title: Gas Analyser Questions and Answers | Sarvesh
url: https://www.sarveshanalytics.com/learn/questions.html
description: 300 questions about online gas analysers, sampling, calibration, hazardous areas and buying, each answered briefly and linked to a full lesson.
updated: 2026-11-02
source: html
---

Learn · Questions

# Every question, answered

300 questions customers and engineers ask about online gas analysis, grouped by track. Each answer links to the lesson that explains it properly.

## Foundations

**What is the difference between a gas analyser and a gas detector?**

An analyser measures a concentration accurately over a defined range for process control or quality. A detector is a safety device that warns when a gas reaches a hazardous level. They are different product classes.

From the lesson [What is online gas analysis? A plain introduction](https://www.sarveshanalytics.com/learn/what-is-online-gas-analysis.html)

**How fast is an online gas analyser?**

From a few seconds to a few minutes, depending on the sensor, the gas and the sample line. See the lesson on response time.

From the lesson [What is online gas analysis? A plain introduction](https://www.sarveshanalytics.com/learn/what-is-online-gas-analysis.html)

**Do I need a sample line for every analyser?**

No. In-situ analysers measure inside the duct or pipe. Most trace and multi-gas analysers do use a sample line.

From the lesson [What is online gas analysis? A plain introduction](https://www.sarveshanalytics.com/learn/what-is-online-gas-analysis.html)

**What does an online analyser output?**

Typically a 4-20 mA signal, relay alarms and digital data such as Modbus, along with a display.

From the lesson [What is online gas analysis? A plain introduction](https://www.sarveshanalytics.com/learn/what-is-online-gas-analysis.html)

**Can portable analysers replace online analysers?**

Not for control or protection. They give a snapshot, not a continuous reading, and need a person present.

From the lesson [Online vs laboratory vs portable gas analysis](https://www.sarveshanalytics.com/learn/online-vs-laboratory-vs-portable-gas-analysis.html)

**Why use a laboratory if I have an online analyser?**

To verify the online instrument, to measure gases it does not cover, and to certify product to a standard.

From the lesson [Online vs laboratory vs portable gas analysis](https://www.sarveshanalytics.com/learn/online-vs-laboratory-vs-portable-gas-analysis.html)

**Is online analysis more accurate than laboratory analysis?**

Usually not. The laboratory can be more accurate; online analysis is faster and continuous.

From the lesson [Online vs laboratory vs portable gas analysis](https://www.sarveshanalytics.com/learn/online-vs-laboratory-vs-portable-gas-analysis.html)

**Which is cheapest?**

A portable instrument has the lowest outlay; online has the lowest cost per reading when you need continuous data.

From the lesson [Online vs laboratory vs portable gas analysis](https://www.sarveshanalytics.com/learn/online-vs-laboratory-vs-portable-gas-analysis.html)

**Why does gas composition not change when I compress it?**

Compression squeezes all components equally, so each keeps its share of the molecules.

From the lesson [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html)

**What is absolute pressure?**

Pressure measured from a perfect vacuum, as opposed to gauge pressure, which is measured from the surrounding atmosphere.

From the lesson [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html)

**Why do analysers need a pressure reading?**

Some sensors respond to partial pressure, so the analyser measures ambient pressure to correct the reading.

From the lesson [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html)

**Is air really 21 % oxygen?**

Dry air is 20.95 %, usually rounded to 21 % or 20.9 %.

From the lesson [How gases behave: pressure, temperature and partial pressure](https://www.sarveshanalytics.com/learn/how-gases-behave-pressure-temperature-and-partial-pressure.html)

**How many ppm is 1 %?**

10,000 ppm.

From the lesson [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html)

**Is ppm a mass or a volume ratio in gas analysis?**

Volume, unless the specification says otherwise.

From the lesson [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html)

**What does 99.999 % purity mean in ppm of impurity?**

Five nines means 10 ppm total impurity, since 100 minus 99.999 is 0.001 %.

From the lesson [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html)

**Why do some limits use mg/m3?**

Environmental rules usually state mass per volume at reference conditions. See the converter.

From the lesson [ppm, percent and ppb explained for gas analysis](https://www.sarveshanalytics.com/learn/ppm-percent-and-ppb-units-explained.html)

**How do I convert ppm to mg/Nm3?**

Multiply by the molar mass and divide by 22.414, for 0 C and 1013.25 hPa.

From the lesson [ppm to mg/Nm3: volume vs mass concentration](https://www.sarveshanalytics.com/learn/volume-mass-and-mg-per-nm3-concentration-units.html)

**What is the difference between Nm3 and Sm3?**

Nm3 refers to normal conditions, usually 0 C; Sm3 to standard conditions, often 15 or 25 C. Always check the definition in the document.

From the lesson [ppm to mg/Nm3: volume vs mass concentration](https://www.sarveshanalytics.com/learn/volume-mass-and-mg-per-nm3-concentration-units.html)

**Is a wet or dry basis used for emission limits?**

Usually dry, but the permit or standard decides.

From the lesson [ppm to mg/Nm3: volume vs mass concentration](https://www.sarveshanalytics.com/learn/volume-mass-and-mg-per-nm3-concentration-units.html)

**Does temperature change ppm?**

No. ppm is a volume fraction. It changes mg/m3.

From the lesson [ppm to mg/Nm3: volume vs mass concentration](https://www.sarveshanalytics.com/learn/volume-mass-and-mg-per-nm3-concentration-units.html)

**Is oxygen higher or lower when measured wet?**

Lower, because the water vapour takes up part of the volume.

From the lesson [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html)

**How much water is in flue gas?**

Commonly 5 to 20 % by volume depending on the fuel. Gas-fired plants carry more than coal-fired.

From the lesson [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html)

**Do dry-gas analysers see water?**

A cooled sample still holds water at the cooler's dew point, which is usually a few percent or less.

From the lesson [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html)

**Can I measure moisture in the same sample?**

Only if the sample is kept above its dew point, which is a different sample system from a cooled dry one.

From the lesson [Dry basis vs wet basis in gas measurement](https://www.sarveshanalytics.com/learn/dry-basis-vs-wet-basis-gas-measurement.html)

**What is the difference between accuracy and precision?**

Accuracy is how close to the true value; precision is how close repeated readings are to each other.

From the lesson [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html)

**What does plus or minus 1 % of span mean?**

The error can be up to 1 % of the full-scale value anywhere on the range.

From the lesson [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html)

**What is resolution?**

The smallest step the instrument can display, which is not the same as its accuracy.

From the lesson [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html)

**Why does an analyser drift?**

Sensor ageing and temperature effects on the sensor and electronics.

From the lesson [Accuracy, precision, repeatability and drift in gas analysers](https://www.sarveshanalytics.com/learn/accuracy-precision-repeatability-and-drift.html)

**What is a good T90 for a gas analyser?**

It depends on the sensor and gas. Process control often needs seconds to a minute; ask for the T90 at your flow and range.

From the lesson [Response time, T90 and sample lag in gas analysis](https://www.sarveshanalytics.com/learn/response-time-t90-and-sample-lag.html)

**Is T90 the same as response time?**

T90 is the usual way to quote response time. Some datasheets use T95 or T63 instead.

From the lesson [Response time, T90 and sample lag in gas analysis](https://www.sarveshanalytics.com/learn/response-time-t90-and-sample-lag.html)

**Why is my reading slower than the datasheet T90?**

The datasheet describes the analyser alone. Your sample line adds lag in front of it.

From the lesson [Response time, T90 and sample lag in gas analysis](https://www.sarveshanalytics.com/learn/response-time-t90-and-sample-lag.html)

**Can software speed up a reading?**

Not the physical delay. Filters in the control system make it smoother and slower, not faster.

From the lesson [Response time, T90 and sample lag in gas analysis](https://www.sarveshanalytics.com/learn/response-time-t90-and-sample-lag.html)

**What is the most important line on a gas analyser datasheet?**

The sample conditions. An analyser outside them will not perform as stated.

From the lesson [How to read a gas analyser datasheet](https://www.sarveshanalytics.com/learn/reading-a-gas-analyser-datasheet.html)

**What does IP rating tell me?**

How well the enclosure keeps out dust and water, for example IP65.

