cement: [Cement kiln gas analysis explained: O2, CO and safety](https://www.sarveshanalytics.com/learn/cement-kiln-gas-analysis-explained.html).

**Coalescing filter**
A filter that merges fine liquid droplets into larger ones that drain away.
Explained in: [Filters for gas analysers: dust, mist, membrane](https://www.sarveshanalytics.com/learn/filters-and-particulate-protection.html)

**Coke oven gas**
A by-product gas of coke making, rich in hydrogen and methane: [Steel plant gases: coke oven, blast furnace, converter](https://www.sarveshanalytics.com/learn/steel-plant-gases-explained.html).

**Combustibles**
Unburnt gases such as CO and hydrogen in flue gas. They can burn at a hot zirconia cell and make the oxygen reading too low.

**Condensate**
Liquid that forms when a gas is cooled below its dew point.
Explained in: [How to design a gas sample handling system](https://www.sarveshanalytics.com/learn/how-to-design-a-gas-sample-handling-system.html), [Measuring carbon monoxide and carbon dioxide online](https://www.sarveshanalytics.com/learn/measuring-carbon-monoxide-and-carbon-dioxide.html), [Measuring methane and biogas composition online](https://www.sarveshanalytics.com/learn/measuring-methane-and-biogas.html)

**Controlled atmosphere (CA)**
Storage with oxygen and CO2 held at set levels to slow ripening: [Controlled-Atmosphere fruit storage explained](https://www.sarveshanalytics.com/learn/controlled-atmosphere-fruit-storage-explained.html).
Explained in: [Controlled-Atmosphere fruit storage explained](https://www.sarveshanalytics.com/learn/controlled-atmosphere-fruit-storage-explained.html)

**Converter gas**
A CO-rich by-product gas of oxygen steelmaking, recovered when its composition allows: [Steel plant gases: coke oven, blast furnace, converter](https://www.sarveshanalytics.com/learn/steel-plant-gases-explained.html).

**Cross-sensitivity**
A sensor's response to gases other than its target: [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html).
Explained in: [What is an NDIR gas analyser?](https://www.sarveshanalytics.com/learn/what-is-an-ndir-gas-analyser.html), [How thermal conductivity gas analysers (TCD) work](https://www.sarveshanalytics.com/learn/how-thermal-conductivity-gas-analysers-work.html), [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html)

**Cryogenic**
Relating to very low temperatures, such as liquid oxygen, nitrogen and argon, which boil at about minus 183, minus 196 and minus 186 C.
Explained in: [Sampling cryogenic and liquefied gases: vaporisers](https://www.sarveshanalytics.com/learn/sampling-cryogenic-and-liquefied-gases.html), [How an air separation plant works and what is measured](https://www.sarveshanalytics.com/learn/how-an-air-separation-plant-works.html)

## D

**DCS**
Distributed control system: a plant control system made of networked controllers and operator stations.
Explained in: [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

**Dead leg**
A branch of tubing with no flow. It holds old gas and adds lag and contamination.
Explained in: [Response time, T90 and sample lag in gas analysis](https://www.sarveshanalytics.com/learn/response-time-t90-and-sample-lag.html), [Trace oxygen measurement and air ingress: ppm done right](https://www.sarveshanalytics.com/learn/trace-oxygen-measurement-and-air-ingress.html), [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html)

**Dead volume**
Volume in a sample system that is not swept by the flow, such as dead legs and oversized fittings.

**Detection limit**
The smallest concentration an analyser can distinguish from zero.
Explained in: [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html)

**Dew point**
The temperature at which water vapour in a gas begins to condense.
Explained in: [What is dew point and how is it measured?](https://www.sarveshanalytics.com/learn/what-is-dew-point-and-how-is-it-measured.html), [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html), [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html)

**Digester**
A sealed vessel in which organic matter is broken down without air to produce biogas: [Biogas, landfill gas and digesters explained](https://www.sarveshanalytics.com/learn/biogas-landfill-gas-and-digesters-explained.html).

**Drain**
A route for condensate to leave a sample system, which must be continuous and safe: [Condensate, moisture and sample cooling for gas analysis](https://www.sarveshanalytics.com/learn/condensate-moisture-and-sample-cooling.html).
Explained in: [Site utilities checklist for gas analysers](https://www.sarveshanalytics.com/learn/site-utilities-checklist-for-analysers.html)

**Drift**
A slow change in an analyser's reading over time with the same input gas.
Explained in: [Online vs laboratory vs portable gas analysis](https://www.sarveshanalytics.com/learn/online-vs-laboratory-vs-portable-gas-analysis.html), [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html), [How often should you calibrate a gas analyser?](https://www.sarveshanalytics.com/learn/how-often-to-calibrate-a-gas-analyser.html)

**Dry basis**
A gas value stated as if all water vapour were removed. Contrast wet basis: [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html).
Explained in: [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html), [Combustion, excess air and flue gas oxygen explained](https://www.sarveshanalytics.com/learn/combustion-excess-air-and-flue-gas-oxygen.html), [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html)

