title: How to Design a Gas Sample Handling System
url: https://www.sarveshanalytics.com/learn/how-to-design-a-gas-sample-handling-system.html
description: A practical guide to gas sample handling: probe, heated line, cooler, pump and filters, common failure modes, and the sample limits of ACCURRA-S and i-Cons.
updated: 2026-11-03
source: html
---

Learn · Sampling and conditioning

# How to design a gas sample handling system for an online analyser

Most analyser trouble starts upstream of the analyser. The sample system is where accuracy is won or lost.

 Intermediate 4 min read Lesson 1 of 10[Sampling and conditioning](https://www.sarveshanalytics.com/learn/track-sampling.html)

Helpful first: [Extractive vs In-Situ gas analysis](https://www.sarveshanalytics.com/learn/extractive-vs-in-situ-gas-analysis.html), [Response time, T90 and sample lag in gas analysis](https://www.sarveshanalytics.com/learn/response-time-t90-and-sample-lag.html)

 Quick answer

A good sample handling system delivers a representative, clean, dry sample at constant pressure and flow. Keep the sample above its dew point until the cooler, keep lines short, filter in stages, and supervise faults. ACCURRA-S expects a dry, non-corrosive, non-flammable sample at 2 to 10 psig and about 1 LPM.

Updated 3 November 2026 · Figures from Sarvesh datasheets

**After this lesson you can**

- List the stages of a sample handling system in order
- Apply the rules that keep a sample representative
- State what an ACCURRA-S expects at its inlet

## The chain

A typical extractive chain has a probe with a filter at the tip, a transport line, a cooler or vaporiser to condition the sample, a pump, fine filters and scrubbers, and then the analyser. Each stage has a job and a failure mode.

Probe, heated line, cooler, pump, fine filter and analyser, with a controller supervising blowback, line temperature and pump interlock.

## Typical failure modes

- A blocked probe filter reduces flow and the analyser reads stale gas
- Condensate in the line absorbs soluble gases and floods the analyser
- A leak lets air in; at ppm level it dominates the reading
- A failed pump gives no flow, with no obvious sign
- Dead volumes and long lines slow the response

## Supervision: let the system say when it is wrong

A sample system fails quietly. The analyser keeps reading, but what it reads is no longer the process. Supervision turns each failure into an alarm:

| Failure | What detects it | What should happen |
|---|---|---|
| Cooler warm | Cooler temperature | Pump stops; SGI-300 runs the external pump only when the sample is below 12 C and dry |
| Water past the cooler | Condensate sensor | Pump stops, alarm, so the analyser is not flooded |
| Probe filter blocking | Flow falls | Blowback, then alarm if flow does not recover |
| Heated line cold | Pt100 on the hose | Alarm; readings flagged |
| Calibration running | Digital inputs from the analyser | Outputs frozen, control system told |

i-Cons reads these inputs and reports values, status and maintenance history over Modbus, so the control room sees the sample system's health, not just a number.

## Design rules

- Keep the sample above its dew point, with a heated probe and line, until the cooler
- Condition in one place: cool, drain and filter before the analyser
- Keep lines short and small in volume, in stainless or PTFE as the gas requires
- Filter in stages: probe filter, then coalescing or fine filter
- Provide blowback for dusty gas, and make it automatic on flow failure
- Measure flow and condensate, and trip the pump on a fault
- Calibrate through the same path where possible

## A worked design: boiler flue gas to an extractive analyser

Flue gas at 180 C with about 12 % water vapour and light dust. Oxygen and CO are to be measured dry. Walk the chain:

1. **Dew point.** 12 % water at atmospheric pressure is a vapour pressure of about 12.2 kPa, a dew point of about 50 C. Everything up to the cooler must stay well above 50 C.
2. **Probe.** SP-300 has a heated 10 micron SS316 sintered filter, for process gas from 120 to 500 C, dust up to 10 g/Nm3 and moisture up to 15 %. Blowback at 4 to 7 kg/cm2 clears the filter.
3. **Heated line.** i-Cons drives heated hose in lengths of 30 to 90 m at 40 W per metre. A 30 m hose draws 1.2 kW; set its temperature well above the 50 C dew point, with Pt100 feedback.
4. **Cooler.** The i-Cons Peltier cooler takes gas at up to 160 C with a dew point up to 65 C and holds the outlet at 5 C. Our gas fits both limits.
5. **What leaves the cooler.** At 5 C the gas keeps about 0.86 % water, so a dry 3.00 % oxygen reads about 2.97 %. Good enough for most control; correct for it in reporting if you need to.
6. **Pump and flow.** The i-Cons pump gives 5 LPM free flow with a fast loop valve. The analyser takes about 1 LPM; the rest keeps the line fresh.
7. **Lag.** 30 m of 4 mm bore PTFE holds about 380 mL. At 4 LPM that is about 6 seconds; at 1 LPM it would be about 23.