From the lesson [How to read a gas analyser datasheet](https://www.sarveshanalytics.com/learn/reading-a-gas-analyser-datasheet.html)

**Are datasheet figures guaranteed on my plant?**

They apply under the stated conditions. Your sample and installation can change them, so ask.

From the lesson [How to read a gas analyser datasheet](https://www.sarveshanalytics.com/learn/reading-a-gas-analyser-datasheet.html)

**Where do I find the ACCURRA-S datasheet?**

On the downloads page of this website.

From the lesson [How to read a gas analyser datasheet](https://www.sarveshanalytics.com/learn/reading-a-gas-analyser-datasheet.html)

## Measurement principles

**Which gas sensor is the most accurate?**

It depends on the gas. Each technology is best for the gases it was built for; the right match matters more than the technology's reputation.

From the lesson [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html)

**Can one sensor measure all gases?**

No. Each uses a physical property that only some gases have.

From the lesson [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html)

**What is the difference between a sensor and an analyser?**

The sensor is the element that responds to the gas. The analyser is the whole instrument with electronics, display and outputs.

From the lesson [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html)

**Does Sarvesh make chromatographs or laser analysers?**

No. The range covers electrochemical, paramagnetic, zirconia, NDIR, thermal conductivity and ceramic moisture sensing.

From the lesson [Gas analysis technologies compared: which sensor for which gas](https://www.sarveshanalytics.com/learn/gas-analysis-technologies-compared.html)

**Which sensor is best for ppm oxygen?**

An electrochemical cell, such as ACCURRA-S 3015, which reads 0 to 1000 ppm with a minimum range of 0 to 100 ppm.

From the lesson [Electrochemical vs paramagnetic oxygen sensors](https://www.sarveshanalytics.com/learn/electrochemical-vs-paramagnetic-oxygen-sensors.html)

**Which sensor has the longest life?**

Paramagnetic sensors have no consumed element. Electrochemical cells wear out; the catalogue gives more than 18 months to 2 years depending on model.

From the lesson [Electrochemical vs paramagnetic oxygen sensors](https://www.sarveshanalytics.com/learn/electrochemical-vs-paramagnetic-oxygen-sensors.html)

**Can a paramagnetic sensor measure trace oxygen?**

Paramagnetic sensors are used for percent and purity measurement. Trace oxygen uses electrochemical cells.

From the lesson [Electrochemical vs paramagnetic oxygen sensors](https://www.sarveshanalytics.com/learn/electrochemical-vs-paramagnetic-oxygen-sensors.html)

**Which sensor for oxygen purity?**

Models 3090 and 3190, both with ambient pressure and temperature compensation. 3090 is electrochemical over 0 to 100 %; 3190 is paramagnetic over 95 to 100 %.

From the lesson [Electrochemical vs paramagnetic oxygen sensors](https://www.sarveshanalytics.com/learn/electrochemical-vs-paramagnetic-oxygen-sensors.html)

**How does a zirconia oxygen sensor work?**

A heated zirconia cell conducts oxygen ions and produces a voltage that depends on the oxygen difference between the flue gas and a reference air stream.

From the lesson [How a zirconia oxygen analyser works](https://www.sarveshanalytics.com/learn/how-a-zirconia-oxygen-analyser-works.html)

**Why does zirconia need heating?**

Zirconia conducts oxygen ions only at high temperature, so the cell is heated to a controlled working temperature.

From the lesson [How a zirconia oxygen analyser works](https://www.sarveshanalytics.com/learn/how-a-zirconia-oxygen-analyser-works.html)

**Can a zirconia analyser read low if there is unburnt fuel?**

Yes. Combustibles consume oxygen at the hot cell, so the reading is lower than the true oxygen content.

From the lesson [How a zirconia oxygen analyser works](https://www.sarveshanalytics.com/learn/how-a-zirconia-oxygen-analyser-works.html)

**How is ACCURRA-Zr calibrated?**

With plain air at a single point.

From the lesson [How a zirconia oxygen analyser works](https://www.sarveshanalytics.com/learn/how-a-zirconia-oxygen-analyser-works.html)

**Can NDIR measure oxygen?**

No. Oxygen does not absorb infrared at these wavelengths. Use electrochemical or paramagnetic sensors.

From the lesson [What is an NDIR gas analyser?](https://www.sarveshanalytics.com/learn/what-is-an-ndir-gas-analyser.html)

**Can NDIR measure hydrogen?**

No. Hydrogen is measured by thermal conductivity, for example ACCURRA-TCD.

From the lesson [What is an NDIR gas analyser?](https://www.sarveshanalytics.com/learn/what-is-an-ndir-gas-analyser.html)

**Does NDIR need a reference gas?**

Calibration uses zero and span gas. Normal operation does not consume gas.

From the lesson [What is an NDIR gas analyser?](https://www.sarveshanalytics.com/learn/what-is-an-ndir-gas-analyser.html)

**How long does an NDIR sensor last?**

The ACCURRA-S catalogue gives more than five years for the NDIR models.

From the lesson [What is an NDIR gas analyser?](https://www.sarveshanalytics.com/learn/what-is-an-ndir-gas-analyser.html)

**Do thermal conductivity analysers need a reference gas?**

ACCURRA-TCD does not, according to the catalogue.

From the lesson [How thermal conductivity gas analysers (TCD) work](https://www.sarveshanalytics.com/learn/how-thermal-conductivity-gas-analysers-work.html)

**Which gases can a thermal conductivity analyser measure?**

Any binary mixture whose components differ in conductivity. The ACCURRA-TCD catalogue lists hydrogen, helium, argon, nitrogen, nitrous oxide, neon and krypton.

From the lesson [How thermal conductivity gas analysers (TCD) work](https://www.sarveshanalytics.com/learn/how-thermal-conductivity-gas-analysers-work.html)

**Can it measure oxygen?**

Only in a binary background of a gas with very different conductivity, such as oxygen in hydrogen or in helium.

From the lesson [How thermal conductivity gas analysers (TCD) work](https://www.sarveshanalytics.com/learn/how-thermal-conductivity-gas-analysers-work.html)

**Why does moisture matter?**

Water vapour has its own conductivity, so wet samples shift the reading. Dry the sample or hold its moisture constant.

From the lesson [How thermal conductivity gas analysers (TCD) work](https://www.sarveshanalytics.com/learn/how-thermal-conductivity-gas-analysers-work.html)

**What dew point is dry enough for nitrogen?**

It depends on the use. Heat treatment and electronics lines often specify dew points well below minus 40 C. Your process specification sets the limit.

From the lesson [What is dew point and how is it measured?](https://www.sarveshanalytics.com/learn/what-is-dew-point-and-how-is-it-measured.html)

**Is a lower dew point always better?**

A lower dew point is drier, but cost rises as you dry further. Specify the dew point the process needs.

From the lesson [What is dew point and how is it measured?](https://www.sarveshanalytics.com/learn/what-is-dew-point-and-how-is-it-measured.html)

**How do I convert dew point to ppm?**

Use the saturation vapour pressure at the dew point divided by the total pressure. At atmospheric pressure, minus 60 C is about 11 ppm by volume.

From the lesson [What is dew point and how is it measured?](https://www.sarveshanalytics.com/learn/what-is-dew-point-and-how-is-it-measured.html)

**Can I calibrate dew point on site?**

No. Field calibration is not possible on M015; ask us how the sensor is checked and recalibrated.

From the lesson [What is dew point and how is it measured?](https://www.sarveshanalytics.com/learn/what-is-dew-point-and-how-is-it-measured.html)

**What is the most accurate way to measure dew point?**

A chilled mirror hygrometer is the usual reference, but it needs clean gas and skilled operation.

From the lesson [Moisture measurement methods compared for dry gas](https://www.sarveshanalytics.com/learn/moisture-measurement-methods-compared.html)

**Can I measure dew point with a humidity sensor?**

Relative humidity sensors suit ambient air. Dry process gas needs a sensor made for low dew point.

From the lesson [Moisture measurement methods compared for dry gas](https://www.sarveshanalytics.com/learn/moisture-measurement-methods-compared.html)

**Why does my moisture reading drift after maintenance?**

The tubing re-wets when opened to air and takes time to dry. Purge before reading.

From the lesson [Moisture measurement methods compared for dry gas](https://www.sarveshanalytics.com/learn/moisture-measurement-methods-compared.html)

**Which range does ACCURRA-S M015 cover?**

Minus 80 to plus 20 C dew point with a ceramic sensor, per the datasheet.