**Dual-channel analyser**
An analyser that measures two gases from one sample in one enclosure, such as ACCURRA-S oxygen with CO2 or oxygen with dew point.
Explained in: [Measuring methane and biogas composition online](https://www.sarveshanalytics.com/learn/measuring-methane-and-biogas.html), [Glove boxes and Controlled-Atmosphere chambers explained](https://www.sarveshanalytics.com/learn/glove-boxes-and-controlled-atmosphere-chambers.html)

## E

**Electrochemical cell**
A sensor that generates a current from a chemical reaction with the target gas, used for trace oxygen and CO. The cell is consumed over time.
Explained in: [Electrochemical vs paramagnetic oxygen sensors](https://www.sarveshanalytics.com/learn/electrochemical-vs-paramagnetic-oxygen-sensors.html), [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html), [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html)

**Electrolyser**
A device that splits water into hydrogen and oxygen using electricity: [Hydrogen electrolysers and gas analysis: what to measure](https://www.sarveshanalytics.com/learn/hydrogen-electrolysers-and-gas-analysis.html).

**Enclosure rating (IP)**
The ingress protection rating that states how well an enclosure keeps out dust and water, for example IP65.
Explained in: [Where to mount a gas analyser: panel, shelter or field](https://www.sarveshanalytics.com/learn/where-to-mount-a-gas-analyser.html)

**Endothermic gas**
A furnace atmosphere made by partly burning natural gas over a catalyst, mostly CO, hydrogen and nitrogen: [Heat treatment furnace atmospheres explained](https://www.sarveshanalytics.com/learn/heat-treatment-furnace-atmospheres-explained.html).
Explained in: [Heat treatment furnace atmospheres explained](https://www.sarveshanalytics.com/learn/heat-treatment-furnace-atmospheres-explained.html)

**ESP (electrostatic precipitator)**
A dust collector that charges particles with high voltage. It can ignite gas if CO and unburnt fuel are present: [Cement kiln gas analysis explained: O2, CO and safety](https://www.sarveshanalytics.com/learn/cement-kiln-gas-analysis-explained.html).

**Excess air**
Air supplied to a burner beyond the amount needed for exact combustion, estimated from flue gas oxygen: [Excess air from flue gas oxygen calculator](https://www.sarveshanalytics.com/calculators/excess-air-calculator.html).
Explained in: [Combustion, excess air and flue gas oxygen explained](https://www.sarveshanalytics.com/learn/combustion-excess-air-and-flue-gas-oxygen.html)

**Explosive atmosphere**
A mixture of air with flammable gas, vapour or dust, within its explosive limits, that can burn or explode if ignited.

**Extractive analysis**
Measuring a gas by drawing a sample from the process to an analyser, usually through a conditioning system.
Explained in: [Extractive vs In-Situ gas analysis](https://www.sarveshanalytics.com/learn/extractive-vs-in-situ-gas-analysis.html), [What is online gas analysis? A plain introduction](https://www.sarveshanalytics.com/learn/what-is-online-gas-analysis.html)

## F

**Failsafe relay**
A relay that moves to its alarm state if power or the instrument fails.
Explained in: [What is limiting oxygen concentration (LOC)?](https://www.sarveshanalytics.com/learn/what-is-limiting-oxygen-concentration.html), [How to read a gas analyser datasheet](https://www.sarveshanalytics.com/learn/reading-a-gas-analyser-datasheet.html), [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html)

**Fine filter**
A final-stage filter before the analyser; in ACCURRA-S a glass micro fibre filter is built in.
Explained in: [How to design a gas sample handling system](https://www.sarveshanalytics.com/learn/how-to-design-a-gas-sample-handling-system.html), [Filters for gas analysers: dust, mist, membrane](https://www.sarveshanalytics.com/learn/filters-and-particulate-protection.html), [Maintenance schedule and spares for gas analysers](https://www.sarveshanalytics.com/learn/maintenance-schedule-and-spares-for-gas-analysers.html)

**Flame arrestor**
A device that stops a flame travelling along a pipe by absorbing its heat in narrow passages.

**Flameproof enclosure (Ex d)**
An enclosure that withstands an internal explosion and prevents it igniting the surrounding atmosphere.
Explained in: [Hazardous area gas analysers explained](https://www.sarveshanalytics.com/learn/hazardous-area-gas-analysers-explained.html), [Protection concepts: flameproof, intrinsic safety, purging](https://www.sarveshanalytics.com/learn/protection-concepts-flameproof-intrinsic-safety-and-purging.html), [Inert gas blanketing in pharma and chemical plants](https://www.sarveshanalytics.com/learn/inert-gas-blanketing-in-pharma-and-chemical-plants.html)

**Flue gas**
The gas leaving a furnace, boiler or kiln after combustion.

**Fractionation**
The separation of a liquid mixture by boiling, the lighter component leaving first. It is why liquid samples must be vaporised completely: [Sampling cryogenic and liquefied gases: vaporisers](https://www.sarveshanalytics.com/learn/sampling-cryogenic-and-liquefied-gases.html).
Explained in: [Sampling cryogenic and liquefied gases: vaporisers](https://www.sarveshanalytics.com/learn/sampling-cryogenic-and-liquefied-gases.html)