**Good to know**

The SGI-300 and i-Cons datasheets limit the sample to SO2, NOx and CO below 1000 ppm and exclude HCl, HF, H2S and other highly corrosive, toxic or flammable gases. A sample outside that needs a different design.

## What the analyser expects

The ACCURRA-S catalogue states the sample it expects: non-corrosive, non-toxic, non-flammable, non-condensing and dry, free from entrained oil, at 2 to 10 psig regulated externally, flow about 1 LPM, dust below 3 micron and a dew point at least 5 degrees below the lowest ambient temperature. Where your sample does not meet these, a conditioning system is designed around your application.

## When to use i-Cons or SGI-300

i-Cons packages the pump, cooler, filter, condensate and flow sensors, blowback and heated-hose control with diagnostics. Its cooler takes sample at up to 160 C with a dew point up to 65 C, and an SP-300 heated probe handles process gas from 120 to 500 C with dust up to 10 g per normal cubic metre. SGI-300 is the cooler with integrated drain pump, filter and condensate sensor for building your own panel. i-Cons and SGI-320 are not suitable for samples containing HCl, HF or H2S.

## Guidance and standards

**API RP 555**, Process Analyzers, is the long-standing industry guide to analyser systems, including sample conditioning and analyser houses. **ISO 10715** gives sampling guidance for natural gas. For stack emission measurement, **EN 15259** covers the measurement site and plan, and **EN 14181** the quality assurance of automated measuring systems once installed: [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html).

Lag and line size: [Sample lines: materials, length, diameter and lag](https://www.sarveshanalytics.com/learn/sample-lines-materials-length-and-lag.html). The tapping: [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html).

**In short**

- The goal is a representative, clean, dry sample at constant pressure and flow.
- Keep the sample above its dew point until the cooler.
- Filter in stages and keep lines short.
- Supervise faults with flow and condensate alarms.

## Check your understanding

Answer in your head or on paper first, then open the answer. Recalling an answer teaches more than rereading.

**Why must the sample stay above its dew point until the cooler?**

Condensation anywhere earlier in the line removes water-soluble gases and wets the line, which changes the sample before it is analysed.

**What sample pressure and flow does an ACCURRA-S expect?**

2 to 10 psig regulated externally, and a flow of about 1 LPM.

**A reading lags a process change by minutes. Which part of the sample system do you suspect?**

Line volume and length: tube internal volume divided by flow is the transport delay. Shorten the line or raise the flow.

## Key terms

- [Sample handling system](https://www.sarveshanalytics.com/learn/glossary.html#t-sample-handling-system)
- [Fine filter](https://www.sarveshanalytics.com/learn/glossary.html#t-fine-filter)
- [Condensate](https://www.sarveshanalytics.com/learn/glossary.html#t-condensate)
- [Peltier cooler](https://www.sarveshanalytics.com/learn/glossary.html#t-peltier-cooler)

## Frequently asked questions

**What is the most common sample system fault?**

A blocked probe filter or condensate in the line. Both are addressed by blowback, heated lines and a condensate sensor with pump trip.

**Why must the sample stay above its dew point?**

Below it, water condenses and takes soluble gases with it, changing the sample before it reaches the analyser.

**How far can the sample line be?**

With i-Cons, up to 60 m with the standard pump and 60 to 90 m with the high-flow pump.

**Can you design the system for me?**

Yes. Send the sample temperature, pressure, dew point, dust level and distance to the analyser.

## Related products

[

### i-Cons

Smart Peltier sample conditioner with diagnostics, auto blowback and Modbus, in a compact 19 inch cabinet. View i-Cons

](https://www.sarveshanalytics.com/smart-systems-for-gas-analyser.html)[

### SGI-300 / SGI-320

Peltier sample cooler with integrated pump, filter and condensate detector; SGI-320 for emission monitoring. View SGI-300 / SGI-320

](https://www.sarveshanalytics.com/sample-coolers.html)[

### SP and SRP sample probes

Heated and non-heated probes with blowback, and retractable probes, in SS316, Inconel or ceramic. View SP and SRP sample probes

](https://www.sarveshanalytics.com/sample-probes.html)

## Where this applies

- [Stack and emission monitoring](https://www.sarveshanalytics.com/applications/stack-and-emission-monitoring-system.html)
- [Steel plant process gases](https://www.sarveshanalytics.com/applications/steel-plant-process-gas-analysis.html)
- [Sampling components and systems](https://www.sarveshanalytics.com/sampling-components.html)
- [Sample probes SP-100, SP-200 and SP-300](https://www.sarveshanalytics.com/sample-probes.html)

## Where to go next

- [Sample lines: materials, length, diameter and lag](https://www.sarveshanalytics.com/learn/sample-lines-materials-length-and-lag.html)