From the lesson [Moisture measurement methods compared for dry gas](https://www.sarveshanalytics.com/learn/moisture-measurement-methods-compared.html)

**What is cross-sensitivity?**

A sensor's response to gases other than its target.

From the lesson [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html)

**How do I find a sensor's interferences?**

In the datasheet or the manufacturer's cross-sensitivity table; ask the supplier about every gas in your sample.

From the lesson [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html)

**Can software correct cross-sensitivity?**

Only if the interfering gas is measured too. Otherwise the instrument cannot tell the two effects apart.

From the lesson [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html)

**Is interference the same as drift?**

No. Drift changes with time at a constant gas. Interference changes with the presence of another gas.

From the lesson [Cross-Sensitivity and interference in gas analysers](https://www.sarveshanalytics.com/learn/cross-sensitivity-and-interference-in-gas-analysers.html)

**Why does my trace oxygen reading never reach the expected low value?**

Usually a leak or permeation in the sample line, not the sensor. Leak-test and purge.

From the lesson [Trace oxygen measurement and air ingress: ppm done right](https://www.sarveshanalytics.com/learn/trace-oxygen-measurement-and-air-ingress.html)

**How long does a trace oxygen cell take to recover after air exposure?**

It can take hours at ppm level, depending on the exposure and the cell. Isolate it during maintenance.

From the lesson [Trace oxygen measurement and air ingress: ppm done right](https://www.sarveshanalytics.com/learn/trace-oxygen-measurement-and-air-ingress.html)

**Which analyser measures trace oxygen?**

ACCURRA-S 3015 covers 0 to 1000 ppm with a minimum range of 0 to 100 ppm.

From the lesson [Trace oxygen measurement and air ingress: ppm done right](https://www.sarveshanalytics.com/learn/trace-oxygen-measurement-and-air-ingress.html)

**Can I use plastic tube if the line is short?**

At ppm levels, no. Use stainless steel; it is a small cost against the error.

From the lesson [Trace oxygen measurement and air ingress: ppm done right](https://www.sarveshanalytics.com/learn/trace-oxygen-measurement-and-air-ingress.html)

**How do you measure hydrogen purity online?**

With a thermal conductivity analyser, for a binary mixture. ACCURRA-TCD needs no reference gas.

From the lesson [Measuring hydrogen: purity, mixtures and safety limits](https://www.sarveshanalytics.com/learn/measuring-hydrogen-purity-and-hydrogen-in-gas-mixtures.html)

**Can hydrogen be measured next to methane?**

Not reliably by thermal conductivity alone, since methane changes the reading. Ask an engineer about the composition.

From the lesson [Measuring hydrogen: purity, mixtures and safety limits](https://www.sarveshanalytics.com/learn/measuring-hydrogen-purity-and-hydrogen-in-gas-mixtures.html)

**What range of hydrogen can ACCURRA-TCD measure?**

0 to 100 % depending on the background gas, per the application notes.

From the lesson [Measuring hydrogen: purity, mixtures and safety limits](https://www.sarveshanalytics.com/learn/measuring-hydrogen-purity-and-hydrogen-in-gas-mixtures.html)

**Is ACCURRA-TCD suitable for hazardous areas?**

Use ACCURRA-Hz for oxygen in hazardous areas. Tell us your area classification for other measurements.

From the lesson [Measuring hydrogen: purity, mixtures and safety limits](https://www.sarveshanalytics.com/learn/measuring-hydrogen-purity-and-hydrogen-in-gas-mixtures.html)

**Which ACCURRA-S models measure CO2?**

6030 (0 to 20 % or 0 to 100 %) and 6015 for ppm ranges, by NDIR.

From the lesson [Measuring carbon monoxide and carbon dioxide online](https://www.sarveshanalytics.com/learn/measuring-carbon-monoxide-and-carbon-dioxide.html)

**Which models measure CO?**

2015 for ppm ranges and 2030 for percent ranges. Confirm ranges against the datasheet.

From the lesson [Measuring carbon monoxide and carbon dioxide online](https://www.sarveshanalytics.com/learn/measuring-carbon-monoxide-and-carbon-dioxide.html)

**Can one analyser read both CO and CO2?**

Yes, in a dual-channel enclosure from one sample.

From the lesson [Measuring carbon monoxide and carbon dioxide online](https://www.sarveshanalytics.com/learn/measuring-carbon-monoxide-and-carbon-dioxide.html)

**Why is the sample cooled before an NDIR analyser?**

To remove condensing water that would disturb the optical path and the reading.

From the lesson [Measuring carbon monoxide and carbon dioxide online](https://www.sarveshanalytics.com/learn/measuring-carbon-monoxide-and-carbon-dioxide.html)

**How do you measure methane in biogas?**

By NDIR on a conditioned sample, for example ACCURRA-S 4030.

From the lesson [Measuring methane and biogas composition online](https://www.sarveshanalytics.com/learn/measuring-methane-and-biogas.html)

**Can one analyser measure methane and CO2 together?**

Yes, a dual-channel enclosure reads both from one sample.

From the lesson [Measuring methane and biogas composition online](https://www.sarveshanalytics.com/learn/measuring-methane-and-biogas.html)

**What does biogas contain besides methane?**

CO2, water vapour, and traces of H2S, hydrogen, nitrogen and oxygen.

From the lesson [Measuring methane and biogas composition online](https://www.sarveshanalytics.com/learn/measuring-methane-and-biogas.html)

**Is a biogas analyser safe-area equipment?**

The standard ACCURRA-S is for safe areas. For classified locations, tell us the zone and we will advise.

From the lesson [Measuring methane and biogas composition online](https://www.sarveshanalytics.com/learn/measuring-methane-and-biogas.html)

**Is in-situ or extractive better?**

Neither is better in general. In-situ suits hot flue gas oxygen; extractive suits trace and multi-gas measurements.

From the lesson [Extractive vs In-Situ gas analysis](https://www.sarveshanalytics.com/learn/extractive-vs-in-situ-gas-analysis.html)

**Why do extractive systems fail?**

Most failures are in the sample path: blocked filters, condensate and pump faults.

From the lesson [Extractive vs In-Situ gas analysis](https://www.sarveshanalytics.com/learn/extractive-vs-in-situ-gas-analysis.html)

**Can I measure CO in situ?**

Extractive measurement is the usual approach for CO in flue gas.

From the lesson [Extractive vs In-Situ gas analysis](https://www.sarveshanalytics.com/learn/extractive-vs-in-situ-gas-analysis.html)

**Which Sarvesh product does extractive emission monitoring?**

SGI-320, with a heated probe, heated line, conditioner and analysers.

From the lesson [Extractive vs In-Situ gas analysis](https://www.sarveshanalytics.com/learn/extractive-vs-in-situ-gas-analysis.html)

**Which analyser for 99.99 % nitrogen?**

Trace model 3015. Residual oxygen up to 100 ppm is the measurement.

From the lesson [Choosing an oxygen analyser: trace, percent or purity](https://www.sarveshanalytics.com/learn/choosing-an-oxygen-analyser-trace-percent-or-purity.html)

**Which for PSA oxygen at 93 %?**

The percentage models 3030 or 3130, which cover 0 to 100 %. The purity models start at 95 %.

From the lesson [Choosing an oxygen analyser: trace, percent or purity](https://www.sarveshanalytics.com/learn/choosing-an-oxygen-analyser-trace-percent-or-purity.html)

**Which for cryogenic oxygen at 99 %?**

Purity model 3090 or 3190, with ambient pressure and temperature compensation.

From the lesson [Choosing an oxygen analyser: trace, percent or purity](https://www.sarveshanalytics.com/learn/choosing-an-oxygen-analyser-trace-percent-or-purity.html)

**Which for boiler flue gas?**

ACCURRA-Zr, an in-situ zirconia analyser.

From the lesson [Choosing an oxygen analyser: trace, percent or purity](https://www.sarveshanalytics.com/learn/choosing-an-oxygen-analyser-trace-percent-or-purity.html)

**Why does barometric pressure affect oxygen analysers?**

Many sensors respond to oxygen partial pressure, which scales with total pressure. A fall in pressure lowers the reading even though the gas is unchanged.