**Frost point**
The temperature at which moisture deposits as ice. Dew points below 0 C in dry gas are usually frost points.
Explained in: [Moisture measurement methods compared for dry gas](https://www.sarveshanalytics.com/learn/moisture-measurement-methods-compared.html)

## G

**Galvanic isolation**
Electrical separation between circuits so no direct current path exists, used to break ground loops and protect electronics.
Explained in: [Wiring, grounding and electrical noise for gas analysers](https://www.sarveshanalytics.com/learn/wiring-grounding-and-electrical-noise.html)

**Gas chromatograph**
An instrument that separates the components of a mixture in a column and measures each. Used for multi-component analysis; not in the Sarvesh range.
Explained in: [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html)

**Gas group IIC**
The most demanding hazardous-area gas group, covering hydrogen and acetylene.
Explained in: [Hazardous area gas analysers explained](https://www.sarveshanalytics.com/learn/hazardous-area-gas-analysers-explained.html), [Measuring hydrogen: purity, mixtures and safety limits](https://www.sarveshanalytics.com/learn/measuring-hydrogen-purity-and-hydrogen-in-gas-mixtures.html), [Flammability, explosive limits and gas safety basics](https://www.sarveshanalytics.com/learn/flammability-explosive-limits-and-gas-safety-basics.html)

**Gauge pressure**
Pressure measured relative to the atmosphere. Add atmospheric pressure to get absolute pressure.

**Ground loop**
A circuit formed by two ground connections at different potentials, which drives current through a signal path and adds noise: [Wiring, grounding and electrical noise for gas analysers](https://www.sarveshanalytics.com/learn/wiring-grounding-and-electrical-noise.html).
Explained in: [Wiring, grounding and electrical noise for gas analysers](https://www.sarveshanalytics.com/learn/wiring-grounding-and-electrical-noise.html)

## H

**H2S**
Hydrogen sulphide: a toxic, corrosive, flammable gas found in biogas and some process streams. It is outside the standard ACCURRA-S sample limits.

**Hazardous area**
A place where an explosive atmosphere may occur. Classified into zones: [Hazardous area classification: zones, gas groups, T-Classes](https://www.sarveshanalytics.com/learn/hazardous-area-classification-zones-gas-groups-temperature-classes.html).

**Heat tracing**
Electrical or steam heating of a line to keep it above the dew point of the gas inside.
Explained in: [Sample lines: materials, length, diameter and lag](https://www.sarveshanalytics.com/learn/sample-lines-materials-length-and-lag.html), [Condensate, moisture and sample cooling for gas analysis](https://www.sarveshanalytics.com/learn/condensate-moisture-and-sample-cooling.html), [Probes and blowback for hot, dusty and tarry gas](https://www.sarveshanalytics.com/learn/hot-dusty-gas-probes-and-blowback.html)

**Hot-wet extraction**
Keeping the sample hot and wet from the probe to a heated analyser so that soluble gases are not lost: [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html).

**HRSG**
Heat recovery steam generator, which makes steam from the hot exhaust of a gas turbine.

**Hygroscopic**
Readily absorbing moisture from the surroundings.
Explained in: [Moisture measurement methods compared for dry gas](https://www.sarveshanalytics.com/learn/moisture-measurement-methods-compared.html)

**Hysteresis**
A deadband that makes an alarm switch on at one value and off at a different value, to prevent chatter.
Explained in: [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

## I

**IECEx**
An international scheme for certifying equipment for explosive atmospheres.
Explained in: [Hazardous area classification: zones, gas groups, T-Classes](https://www.sarveshanalytics.com/learn/hazardous-area-classification-zones-gas-groups-temperature-classes.html)

**In-situ analysis**
Measuring the gas where it flows, with the sensor in the duct or pipe and no sample line.
Explained in: [How a zirconia oxygen analyser works](https://www.sarveshanalytics.com/learn/how-a-zirconia-oxygen-analyser-works.html), [Extractive vs In-Situ gas analysis](https://www.sarveshanalytics.com/learn/extractive-vs-in-situ-gas-analysis.html), [What is online gas analysis? A plain introduction](https://www.sarveshanalytics.com/learn/what-is-online-gas-analysis.html)

**Inerting**
Replacing air in a vessel with an inert gas such as nitrogen to keep oxygen below the level that supports combustion.
Explained in: [What is limiting oxygen concentration (LOC)?](https://www.sarveshanalytics.com/learn/what-is-limiting-oxygen-concentration.html), [Flammability, explosive limits and gas safety basics](https://www.sarveshanalytics.com/learn/flammability-explosive-limits-and-gas-safety-basics.html), [Inerting and purging: how they work and how to prove them](https://www.sarveshanalytics.com/learn/inerting-and-purging-how-they-work.html)

**Interference**
An error in a reading caused by another gas or condition in the sample.
Explained in: [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html)

