# Maintenance schedule and spares for gas analysers

Most analyser downtime is preventable and starts in the sample system. A short schedule and a few spares remove most of it.

 Intermediate 3 min read Lesson 6 of 8[Calibration and maintenance](https://www.sarveshanalytics.com/learn/track-calibration-and-maintenance.html)

Helpful first: [Troubleshooting gas analyser readings: high, low, noisy, slow](https://www.sarveshanalytics.com/learn/troubleshooting-gas-analyser-readings.html)

 Quick answer

Maintain analysers by routine: a daily look at flow, pressure and condensate; a weekly check of filters and drain; a monthly calibration check; and an annual full calibration, sensor check and leak test. Keep spare filter elements, tubing fittings, a sensor or cell and a few seals. Log every action so that trends and repeat faults are visible.

Updated 3 November 2026 · Figures from Sarvesh datasheets

**After this lesson you can**

- Build a maintenance routine by frequency
- Select spares to hold
- Keep a useful log

## A starting schedule

| Interval | Tasks |
|---|---|
| Daily or each shift | Look at the display, flow, pressure and any alarms; check the condensate drain and cooler status |
| Weekly | Inspect filters and the rotameter; check calibration gas pressure; look for leaks or moisture in the sample line |
| Monthly | Calibration check with zero and span, recording the as-found deviation; clean or replace filter elements as needed |
| Quarterly | Check heated lines and probe condition, blowback operation and cooler performance |
| Annually | Full calibration, leak test of the sample system, sensor and cell review, check of alarms and outputs, review of the maintenance log |

These are starting points. Lengthen or shorten each from your own records: [How often should you calibrate a gas analyser](https://www.sarveshanalytics.com/learn/how-often-to-calibrate-a-gas-analyser.html).

## Spares to hold

- Filter elements for each filter in the sample line.
- Tubing, ferrules and fittings of the same material as installed.
- A spare sensor or cell for critical analysers, particularly electrochemical cells with a limited life: [Sensor life, poisoning and replacement in gas analysers](https://www.sarveshanalytics.com/learn/sensor-life-poisoning-and-replacement.html).
- Pump diaphragms or a spare pump where one is used, and seals.
- Calibration gas, regulators and the connectors to use them.
- Fuses and any items the manual lists as consumables.

Stock to match the criticality: a safety interlock analyser deserves a spare sensor on the shelf.

## The log

- Date, who, what was done.
- As-found and as-left calibration values and the gas lot.
- Flow and pressure at the inlet.
- Parts replaced and their lot or date.
- Any faults and the actions taken.

A log shows when a part fails early, when drift is growing and when an interval can be extended.

## Training and support

A trained technician is the best maintenance tool. Operators should know what each alarm means and when to call. For a plant starting out, ask about service and training from the supplier.

**In short**

- Run a daily, weekly, monthly and annual routine.
- Hold spare filters, fittings, sensors and calibration gas.
- Log as-found values, flow, pressure and parts.
- Adjust intervals from data.

## Check your understanding

Answer in your head or on paper first, then open the answer. Recalling an answer teaches more than rereading.

**Which two checks belong in a daily routine?**

Flow and pressure, and the condensate drain and cooler.

**Why keep a spare sensor?**

A critical analyser should not be down while waiting for a sensor to arrive.

**What does a log let you do?**

See trends, repeat faults and decide when to lengthen or shorten intervals.

## Key terms

- [Calibration](https://www.sarveshanalytics.com/learn/glossary.html#t-calibration)
- [Sensor life](https://www.sarveshanalytics.com/learn/glossary.html#t-sensor-life)
- [Fine filter](https://www.sarveshanalytics.com/learn/glossary.html#t-fine-filter)
- [Condensate](https://www.sarveshanalytics.com/learn/glossary.html#t-condensate)

## Frequently asked questions

**How often should an analyser be serviced?**

A routine of daily, weekly, monthly and annual tasks, adjusted from your records.

**Which spares are essential?**

Filter elements, fittings, calibration gas and, for critical analysers, a spare sensor.

**Can Sarvesh service my analyser?**

Contact us with the model and site; support is part of how we work.

**Who should maintain analysers?**

Trained instrument technicians following the manual.

## Related products

[

### ACCURRA-S

Online analyser for trace, percent and purity O2, CO, CO2, CH4 and dew point; can measure two gases from one sample. View ACCURRA-S

](https://www.sarveshanalytics.com/safe-gas-analyser.html)[

### Headline filters

Particulate, coalescing and PTFE membrane filters from Headline Filters, UK, for analyser protection. View Headline filters

](https://www.sarveshanalytics.com/headline-filters.html)

## Where this applies

- [Gas purity certification and cylinder filling](https://www.sarveshanalytics.com/applications/gas-purity-certification-and-cylinder-filling.html)

## Where to go next

- [Records, audit trails and qualification for regulated plants](https://www.sarveshanalytics.com/learn/records-audit-trails-and-qualification-for-regulated-plants.html)