From the lesson [Why ambient pressure affects oxygen purity readings](https://www.sarveshanalytics.com/learn/why-ambient-pressure-affects-oxygen-purity-readings.html)

**How much error can pressure cause?**

The ACCURRA-S catalogue states up to 10 % over a day and night swing in the 90 to 100 % range, without compensation.

From the lesson [Why ambient pressure affects oxygen purity readings](https://www.sarveshanalytics.com/learn/why-ambient-pressure-affects-oxygen-purity-readings.html)

**Which ACCURRA-S models compensate?**

Purity models 3090 (electrochemical, 0 to 100 %) and 3190 (paramagnetic, 95 to 100 %).

From the lesson [Why ambient pressure affects oxygen purity readings](https://www.sarveshanalytics.com/learn/why-ambient-pressure-affects-oxygen-purity-readings.html)

**Does compensation remove the need for calibration?**

No. Calibration is still needed, but compensation keeps the calibration valid across weather and seasons.

From the lesson [Why ambient pressure affects oxygen purity readings](https://www.sarveshanalytics.com/learn/why-ambient-pressure-affects-oxygen-purity-readings.html)

## Sampling and conditioning

**What is the most common sample system fault?**

A blocked probe filter or condensate in the line. Both are addressed by blowback, heated lines and a condensate sensor with pump trip.

From the lesson [How to design a gas sample handling system](https://www.sarveshanalytics.com/learn/how-to-design-a-gas-sample-handling-system.html)

**Why must the sample stay above its dew point?**

Below it, water condenses and takes soluble gases with it, changing the sample before it reaches the analyser.

From the lesson [How to design a gas sample handling system](https://www.sarveshanalytics.com/learn/how-to-design-a-gas-sample-handling-system.html)

**How far can the sample line be?**

With i-Cons, up to 60 m with the standard pump and 60 to 90 m with the high-flow pump.

From the lesson [How to design a gas sample handling system](https://www.sarveshanalytics.com/learn/how-to-design-a-gas-sample-handling-system.html)

**Can you design the system for me?**

Yes. Send the sample temperature, pressure, dew point, dust level and distance to the analyser.

From the lesson [How to design a gas sample handling system](https://www.sarveshanalytics.com/learn/how-to-design-a-gas-sample-handling-system.html)

**Where should a sample point be?**

At a well-mixed location, upstream of air leaks, at known process conditions and accessible for maintenance.

From the lesson [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html)

**How far downstream of a gas injection point should I sample?**

Far enough for mixing, typically several pipe diameters of turbulent flow or a mixing element. Ask for advice for your case.

From the lesson [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html)

**Does the tapping affect response time?**

Yes. A large or dead tapping volume adds lag.

From the lesson [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html)

**Can you advise on the sample point?**

Yes. Send a drawing or photographs with the application data sheet.

From the lesson [Choosing the sample point for a gas analyser](https://www.sarveshanalytics.com/learn/choosing-the-sample-point.html)

**What tubing is best for gas analysis?**

316 stainless steel for most trace and reactive work.

From the lesson [Sample lines: materials, length, diameter and lag](https://www.sarveshanalytics.com/learn/sample-lines-materials-length-and-lag.html)

**How long can a sample line be?**

As short as practical; the limit is the acceptable lag, which depends on bore, flow and pressure.

From the lesson [Sample lines: materials, length, diameter and lag](https://www.sarveshanalytics.com/learn/sample-lines-materials-length-and-lag.html)

**Can I use flexible hose?**

For short connections to the analyser in non-critical percent measurements. Use metal for ppm and moisture.

From the lesson [Sample lines: materials, length, diameter and lag](https://www.sarveshanalytics.com/learn/sample-lines-materials-length-and-lag.html)

**Do I need a heated line?**

If the sample is wet and hot, yes, until it reaches a cooler. Dry gases at ambient temperature do not.

From the lesson [Sample lines: materials, length, diameter and lag](https://www.sarveshanalytics.com/learn/sample-lines-materials-length-and-lag.html)

**How often should analyser filters be changed?**

By experience at the site; start with a regular interval and shorten it if the pressure drop or flow shows blocking.

From the lesson [Filters for gas analysers: dust, mist, membrane](https://www.sarveshanalytics.com/learn/filters-and-particulate-protection.html)

**Can I clean and reuse a filter element?**

Not reliably. Replace it.

From the lesson [Filters for gas analysers: dust, mist, membrane](https://www.sarveshanalytics.com/learn/filters-and-particulate-protection.html)

**What dust size can ACCURRA-S accept?**

The catalogue specifies dust below 3 micron.

From the lesson [Filters for gas analysers: dust, mist, membrane](https://www.sarveshanalytics.com/learn/filters-and-particulate-protection.html)

**Who supplies the filters?**

Sarvesh represents Headline Filters, UK, in India.

From the lesson [Filters for gas analysers: dust, mist, membrane](https://www.sarveshanalytics.com/learn/filters-and-particulate-protection.html)

**What temperature should a sample cooler run at?**

Often 3 to 5 C, low enough to remove most water without freezing. Follow the cooler's specification.

From the lesson [Condensate, moisture and sample cooling for gas analysis](https://www.sarveshanalytics.com/learn/condensate-moisture-and-sample-cooling.html)

**What is the difference between SGI-300 and i-Cons?**

SGI-300 is a Peltier cooler with pump, filter and condensate detector; i-Cons is a smart Peltier conditioner with diagnostics, auto blowback and Modbus in a compact 19 inch cabinet.

From the lesson [Condensate, moisture and sample cooling for gas analysis](https://www.sarveshanalytics.com/learn/condensate-moisture-and-sample-cooling.html)

**Can I measure moisture after a cooler?**

Only to the cooler's dew point. If you need the process moisture, the sample must not be cooled.

From the lesson [Condensate, moisture and sample cooling for gas analysis](https://www.sarveshanalytics.com/learn/condensate-moisture-and-sample-cooling.html)

**Why does my CO2 reading differ before and after the cooler?**

Soluble CO2 partly dissolves in condensate; the difference should be small and steady.

From the lesson [Condensate, moisture and sample cooling for gas analysis](https://www.sarveshanalytics.com/learn/condensate-moisture-and-sample-cooling.html)

**What flow does an ACCURRA-S need?**

About 1 LPM, per the catalogue, with a regulated pressure of 2 to 10 psig.

From the lesson [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html)

**Do I need a pressure regulator?**

Yes when the process pressure is higher than the analyser's range or varies.

From the lesson [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html)

**Where should the sample vent?**

To a safe place suited to the gas; flammable or toxic samples go to a flare or recovery.

From the lesson [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html)

**What is a fast loop?**

A bypass that circulates a large flow near the tapping so the analyser takes a quick, small slice.

From the lesson [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html)

**What material should a probe be?**

SS316 for general duty, Inconel for higher temperature or corrosion, ceramic for the hottest duties.

From the lesson [Probes and blowback for hot, dusty and tarry gas](https://www.sarveshanalytics.com/learn/hot-dusty-gas-probes-and-blowback.html)

**How often should blowback run?**

Set by the dust load; start frequent and relax once the probe stays clean.

From the lesson [Probes and blowback for hot, dusty and tarry gas](https://www.sarveshanalytics.com/learn/hot-dusty-gas-probes-and-blowback.html)

**Can I sample tarry gas?**

Yes with a heated probe, heated line and conditioning matched to the tar, after review of the composition.

From the lesson [Probes and blowback for hot, dusty and tarry gas](https://www.sarveshanalytics.com/learn/hot-dusty-gas-probes-and-blowback.html)

**What is the difference between SP and SRP probes?**

SP are fixed probes; SRP are retractable.

From the lesson [Probes and blowback for hot, dusty and tarry gas](https://www.sarveshanalytics.com/learn/hot-dusty-gas-probes-and-blowback.html)

**Why can't I analyse liquid oxygen directly?**

Analysers measure gas. The liquid must be vaporised, and completely, to keep its composition.

From the lesson [Sampling cryogenic and liquefied gases: vaporisers](https://www.sarveshanalytics.com/learn/sampling-cryogenic-and-liquefied-gases.html)

**Which vaporiser does Sarvesh offer for cryogenic samples?**

SV-200, with a bypass line; SV-400 is also available.

From the lesson [Sampling cryogenic and liquefied gases: vaporisers](https://www.sarveshanalytics.com/learn/sampling-cryogenic-and-liquefied-gases.html)

**Can I analyse LPG with ACCURRA-S?**

Flammable samples are outside the standard ACCURRA-S sample specification; ask an engineer.