**Interlock**
A control or safety logic that stops or prevents an action when a condition, such as high oxygen, is met.

**Intrinsic safety (Ex i)**
A protection concept that limits electrical energy so that no spark or heat can ignite a flammable atmosphere: [Protection concepts: flameproof, intrinsic safety, purging](https://www.sarveshanalytics.com/learn/protection-concepts-flameproof-intrinsic-safety-and-purging.html).
Explained in: [Protection concepts: flameproof, intrinsic safety, purging](https://www.sarveshanalytics.com/learn/protection-concepts-flameproof-intrinsic-safety-and-purging.html)

## L

**Lambda**
The ratio of actual air to the air needed for exact combustion. Lambda 1 is stoichiometric: [Excess air from flue gas oxygen calculator](https://www.sarveshanalytics.com/calculators/excess-air-calculator.html).
Explained in: [Combustion, excess air and flue gas oxygen explained](https://www.sarveshanalytics.com/learn/combustion-excess-air-and-flue-gas-oxygen.html)

**Landfill gas**
A methane-rich gas produced by waste in landfills, drawn from wells and used or flared.

**Leak test**
A test, with pressure and a detection method, that shows whether a sample system or enclosure leaks.

**LEL**
Lower explosive limit: the lowest concentration of a flammable gas in air that can burn or explode. About 4 % for hydrogen and 5 % for methane: [Flammability, explosive limits and gas safety basics](https://www.sarveshanalytics.com/learn/flammability-explosive-limits-and-gas-safety-basics.html).
Explained in: [Measuring hydrogen: purity, mixtures and safety limits](https://www.sarveshanalytics.com/learn/measuring-hydrogen-purity-and-hydrogen-in-gas-mixtures.html), [Flammability, explosive limits and gas safety basics](https://www.sarveshanalytics.com/learn/flammability-explosive-limits-and-gas-safety-basics.html), [Hydrogen electrolysers and gas analysis: what to measure](https://www.sarveshanalytics.com/learn/hydrogen-electrolysers-and-gas-analysis.html)

**Limiting oxygen concentration (LOC)**
The oxygen level below which a fuel and inert gas mixture cannot sustain a flame.
Explained in: [What is limiting oxygen concentration (LOC)?](https://www.sarveshanalytics.com/learn/what-is-limiting-oxygen-concentration.html), [Inerting and purging: how they work and how to prove them](https://www.sarveshanalytics.com/learn/inerting-and-purging-how-they-work.html), [Inert gas blanketing in pharma and chemical plants](https://www.sarveshanalytics.com/learn/inert-gas-blanketing-in-pharma-and-chemical-plants.html)

**Linearity**
How closely the response of an analyser follows a straight line over its range.
Explained in: [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html)

**Live zero**
The use of 4 mA, not 0 mA, for the bottom of a 4-20 mA range, so a broken wire can be recognised.
Explained in: [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html)

## M

**Membrane filter**
A thin porous membrane, often PTFE, that removes fine particles and aerosols and repels water: [Filters for gas analysers: dust, mist, membrane](https://www.sarveshanalytics.com/learn/filters-and-particulate-protection.html).
Explained in: [Filters for gas analysers: dust, mist, membrane](https://www.sarveshanalytics.com/learn/filters-and-particulate-protection.html)

**Modbus RTU**
A serial communication protocol, here over RS 485, used to read measurements, status and history from an analyser.
Explained in: [How to read a gas analyser datasheet](https://www.sarveshanalytics.com/learn/reading-a-gas-analyser-datasheet.html), [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html), [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

**Molar volume**
The volume of one mole of gas: 22.414 litres at 0 C and 1013.25 hPa for an ideal gas.
Explained in: [ppm to mg/Nm3: volume vs mass concentration](https://www.sarveshanalytics.com/learn/volume-mass-and-mg-per-nm3-concentration-units.html)

**Mole fraction**
The share of molecules in a mixture that are a given gas. Percent and ppm by volume are mole fractions.
Explained in: [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html)

**Molecular sieve**
A porous solid that adsorbs molecules by size, used to dry gas and remove CO2 and in PSA beds.

## N

**NDIR**
Non-dispersive infrared: a method that measures gases such as CO, CO2 and methane by their absorption of infrared light.
Explained in: [What is an NDIR gas analyser?](https://www.sarveshanalytics.com/learn/what-is-an-ndir-gas-analyser.html), [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html), [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html)

**Needle valve**
A fine adjustment valve used to set sample flow.

**Normal cubic metre (Nm3)**
A cubic metre of gas at normal conditions, usually 0 C and 1013.25 hPa: [ppm to mg/Nm3: volume vs mass concentration](https://www.sarveshanalytics.com/learn/volume-mass-and-mg-per-nm3-concentration-units.html).
Explained in: [ppm to mg/Nm3: volume vs mass concentration](https://www.sarveshanalytics.com/learn/volume-mass-and-mg-per-nm3-concentration-units.html)

## O

**Oxygen deficiency**
An atmosphere with less oxygen than normal, commonly below about 19.5 %, which endangers life: [Oxygen enrichment and oxygen deficiency hazards](https://www.sarveshanalytics.com/learn/oxygen-enrichment-and-oxygen-deficiency-hazards.html).
Explained in: [Oxygen enrichment and oxygen deficiency hazards](https://www.sarveshanalytics.com/learn/oxygen-enrichment-and-oxygen-deficiency-hazards.html), [Controlled-Atmosphere fruit storage explained](https://www.sarveshanalytics.com/learn/controlled-atmosphere-fruit-storage-explained.html)