From the lesson [Sampling cryogenic and liquefied gases: vaporisers](https://www.sarveshanalytics.com/learn/sampling-cryogenic-and-liquefied-gases.html)

**How do I flush a vaporiser?**

Use the bypass line, per the product instructions.

From the lesson [Sampling cryogenic and liquefied gases: vaporisers](https://www.sarveshanalytics.com/learn/sampling-cryogenic-and-liquefied-gases.html)

**Can ACCURRA-S measure corrosive gases?**

Not in the standard configuration. Send the stream composition and we will advise.

From the lesson [What samples an analyser cannot take: know the limits](https://www.sarveshanalytics.com/learn/what-samples-an-analyser-cannot-take.html)

**Can you analyse H2S?**

No; H2S is outside the ACCURRA-S gas list and the i-Cons and SGI-320 limits.

From the lesson [What samples an analyser cannot take: know the limits](https://www.sarveshanalytics.com/learn/what-samples-an-analyser-cannot-take.html)

**Can I put an analyser on raw biogas?**

Only after conditioning; moisture and H2S must be dealt with first.

From the lesson [What samples an analyser cannot take: know the limits](https://www.sarveshanalytics.com/learn/what-samples-an-analyser-cannot-take.html)

**How do I find out if my sample is suitable?**

Send the gas composition, pressure, temperature, dust and moisture with an application data sheet.

From the lesson [What samples an analyser cannot take: know the limits](https://www.sarveshanalytics.com/learn/what-samples-an-analyser-cannot-take.html)

**Why is my analyser reading wrong?**

Check the sample first: flow, filters, leaks, condensate. Then apply a known gas.

From the lesson [Sample system troubleshooting: symptoms to causes](https://www.sarveshanalytics.com/learn/sample-system-troubleshooting.html)

**How do I find a leak?**

Pressurise with inert gas and use a leak-detection fluid or a pressure-hold test.

From the lesson [Sample system troubleshooting: symptoms to causes](https://www.sarveshanalytics.com/learn/sample-system-troubleshooting.html)

**Why does water get into the analyser?**

A failed cooler or blocked drain, or a line below the dew point.

From the lesson [Sample system troubleshooting: symptoms to causes](https://www.sarveshanalytics.com/learn/sample-system-troubleshooting.html)

**When should I call an engineer?**

When known gas reads wrong, or after the sample checks are clean.

From the lesson [Sample system troubleshooting: symptoms to causes](https://www.sarveshanalytics.com/learn/sample-system-troubleshooting.html)

## Installation and integration

**Can an ACCURRA-S be field mounted?**

It is built for panel or rack mounting. For field use, ask about an enclosure suited to the site.

From the lesson [Where to mount a gas analyser: panel, shelter or field](https://www.sarveshanalytics.com/learn/where-to-mount-a-gas-analyser.html)

**How far can the analyser be from the sample point?**

As far as the acceptable lag allows; use a fast loop for long runs.

From the lesson [Where to mount a gas analyser: panel, shelter or field](https://www.sarveshanalytics.com/learn/where-to-mount-a-gas-analyser.html)

**Do analysers need heating in winter?**

Ambient limits apply; in cold sites a heated shelter or enclosure avoids condensation.

From the lesson [Where to mount a gas analyser: panel, shelter or field](https://www.sarveshanalytics.com/learn/where-to-mount-a-gas-analyser.html)

**What IP rating do I need?**

Choose by the dust and water exposure at the site; indoors IP20 to IP54 is common, outdoors IP65 or higher.

From the lesson [Where to mount a gas analyser: panel, shelter or field](https://www.sarveshanalytics.com/learn/where-to-mount-a-gas-analyser.html)

**What is a 4-20 mA signal?**

A current that represents the reading, 4 mA at the bottom of the range and 20 mA at the top.

From the lesson [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html)

**Which outputs does ACCURRA-S have?**

Configurable 4-20 mA, relay alarms and Modbus RTU, according to the catalogue.

From the lesson [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html)

**What is the maximum cable length for 4-20 mA?**

Hundreds of metres, limited by the loop's compliance voltage and wire resistance.

From the lesson [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html)

**Can Modbus start a calibration?**

On models that support it; ask for the register map.

From the lesson [Analyser outputs: 4-20 mA, relays and Modbus explained](https://www.sarveshanalytics.com/learn/analyser-outputs-4-20-ma-relays-and-modbus.html)

**How far apart should power and signal cables run?**

Use separate trays; a few tens of centimetres of separation is common, and follow your plant's standard.

From the lesson [Wiring, grounding and electrical noise for gas analysers](https://www.sarveshanalytics.com/learn/wiring-grounding-and-electrical-noise.html)

**Can I run Modbus beside power cables?**

Avoid it; use shielded twisted pair and keep it separate.

From the lesson [Wiring, grounding and electrical noise for gas analysers](https://www.sarveshanalytics.com/learn/wiring-grounding-and-electrical-noise.html)

**What causes a noisy 4-20 mA signal?**

Ground loops, unshielded cable, cables beside power, or a loose terminal.

From the lesson [Wiring, grounding and electrical noise for gas analysers](https://www.sarveshanalytics.com/learn/wiring-grounding-and-electrical-noise.html)

**Do I need surge protection?**

For long outdoor cables, yes.

From the lesson [Wiring, grounding and electrical noise for gas analysers](https://www.sarveshanalytics.com/learn/wiring-grounding-and-electrical-noise.html)

**How do I scale a 4-20 mA input?**

Set 4 mA to the low end and 20 mA to the high end of the analyser range, in the same units.

From the lesson [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

**What should happen when the analyser fails?**

That is a process decision: hold, switch to manual, alarm or trip. Decide it in advance.

From the lesson [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

**Can the control system start an analyser calibration?**

Where the analyser supports it by Modbus or a digital input.

From the lesson [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

**How do I avoid nuisance alarms?**

Use a deadband and a short delay, and separate warning from trip levels.

From the lesson [Connecting a gas analyser to PLC, DCS and SCADA](https://www.sarveshanalytics.com/learn/connecting-an-analyser-to-plc-dcs-and-scada.html)

**What power does an analyser need?**

Check the model datasheet for voltage and consumption; use a clean supply.

From the lesson [Site utilities checklist for gas analysers](https://www.sarveshanalytics.com/learn/site-utilities-checklist-for-analysers.html)

**Do I need instrument air for every analyser?**

Only for functions such as blowback or air purges, and for some probes.

From the lesson [Site utilities checklist for gas analysers](https://www.sarveshanalytics.com/learn/site-utilities-checklist-for-analysers.html)

**Where do I store calibration cylinders?**

In a ventilated, secured place, with cylinders restrained, away from heat and incompatible gases.

From the lesson [Site utilities checklist for gas analysers](https://www.sarveshanalytics.com/learn/site-utilities-checklist-for-analysers.html)

**Can I vent to the room?**

Only if the sample is harmless in the quantity vented; follow your site safety rules.

From the lesson [Site utilities checklist for gas analysers](https://www.sarveshanalytics.com/learn/site-utilities-checklist-for-analysers.html)

**How long does commissioning take?**

From a few hours for a simple rack analyser to days for a system with probes and heated lines.

From the lesson [Commissioning an online gas analyser: A Step-by-Step guide](https://www.sarveshanalytics.com/learn/commissioning-an-online-gas-analyser.html)

**Should I calibrate before or after purging?**

After purging and leak-testing, so the calibration reflects a clean system.

From the lesson [Commissioning an online gas analyser: A Step-by-Step guide](https://www.sarveshanalytics.com/learn/commissioning-an-online-gas-analyser.html)

**What documents do I need at hand-over?**

Drawings, register map, calibration record, spares and maintenance intervals.

From the lesson [Commissioning an online gas analyser: A Step-by-Step guide](https://www.sarveshanalytics.com/learn/commissioning-an-online-gas-analyser.html)

**Can you commission for me?**

Ask us about service support for your site.

From the lesson [Commissioning an online gas analyser: A Step-by-Step guide](https://www.sarveshanalytics.com/learn/commissioning-an-online-gas-analyser.html)

## Calibration and maintenance

**What is the difference between calibration and a calibration check?**

A check compares and records; a calibration also adjusts the instrument.

From the lesson [How gas analyser calibration works: zero and span](https://www.sarveshanalytics.com/learn/how-gas-analyser-calibration-works.html)

**What span gas concentration should I use?**

Typically 70 to 90 % of the range, in a background similar to the sample, with a certificate.

From the lesson [How gas analyser calibration works: zero and span](https://www.sarveshanalytics.com/learn/how-gas-analyser-calibration-works.html)

**Can I calibrate oxygen with air?**

Yes for percent ranges, using clean dry air at 20.9 %; not for trace ranges.

From the lesson [How gas analyser calibration works: zero and span](https://www.sarveshanalytics.com/learn/how-gas-analyser-calibration-works.html)

**How do I know calibration worked?**

Re-apply the gas and confirm the reading; then confirm the process reading returns.

From the lesson [How gas analyser calibration works: zero and span](https://www.sarveshanalytics.com/learn/how-gas-analyser-calibration-works.html)

**What calibration gas do I need for ACCURRA-S 3015?**

A zero gas with very low oxygen and a span gas in nitrogen near the top of your range, with a certificate.

From the lesson [Calibration gases: choosing, handling and certificates](https://www.sarveshanalytics.com/learn/calibration-gases-choosing-and-handling.html)

**How long does a calibration cylinder last?**

Until the expiry on the certificate, and until pressure falls to the cylinder maker's minimum.

From the lesson [Calibration gases: choosing, handling and certificates](https://www.sarveshanalytics.com/learn/calibration-gases-choosing-and-handling.html)

**Can I make my own calibration gas?**

Not for a measurement that matters. Certified gas gives a traceable result.

From the lesson [Calibration gases: choosing, handling and certificates](https://www.sarveshanalytics.com/learn/calibration-gases-choosing-and-handling.html)

**Do I need different gas for each analyser?**

Only if the ranges, gases or backgrounds differ.

From the lesson [Calibration gases: choosing, handling and certificates](https://www.sarveshanalytics.com/learn/calibration-gases-choosing-and-handling.html)

**How often should an oxygen analyser be calibrated?**

Start monthly and adjust with as-found data; safety and purity measurements may justify more often.

From the lesson [How often should you calibrate a gas analyser?](https://www.sarveshanalytics.com/learn/how-often-to-calibrate-a-gas-analyser.html)

**Is annual calibration enough?**

Only if drift records show it, and the consequence of error is small.

From the lesson [How often should you calibrate a gas analyser?](https://www.sarveshanalytics.com/learn/how-often-to-calibrate-a-gas-analyser.html)

**Does auto-calibration replace manual calibration?**

It can reduce manual work, but you still verify with certified gas and review the results.

From the lesson [How often should you calibrate a gas analyser?](https://www.sarveshanalytics.com/learn/how-often-to-calibrate-a-gas-analyser.html)

**Who sets the interval?**

You do, from your tolerance and records, unless a standard or quality system specifies one.

From the lesson [How often should you calibrate a gas analyser?](https://www.sarveshanalytics.com/learn/how-often-to-calibrate-a-gas-analyser.html)

**Is auto-calibration available on all ACCURRA-S models?**

It is optional on most; 3015 and M015 do not have it.

From the lesson [Auto-Calibration and validation checks for gas analysers](https://www.sarveshanalytics.com/learn/auto-calibration-and-validation-checks.html)

**How often should auto-calibration run?**

Daily or weekly is a common start; adjust to your drift records.

From the lesson [Auto-Calibration and validation checks for gas analysers](https://www.sarveshanalytics.com/learn/auto-calibration-and-validation-checks.html)

**Does auto-calibration reduce the need for cylinders?**

No, it needs certified gas connected to the analyser.

From the lesson [Auto-Calibration and validation checks for gas analysers](https://www.sarveshanalytics.com/learn/auto-calibration-and-validation-checks.html)

**Can it start from the control system?**

Via Modbus command on models that support it.

From the lesson [Auto-Calibration and validation checks for gas analysers](https://www.sarveshanalytics.com/learn/auto-calibration-and-validation-checks.html)

**Why is my analyser reading zero?**

Check the output scaling, fault state, sensor and valve positions; then apply span gas.

From the lesson [Troubleshooting gas analyser readings: high, low, noisy, slow](https://www.sarveshanalytics.com/learn/troubleshooting-gas-analyser-readings.html)

**Why does the reading drift upward?**

Temperature change, a leak, calibration gas exhausted or sensor ageing.

From the lesson [Troubleshooting gas analyser readings: high, low, noisy, slow](https://www.sarveshanalytics.com/learn/troubleshooting-gas-analyser-readings.html)

**How do I know if the sensor is dead?**

Large span corrections, slow response or a zero that will not settle.

From the lesson [Troubleshooting gas analyser readings: high, low, noisy, slow](https://www.sarveshanalytics.com/learn/troubleshooting-gas-analyser-readings.html)

**Where are the alarm and calibration records?**

ACCURRA-S saves calibration with deviation values and alarm history in the analyser.

From the lesson [Troubleshooting gas analyser readings: high, low, noisy, slow](https://www.sarveshanalytics.com/learn/troubleshooting-gas-analyser-readings.html)

**How often should an analyser be serviced?**

A routine of daily, weekly, monthly and annual tasks, adjusted from your records.

From the lesson [Maintenance schedule and spares for gas analysers](https://www.sarveshanalytics.com/learn/maintenance-schedule-and-spares-for-gas-analysers.html)

**Which spares are essential?**

Filter elements, fittings, calibration gas and, for critical analysers, a spare sensor.

From the lesson [Maintenance schedule and spares for gas analysers](https://www.sarveshanalytics.com/learn/maintenance-schedule-and-spares-for-gas-analysers.html)

**Can Sarvesh service my analyser?**

Contact us with the model and site; support is part of how we work.

From the lesson [Maintenance schedule and spares for gas analysers](https://www.sarveshanalytics.com/learn/maintenance-schedule-and-spares-for-gas-analysers.html)

**Who should maintain analysers?**

Trained instrument technicians following the manual.

From the lesson [Maintenance schedule and spares for gas analysers](https://www.sarveshanalytics.com/learn/maintenance-schedule-and-spares-for-gas-analysers.html)

**How long do gas sensors last?**

It depends on the type and the sample. Electrochemical cells months to years; others longer.

From the lesson [Sensor life, poisoning and replacement in gas analysers](https://www.sarveshanalytics.com/learn/sensor-life-poisoning-and-replacement.html)

**Can a poisoned sensor recover?**

Sometimes slowly if the exposure was brief; usually replace it.

From the lesson [Sensor life, poisoning and replacement in gas analysers](https://www.sarveshanalytics.com/learn/sensor-life-poisoning-and-replacement.html)

**Should I replace sensors on a schedule?**

For critical points yes, informed by the calibration trend.

From the lesson [Sensor life, poisoning and replacement in gas analysers](https://www.sarveshanalytics.com/learn/sensor-life-poisoning-and-replacement.html)

**Does ACCURRA-S sensor choice affect life?**

Yes: electrochemical cells are consumed; paramagnetic models have no consumed part.

From the lesson [Sensor life, poisoning and replacement in gas analysers](https://www.sarveshanalytics.com/learn/sensor-life-poisoning-and-replacement.html)

**Does ACCURRA-S support audit trails?**

It stores calibration with deviation values and alarm history, and password-protects set-up. Tell us your exact requirements.

From the lesson [Records, audit trails and qualification for regulated plants](https://www.sarveshanalytics.com/learn/records-audit-trails-and-qualification-for-regulated-plants.html)

**Is ACCURRA-S compliant with a specific regulation?**

We make no such claim; ask for a point-by-point answer for your requirement.

From the lesson [Records, audit trails and qualification for regulated plants](https://www.sarveshanalytics.com/learn/records-audit-trails-and-qualification-for-regulated-plants.html)

**Who writes the qualification documents?**

Your quality team, with documents and data from the supplier.

From the lesson [Records, audit trails and qualification for regulated plants](https://www.sarveshanalytics.com/learn/records-audit-trails-and-qualification-for-regulated-plants.html)

**Can records be archived electronically?**

Via the analyser's digital outputs into your own system.