**Oxygen enrichment**
An atmosphere with more oxygen than normal, commonly above about 23.5 %, in which materials ignite easily.
Explained in: [Oxygen enrichment and oxygen deficiency hazards](https://www.sarveshanalytics.com/learn/oxygen-enrichment-and-oxygen-deficiency-hazards.html)

## P

**Paramagnetic sensor**
An oxygen sensor that uses oxygen's attraction to a magnetic field. It has no consumed element.
Explained in: [Electrochemical vs paramagnetic oxygen sensors](https://www.sarveshanalytics.com/learn/electrochemical-vs-paramagnetic-oxygen-sensors.html), [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html)

**Partial pressure**
The pressure a gas in a mixture would exert alone: its mole fraction times the total pressure: [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html).
Explained in: [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html)

**Peltier cooler**
A solid-state cooler with no refrigerant or compressor, used to cool and dry sample gas.
Explained in: [How to design a gas sample handling system](https://www.sarveshanalytics.com/learn/how-to-design-a-gas-sample-handling-system.html), [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html), [Condensate, moisture and sample cooling for gas analysis](https://www.sarveshanalytics.com/learn/condensate-moisture-and-sample-cooling.html)

**Permeation**
The slow passage of gas through the wall of a material such as plastic tubing, a cause of oxygen and moisture error at trace levels.
Explained in: [Sample lines: materials, length, diameter and lag](https://www.sarveshanalytics.com/learn/sample-lines-materials-length-and-lag.html)

**PLC**
Programmable logic controller: a small industrial computer that runs control logic.
Explained in: [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

**Poisoning (sensor)**
Loss of sensor performance caused by a contaminant, such as sulphur, chlorine or silicone: [Sensor life, poisoning and replacement in gas analysers](https://www.sarveshanalytics.com/learn/sensor-life-poisoning-and-replacement.html).
Explained in: [Sensor life, poisoning and replacement in gas analysers](https://www.sarveshanalytics.com/learn/sensor-life-poisoning-and-replacement.html)

**ppb**
Parts per billion by volume: one part in a billion. 1000 ppb is 1 ppm: [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html).
Explained in: [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html)

**ppm**
Parts per million by volume. 10,000 ppm is 1 %.
Explained in: [What is dew point and how is it measured?](https://www.sarveshanalytics.com/learn/what-is-dew-point-and-how-is-it-measured.html), [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html), [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html)

**Precision**
How closely repeated readings of the same gas agree with each other.
Explained in: [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html)

**Preheater**
A tower of cyclones that heats raw meal with kiln gas before the kiln in a cement plant.

**Pressure dew point**
The dew point of a gas at its line pressure, usually much higher than the dew point after expansion to atmosphere: [Compressed air quality and dew point (ISO 8573-1)](https://www.sarveshanalytics.com/learn/compressed-air-quality-and-dew-point.html).
Explained in: [Compressed air quality and dew point (ISO 8573-1)](https://www.sarveshanalytics.com/learn/compressed-air-quality-and-dew-point.html)

**Pressure regulator**
A device that reduces and holds a steady downstream pressure.
Explained in: [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html)

**Probe**
The first part of a sampling system, inserted in the process to draw a sample through a filter.
Explained in: [Extractive vs In-Situ gas analysis](https://www.sarveshanalytics.com/learn/extractive-vs-in-situ-gas-analysis.html), [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html), [Probes and blowback for hot, dusty and tarry gas](https://www.sarveshanalytics.com/learn/hot-dusty-gas-probes-and-blowback.html)

**Proof test**
A periodic test that reveals hidden failures in a safety function, such as applying test gas to an analyser.
Explained in: [What SIL means for gas analysers, and what it does not](https://www.sarveshanalytics.com/learn/what-sil-means-for-gas-analysers.html)

**PSA / VPSA**
Pressure swing adsorption and vacuum pressure swing adsorption, processes that make nitrogen or oxygen on site using adsorbent beds.
Explained in: [PSA, VPSA and membrane gas generators explained](https://www.sarveshanalytics.com/learn/nitrogen-and-oxygen-generators-psa-vpsa-and-membrane.html)

**Purge**
To replace the gas in a vessel or line with another gas, usually inert, before use or opening: [Inerting and purging: how they work and how to prove them](https://www.sarveshanalytics.com/learn/inerting-and-purging-how-they-work.html).
Explained in: [Trace oxygen measurement and air ingress: ppm done right](https://www.sarveshanalytics.com/learn/trace-oxygen-measurement-and-air-ingress.html), [Commissioning an online gas analyser: A Step-by-Step guide](https://www.sarveshanalytics.com/learn/commissioning-an-online-gas-analyser.html), [Inerting and purging: how they work and how to prove them](https://www.sarveshanalytics.com/learn/inerting-and-purging-how-they-work.html)