From the lesson [Records, audit trails and qualification for regulated plants](https://www.sarveshanalytics.com/learn/records-audit-trails-and-qualification-for-regulated-plants.html)

## Safety and hazardous areas

**What is the explosive range of hydrogen?**

About 4 to 75 % by volume in air.

From the lesson [Flammability, explosive limits and gas safety basics](https://www.sarveshanalytics.com/learn/flammability-explosive-limits-and-gas-safety-basics.html)

**Is a process analyser a safety device?**

It can be part of a protection scheme, but it is not a personal gas detector and carries no SIL claim from us.

From the lesson [Flammability, explosive limits and gas safety basics](https://www.sarveshanalytics.com/learn/flammability-explosive-limits-and-gas-safety-basics.html)

**What does LEL stand for?**

Lower explosive limit, also called lower flammable limit.

From the lesson [Flammability, explosive limits and gas safety basics](https://www.sarveshanalytics.com/learn/flammability-explosive-limits-and-gas-safety-basics.html)

**Does pressure change the explosive limits?**

Yes; the range widens with pressure and temperature. Use design data for your conditions.

From the lesson [Flammability, explosive limits and gas safety basics](https://www.sarveshanalytics.com/learn/flammability-explosive-limits-and-gas-safety-basics.html)

**What is a typical LOC?**

For many organic vapours in nitrogen, roughly 8 to 12 % oxygen; for hydrogen in nitrogen, near 5 %. Use values for your own materials.

From the lesson [What is limiting oxygen concentration (LOC)?](https://www.sarveshanalytics.com/learn/what-is-limiting-oxygen-concentration.html)

**What oxygen alarm should I set?**

The value comes from your hazard study, as a margin below the LOC. The analyser alarm points are configurable.

From the lesson [What is limiting oxygen concentration (LOC)?](https://www.sarveshanalytics.com/learn/what-is-limiting-oxygen-concentration.html)

**Which analyser for a blanketed tank?**

ACCURRA-S 3015 for ppm, 3030 or 3130 for percent, and ACCURRA-Hz or INFINITY-Tx in hazardous areas.

From the lesson [What is limiting oxygen concentration (LOC)?](https://www.sarveshanalytics.com/learn/what-is-limiting-oxygen-concentration.html)

**Does nitrogen blanketing need continuous monitoring?**

Continuous monitoring is the usual way to demonstrate the barrier holds, and it permits a smaller margin under common guidance.

From the lesson [What is limiting oxygen concentration (LOC)?](https://www.sarveshanalytics.com/learn/what-is-limiting-oxygen-concentration.html)

**How do you purge a tank with nitrogen?**

By sweep-through or dilution to a target oxygen, then confirm with an analyser.

From the lesson [Inerting and purging: how they work and how to prove them](https://www.sarveshanalytics.com/learn/inerting-and-purging-how-they-work.html)

**How much nitrogen do I need to purge a vessel?**

About three vessel volumes for 1 % oxygen if well mixed, more in practice.

From the lesson [Inerting and purging: how they work and how to prove them](https://www.sarveshanalytics.com/learn/inerting-and-purging-how-they-work.html)

**Which standard sets inerting margins?**

Standards such as NFPA 69 give guidance; follow the one that applies to your site.

From the lesson [Inerting and purging: how they work and how to prove them](https://www.sarveshanalytics.com/learn/inerting-and-purging-how-they-work.html)

**What oxygen alarm should I set?**

Below the LOC by the margin your safety assessment requires.

From the lesson [Inerting and purging: how they work and how to prove them](https://www.sarveshanalytics.com/learn/inerting-and-purging-how-they-work.html)

**What does Zone 2 mean?**

An area where an explosive gas atmosphere is not likely in normal operation and, if it occurs, lasts only briefly.

From the lesson [Hazardous area classification: zones, gas groups, T-Classes](https://www.sarveshanalytics.com/learn/hazardous-area-classification-zones-gas-groups-temperature-classes.html)

**Is IIC better than IIB?**

IIC equipment is suitable for a wider range of gases, since it meets a more demanding test.

From the lesson [Hazardous area classification: zones, gas groups, T-Classes](https://www.sarveshanalytics.com/learn/hazardous-area-classification-zones-gas-groups-temperature-classes.html)

**What does T6 mean?**

The equipment's surface will not exceed 85 C.

From the lesson [Hazardous area classification: zones, gas groups, T-Classes](https://www.sarveshanalytics.com/learn/hazardous-area-classification-zones-gas-groups-temperature-classes.html)

**Who decides the zone?**

The plant owner or its consultant, documented on the area classification drawings.

From the lesson [Hazardous area classification: zones, gas groups, T-Classes](https://www.sarveshanalytics.com/learn/hazardous-area-classification-zones-gas-groups-temperature-classes.html)

**What does Gas Group IIC mean?**

The most demanding gas group, which includes hydrogen and acetylene. IIC-certified equipment also suits IIA and IIB gases.

From the lesson [Hazardous area gas analysers explained](https://www.sarveshanalytics.com/learn/hazardous-area-gas-analysers-explained.html)

**What does T6 mean?**

A maximum surface temperature of 85 C.

From the lesson [Hazardous area gas analysers explained](https://www.sarveshanalytics.com/learn/hazardous-area-gas-analysers-explained.html)

**What is a flameproof enclosure?**

An Ex d enclosure that withstands an internal explosion and prevents it igniting the outside atmosphere.

From the lesson [Hazardous area gas analysers explained](https://www.sarveshanalytics.com/learn/hazardous-area-gas-analysers-explained.html)

**Is ACCURRA-Hz SIL rated?**

We make no SIL claim. If your project requires SIL-rated devices, tell us at the start.

From the lesson [Hazardous area gas analysers explained](https://www.sarveshanalytics.com/learn/hazardous-area-gas-analysers-explained.html)

**What does Ex d mean?**

Flameproof: the enclosure contains an internal explosion and prevents it igniting the outside atmosphere.

From the lesson [Protection concepts: flameproof, intrinsic safety, purging](https://www.sarveshanalytics.com/learn/protection-concepts-flameproof-intrinsic-safety-and-purging.html)

**What is intrinsic safety?**

A concept that limits electrical energy so no spark or heat can ignite gas.

From the lesson [Protection concepts: flameproof, intrinsic safety, purging](https://www.sarveshanalytics.com/learn/protection-concepts-flameproof-intrinsic-safety-and-purging.html)

**Can I open a flameproof analyser in a hazardous area?**

Only following the manufacturer's rules; ACCURRA-Hz allows calibration without opening the covers.

From the lesson [Protection concepts: flameproof, intrinsic safety, purging](https://www.sarveshanalytics.com/learn/protection-concepts-flameproof-intrinsic-safety-and-purging.html)

**Do I need a barrier for INFINITY-Tx?**

Check the certificate and the loop's entity parameters; ask us for the installation requirements.

From the lesson [Protection concepts: flameproof, intrinsic safety, purging](https://www.sarveshanalytics.com/learn/protection-concepts-flameproof-intrinsic-safety-and-purging.html)

**Are ACCURRA analysers SIL rated?**

We make no SIL claim. Tell us your requirement and we will respond point by point.

From the lesson [What SIL means for gas analysers, and what it does not](https://www.sarveshanalytics.com/learn/what-sil-means-for-gas-analysers.html)

**What is the difference between SIL and Ex certification?**

Ex certification addresses ignition of explosive atmospheres; SIL addresses the reliability of a safety function.

From the lesson [What SIL means for gas analysers, and what it does not](https://www.sarveshanalytics.com/learn/what-sil-means-for-gas-analysers.html)

**Can a gas analyser be part of a safety loop?**

Yes, with a proper safety design, but it needs careful treatment of sample, drift and response time.

From the lesson [What SIL means for gas analysers, and what it does not](https://www.sarveshanalytics.com/learn/what-sil-means-for-gas-analysers.html)

**What is proof testing?**

A periodic test that reveals hidden failures in a safety function.

From the lesson [What SIL means for gas analysers, and what it does not](https://www.sarveshanalytics.com/learn/what-sil-means-for-gas-analysers.html)

**What oxygen level is safe to breathe?**

Normal air is 20.95 %; below about 19.5 % is generally treated as deficient and above about 23.5 % as enriched.

From the lesson [Oxygen enrichment and oxygen deficiency hazards](https://www.sarveshanalytics.com/learn/oxygen-enrichment-and-oxygen-deficiency-hazards.html)

**Is nitrogen toxic?**

No, but it displaces oxygen and can cause asphyxiation in closed spaces.