**Purging (Ex p)**
A protection concept in which an enclosure is kept at a slight positive pressure of clean air or inert gas to keep hazardous gas out.
Explained in: [Protection concepts: flameproof, intrinsic safety, purging](https://www.sarveshanalytics.com/learn/protection-concepts-flameproof-intrinsic-safety-and-purging.html)

**Purity analyser**
An analyser that measures a high concentration of a gas, near 100 %, such as oxygen purity from 95 to 100 %.
Explained in: [Why ambient pressure affects oxygen purity readings](https://www.sarveshanalytics.com/learn/why-ambient-pressure-affects-oxygen-purity-readings.html), [Choosing an oxygen analyser: trace, percent or purity](https://www.sarveshanalytics.com/learn/choosing-an-oxygen-analyser-trace-percent-or-purity.html), [How an air separation plant works and what is measured](https://www.sarveshanalytics.com/learn/how-an-air-separation-plant-works.html)

## R

**Range**
The span of concentration an analyser measures, for example 0 to 1000 ppm.
Explained in: [Why ambient pressure affects oxygen purity readings](https://www.sarveshanalytics.com/learn/why-ambient-pressure-affects-oxygen-purity-readings.html), [Electrochemical vs paramagnetic oxygen sensors](https://www.sarveshanalytics.com/learn/electrochemical-vs-paramagnetic-oxygen-sensors.html), [Choosing an oxygen analyser: trace, percent or purity](https://www.sarveshanalytics.com/learn/choosing-an-oxygen-analyser-trace-percent-or-purity.html)

**Reference conditions**
The temperature and pressure, dry or wet basis and reference oxygen to which a concentration is stated.
Explained in: [ppm to mg/Nm3: volume vs mass concentration](https://www.sarveshanalytics.com/learn/volume-mass-and-mg-per-nm3-concentration-units.html), [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html)

**Reference gas**
A second gas supplied to some analysers for comparison. ACCURRA-TCD needs none.
Explained in: [How a zirconia oxygen analyser works](https://www.sarveshanalytics.com/learn/how-a-zirconia-oxygen-analyser-works.html), [How thermal conductivity gas analysers (TCD) work](https://www.sarveshanalytics.com/learn/how-thermal-conductivity-gas-analysers-work.html)

**Repeatability**
The agreement between repeated readings of the same gas under the same conditions.
Explained in: [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html)

**Resolution**
The smallest change an instrument can display or output.
Explained in: [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html)

**Respiratory quotient**
The ratio of CO2 produced to oxygen consumed by respiring fruit or cells. Used in some storage and fermentation controls.
Explained in: [Controlled-Atmosphere fruit storage explained](https://www.sarveshanalytics.com/learn/controlled-atmosphere-fruit-storage-explained.html)

**Response time (T90)**
The time for a reading to reach 90 % of a step change in concentration.
Explained in: [Response time, T90 and sample lag in gas analysis](https://www.sarveshanalytics.com/learn/response-time-t90-and-sample-lag.html), [How to read a gas analyser datasheet](https://www.sarveshanalytics.com/learn/reading-a-gas-analyser-datasheet.html)

**Rotameter**
A flow meter with a float in a tapered tube, used to see and set sample flow.
Explained in: [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html), [Controlled-Atmosphere fruit storage explained](https://www.sarveshanalytics.com/learn/controlled-atmosphere-fruit-storage-explained.html)

**RS-485**
A two-wire serial communication standard used by Modbus RTU, allowing many devices on one cable over long distances.
Explained in: [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html)

## S

**Safe area**
A place classified as free of explosive atmospheres, where ordinary equipment may be used.

**Safety instrumented function**
A function that detects a dangerous condition and brings the process to a safe state, with a defined safety integrity level: [What SIL means for gas analysers, and what it does not](https://www.sarveshanalytics.com/learn/what-sil-means-for-gas-analysers.html).
Explained in: [What SIL means for gas analysers, and what it does not](https://www.sarveshanalytics.com/learn/what-sil-means-for-gas-analysers.html)

**Sample conditioning**
Treating a sample, by filtering, cooling, drying and regulating, so that it suits the analyser.

**Sample handling system**
The probe, lines, cooler, pump and filters that deliver a conditioned sample to an analyser.
Explained in: [Extractive vs In-Situ gas analysis](https://www.sarveshanalytics.com/learn/extractive-vs-in-situ-gas-analysis.html), [How to design a gas sample handling system](https://www.sarveshanalytics.com/learn/how-to-design-a-gas-sample-handling-system.html), [What is online gas analysis? A plain introduction](https://www.sarveshanalytics.com/learn/what-is-online-gas-analysis.html)

**Sample lag**
The time for a sample to travel from the tapping to the analyser, equal to line volume divided by flow: [Gas sample line lag time calculator](https://www.sarveshanalytics.com/calculators/sample-lag-time-calculator.html).
Explained in: [Response time, T90 and sample lag in gas analysis](https://www.sarveshanalytics.com/learn/response-time-t90-and-sample-lag.html), [Sample lines: materials, length, diameter and lag](https://www.sarveshanalytics.com/learn/sample-lines-materials-length-and-lag.html)

**Sample point (tapping)**
The connection on the process from which gas is taken for analysis: [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html).
Explained in: [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html)

**SCADA**
Supervisory control and data acquisition: software that collects plant data, shows it to operators and stores it.
Explained in: [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

**Scrubber**
A filter or chemical bed that removes an interfering or harmful gas from a sample before the sensor.
Explained in: [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html), [What samples an analyser cannot take: know the limits](https://www.sarveshanalytics.com/learn/what-samples-an-analyser-cannot-take.html)

**Selectivity**
The ability of a sensor to respond only to its target gas.

**Sensor bypass valve**
A valve that isolates trace oxygen or dew point sensors during purging or upsets so they recover quickly.
Explained in: [What is dew point and how is it measured?](https://www.sarveshanalytics.com/learn/what-is-dew-point-and-how-is-it-measured.html), [Response time, T90 and sample lag in gas analysis](https://www.sarveshanalytics.com/learn/response-time-t90-and-sample-lag.html), [Moisture measurement methods compared for dry gas](https://www.sarveshanalytics.com/learn/moisture-measurement-methods-compared.html)

**Sensor life**
The time over which a sensor performs within specification, which depends on the type and the sample: [Sensor life, poisoning and replacement in gas analysers](https://www.sarveshanalytics.com/learn/sensor-life-poisoning-and-replacement.html).
Explained in: [Maintenance schedule and spares for gas analysers](https://www.sarveshanalytics.com/learn/maintenance-schedule-and-spares-for-gas-analysers.html), [Sensor life, poisoning and replacement in gas analysers](https://www.sarveshanalytics.com/learn/sensor-life-poisoning-and-replacement.html), [Total cost of ownership of a gas analyser](https://www.sarveshanalytics.com/learn/total-cost-of-ownership-of-a-gas-analyser.html)

**Shielded cable**
A cable with a conductive screen that protects the signal from electrical noise.
Explained in: [Wiring, grounding and electrical noise for gas analysers](https://www.sarveshanalytics.com/learn/wiring-grounding-and-electrical-noise.html)

**SIL**
Safety integrity level, a measure of the reliability of a safety function under functional safety standards. Sarvesh makes no SIL claim for its analysers.
Explained in: [Hazardous area gas analysers explained](https://www.sarveshanalytics.com/learn/hazardous-area-gas-analysers-explained.html), [What SIL means for gas analysers, and what it does not](https://www.sarveshanalytics.com/learn/what-sil-means-for-gas-analysers.html), [Hydrogen electrolysers and gas analysis: what to measure](https://www.sarveshanalytics.com/learn/hydrogen-electrolysers-and-gas-analysis.html)

**Siloxane**
A silicon-containing organic compound found in landfill gas and some biogas that can damage sensors and engines.

**Span**
The difference between the top and bottom of an analyser's range. A span error scales readings; compare zero error.
Explained in: [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html), [How gas analyser calibration works: zero and span](https://www.sarveshanalytics.com/learn/how-gas-analyser-calibration-works.html)

**Span drift**
A change over time in the scale of an analyser's response.

**Span gas**
Gas of known concentration near the top of the range, used to set the analyser's gain during calibration.
Explained in: [What is an NDIR gas analyser?](https://www.sarveshanalytics.com/learn/what-is-an-ndir-gas-analyser.html), [Online vs laboratory vs portable gas analysis](https://www.sarveshanalytics.com/learn/online-vs-laboratory-vs-portable-gas-analysis.html), [Sample system troubleshooting: symptoms to causes](https://www.sarveshanalytics.com/learn/sample-system-troubleshooting.html)

**Stoichiometric**
Exactly the proportions of fuel and air needed for complete combustion.
Explained in: [Combustion, excess air and flue gas oxygen explained](https://www.sarveshanalytics.com/learn/combustion-excess-air-and-flue-gas-oxygen.html)

**Stratification**
Layering of a gas stream so that composition or temperature differs across a duct: [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html).
Explained in: [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html)

**Syngas**
Synthesis gas, a mixture mainly of carbon monoxide and hydrogen made by gasification.
Explained in: [Steel plant gases: coke oven, blast furnace, converter](https://www.sarveshanalytics.com/learn/steel-plant-gases-explained.html)

## T

**TCD (thermal conductivity detector)**
A sensor that measures a gas by how well it carries heat away from a heated element.
Explained in: [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html), [How thermal conductivity gas analysers (TCD) work](https://www.sarveshanalytics.com/learn/how-thermal-conductivity-gas-analysers-work.html), [Measuring hydrogen: purity, mixtures and safety limits](https://www.sarveshanalytics.com/learn/measuring-hydrogen-purity-and-hydrogen-in-gas-mixtures.html)

**TDLAS**
Tunable diode laser absorption spectroscopy: measuring a gas by laser light absorbed on one narrow line. Not in the Sarvesh range.
Explained in: [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html)

**Temperature class**
A marking, T1 to T6, giving the maximum surface temperature of hazardous-area equipment, from 450 C to 85 C.
Explained in: [Hazardous area classification: zones, gas groups, T-Classes](https://www.sarveshanalytics.com/learn/hazardous-area-classification-zones-gas-groups-temperature-classes.html)

**Trace measurement**
Measurement at low concentration, typically ppm.
Explained in: [What is dew point and how is it measured?](https://www.sarveshanalytics.com/learn/what-is-dew-point-and-how-is-it-measured.html), [Choosing an oxygen analyser: trace, percent or purity](https://www.sarveshanalytics.com/learn/choosing-an-oxygen-analyser-trace-percent-or-purity.html), [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html)

**Traceability**
The unbroken chain of comparisons linking a measurement to a national or international standard.
Explained in: [Calibration gases: choosing, handling and certificates](https://www.sarveshanalytics.com/learn/calibration-gases-choosing-and-handling.html), [How often should you calibrate a gas analyser?](https://www.sarveshanalytics.com/learn/how-often-to-calibrate-a-gas-analyser.html), [Records, audit trails and qualification for regulated plants](https://www.sarveshanalytics.com/learn/records-audit-trails-and-qualification-for-regulated-plants.html)

## U

**UEL**
Upper explosive limit: the highest concentration of a flammable gas in air that can still burn. About 75 % for hydrogen.
Explained in: [Flammability, explosive limits and gas safety basics](https://www.sarveshanalytics.com/learn/flammability-explosive-limits-and-gas-safety-basics.html)

**ULO (ultra-low oxygen)**
A controlled-atmosphere regime that holds oxygen at around 1 to 2 % or lower.

**Uncertainty**
A stated range within which the true value is expected to lie, given on calibration certificates.

## V

**Validation**
Checking, with a known gas, that an analyser reads correctly, without necessarily adjusting it: [Auto-Calibration and validation checks for gas analysers](https://www.sarveshanalytics.com/learn/auto-calibration-and-validation-checks.html).
Explained in: [Records, audit trails and qualification for regulated plants](https://www.sarveshanalytics.com/learn/records-audit-trails-and-qualification-for-regulated-plants.html)

**Vaporiser**
A heated device that turns a liquid or cold sample into gas of the same composition.
Explained in: [Sampling cryogenic and liquefied gases: vaporisers](https://www.sarveshanalytics.com/learn/sampling-cryogenic-and-liquefied-gases.html), [How an air separation plant works and what is measured](https://www.sarveshanalytics.com/learn/how-an-air-separation-plant-works.html)

**Vent**
The outlet where a sample leaves an analyser, which must be safe and have steady low back pressure.
Explained in: [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html), [Site utilities checklist for gas analysers](https://www.sarveshanalytics.com/learn/site-utilities-checklist-for-analysers.html), [Inerting and purging: how they work and how to prove them](https://www.sarveshanalytics.com/learn/inerting-and-purging-how-they-work.html)

**Volume fraction**
The share of a mixture's volume taken by a component. For gases, equal to mole fraction.
Explained in: [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html)

## W

**Warm-up time**
The time after power-up for a sensor or heater to stabilise before readings are reliable.

**Wet basis**
A gas value stated with water vapour counted as part of the gas: [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html).
Explained in: [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html)

## Z

**Zeolite**
A porous mineral used as an adsorbent, for example in oxygen PSA generators, where it adsorbs nitrogen.

**Zero**
The analyser reading with no target gas present. A zero error shifts every reading equally.
Explained in: [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html), [How gas analyser calibration works: zero and span](https://www.sarveshanalytics.com/learn/how-gas-analyser-calibration-works.html)

**Zero drift**
A change over time in the analyser's reading at zero gas.

**Zero gas**
Gas free of the target component, used to set the analyser's zero during calibration.
Explained in: [What is an NDIR gas analyser?](https://www.sarveshanalytics.com/learn/what-is-an-ndir-gas-analyser.html), [Sample system troubleshooting: symptoms to causes](https://www.sarveshanalytics.com/learn/sample-system-troubleshooting.html), [Commissioning an online gas analyser: A Step-by-Step guide](https://www.sarveshanalytics.com/learn/commissioning-an-online-gas-analyser.html)

**Zirconia sensor**
A heated ceramic sensor that measures oxygen from the voltage generated between two sides of a zirconia element.
Explained in: [How a zirconia oxygen analyser works](https://www.sarveshanalytics.com/learn/how-a-zirconia-oxygen-analyser-works.html), [Electrochemical vs paramagnetic oxygen sensors](https://www.sarveshanalytics.com/learn/electrochemical-vs-paramagnetic-oxygen-sensors.html), [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html)

**Zone 1 / Zone 2**
Hazardous-area zones where flammable gas may occur occasionally (Zone 1) or is not likely in normal operation (Zone 2).
Explained in: [Hazardous area gas analysers explained](https://www.sarveshanalytics.com/learn/hazardous-area-gas-analysers-explained.html), [Where to mount a gas analyser: panel, shelter or field](https://www.sarveshanalytics.com/learn/where-to-mount-a-gas-analyser.html), [Hazardous area classification: zones, gas groups, T-Classes](https://www.sarveshanalytics.com/learn/hazardous-area-classification-zones-gas-groups-temperature-classes.html)