From the lesson [Oxygen enrichment and oxygen deficiency hazards](https://www.sarveshanalytics.com/learn/oxygen-enrichment-and-oxygen-deficiency-hazards.html)

**Can a process oxygen analyser be used as a personal monitor?**

No; use a device built and certified for personal or area monitoring.

From the lesson [Oxygen enrichment and oxygen deficiency hazards](https://www.sarveshanalytics.com/learn/oxygen-enrichment-and-oxygen-deficiency-hazards.html)

**How should an oxygen analyser vent?**

To a safe place suited to the gas, away from people and ignition sources.

From the lesson [Oxygen enrichment and oxygen deficiency hazards](https://www.sarveshanalytics.com/learn/oxygen-enrichment-and-oxygen-deficiency-hazards.html)

## Process primers

**What oxygen level should a boiler run at?**

It depends on fuel and burner; take the target from the burner supplier and your commissioning, and trim to just above the CO breakpoint.

From the lesson [Combustion, excess air and flue gas oxygen explained](https://www.sarveshanalytics.com/learn/combustion-excess-air-and-flue-gas-oxygen.html)

**Why is oxygen measured in flue gas?**

It is the simplest indicator of excess air and therefore of fuel efficiency.

From the lesson [Combustion, excess air and flue gas oxygen explained](https://www.sarveshanalytics.com/learn/combustion-excess-air-and-flue-gas-oxygen.html)

**Why measure CO too?**

To make sure that reducing excess air does not cause incomplete combustion.

From the lesson [Combustion, excess air and flue gas oxygen explained](https://www.sarveshanalytics.com/learn/combustion-excess-air-and-flue-gas-oxygen.html)

**Which Sarvesh product measures flue gas oxygen?**

ACCURRA-Zr, an in-situ zirconia analyser.

From the lesson [Combustion, excess air and flue gas oxygen explained](https://www.sarveshanalytics.com/learn/combustion-excess-air-and-flue-gas-oxygen.html)

**What are the products of an air separation plant?**

Oxygen, nitrogen and argon, as gas and liquid, often with instrument air.

From the lesson [How an air separation plant works and what is measured](https://www.sarveshanalytics.com/learn/how-an-air-separation-plant-works.html)

**Which measurements matter most in an ASU?**

Oxygen purity, trace oxygen in nitrogen and argon, and moisture.

From the lesson [How an air separation plant works and what is measured](https://www.sarveshanalytics.com/learn/how-an-air-separation-plant-works.html)

**Why measure CO2 at the purifier?**

A rising CO2 shows the adsorber is exhausted and CO2 could freeze downstream.

From the lesson [How an air separation plant works and what is measured](https://www.sarveshanalytics.com/learn/how-an-air-separation-plant-works.html)

**Do you supply analysers for ASUs?**

Yes; see the industrial gas plant pages.

From the lesson [How an air separation plant works and what is measured](https://www.sarveshanalytics.com/learn/how-an-air-separation-plant-works.html)

**What is PSA?**

Pressure swing adsorption: beds of adsorbent separate gases at pressure and are regenerated by lowering pressure.

From the lesson [PSA, VPSA and membrane gas generators explained](https://www.sarveshanalytics.com/learn/nitrogen-and-oxygen-generators-psa-vpsa-and-membrane.html)

**How pure is PSA nitrogen?**

From about 95 % to 99.999 % nitrogen, depending on design and flow.

From the lesson [PSA, VPSA and membrane gas generators explained](https://www.sarveshanalytics.com/learn/nitrogen-and-oxygen-generators-psa-vpsa-and-membrane.html)

**How pure is PSA oxygen?**

Typically 90 to 95 %.

From the lesson [PSA, VPSA and membrane gas generators explained](https://www.sarveshanalytics.com/learn/nitrogen-and-oxygen-generators-psa-vpsa-and-membrane.html)

**Do I need an analyser on a generator?**

If purity matters, yes. It shows when output drifts out of specification.

From the lesson [PSA, VPSA and membrane gas generators explained](https://www.sarveshanalytics.com/learn/nitrogen-and-oxygen-generators-psa-vpsa-and-membrane.html)

**What is carbon potential?**

The carbon content steel would reach at equilibrium with the furnace gas.

From the lesson [Heat treatment furnace atmospheres explained](https://www.sarveshanalytics.com/learn/heat-treatment-furnace-atmospheres-explained.html)

**What analyser measures hydrogen in nitrogen?**

ACCURRA-TCD, by thermal conductivity.

From the lesson [Heat treatment furnace atmospheres explained](https://www.sarveshanalytics.com/learn/heat-treatment-furnace-atmospheres-explained.html)

**How do you protect against air leaks in a furnace?**

Measure trace oxygen and alarm on a rise; maintain seals and positive pressure.

From the lesson [Heat treatment furnace atmospheres explained](https://www.sarveshanalytics.com/learn/heat-treatment-furnace-atmospheres-explained.html)

**What does the dew point tell you?**

How much water is in the atmosphere, which affects oxidising or reducing behaviour.

From the lesson [Heat treatment furnace atmospheres explained](https://www.sarveshanalytics.com/learn/heat-treatment-furnace-atmospheres-explained.html)

**What gases do I measure around an electrolyser?**

Hydrogen in oxygen, oxygen in hydrogen, hydrogen purity and moisture.

From the lesson [Hydrogen electrolysers and gas analysis: what to measure](https://www.sarveshanalytics.com/learn/hydrogen-electrolysers-and-gas-analysis.html)

**What alarm level for hydrogen in oxygen?**

Set by the manufacturer and your safety study, far below 4 %.

From the lesson [Hydrogen electrolysers and gas analysis: what to measure](https://www.sarveshanalytics.com/learn/hydrogen-electrolysers-and-gas-analysis.html)

**Are ACCURRA-Hz analysers suitable for electrolyser areas?**

They are certified for Gas Group IIC, Zone 1 and 2, T6; confirm your zone and requirements.

From the lesson [Hydrogen electrolysers and gas analysis: what to measure](https://www.sarveshanalytics.com/learn/hydrogen-electrolysers-and-gas-analysis.html)

**Do your analysers have SIL?**

We make no SIL claim.

From the lesson [Hydrogen electrolysers and gas analysis: what to measure](https://www.sarveshanalytics.com/learn/hydrogen-electrolysers-and-gas-analysis.html)

**What is biogas made of?**

Mostly methane and CO2, with water vapour and traces of H2S.

From the lesson [Biogas, landfill gas and digesters explained](https://www.sarveshanalytics.com/learn/biogas-landfill-gas-and-digesters-explained.html)

**What is CBG?**

Compressed biogas: biogas upgraded to high methane content and compressed for use as fuel.

From the lesson [Biogas, landfill gas and digesters explained](https://www.sarveshanalytics.com/learn/biogas-landfill-gas-and-digesters-explained.html)

**Why is H2S a problem?**

It is corrosive and toxic and damages sensors and engines.

From the lesson [Biogas, landfill gas and digesters explained](https://www.sarveshanalytics.com/learn/biogas-landfill-gas-and-digesters-explained.html)

**Which analyser measures biogas methane?**

ACCURRA-S 4030, by NDIR, on a conditioned sample.

From the lesson [Biogas, landfill gas and digesters explained](https://www.sarveshanalytics.com/learn/biogas-landfill-gas-and-digesters-explained.html)

**What oxygen level is used in ULO storage?**

Around 1 to 2 % or lower, depending on the fruit; follow your guideline.

From the lesson [Controlled-Atmosphere fruit storage explained](https://www.sarveshanalytics.com/learn/controlled-atmosphere-fruit-storage-explained.html)

**Which analyser measures CA store gas?**

ACCURRA-S 3030 for oxygen and 6030 for CO2; confirm range and accuracy for ULO.

From the lesson [Controlled-Atmosphere fruit storage explained](https://www.sarveshanalytics.com/learn/controlled-atmosphere-fruit-storage-explained.html)

**Can one analyser serve several rooms?**

With a sampling manifold that switches rooms in turn; discuss the arrangement.

From the lesson [Controlled-Atmosphere fruit storage explained](https://www.sarveshanalytics.com/learn/controlled-atmosphere-fruit-storage-explained.html)

**Why measure CO2 as well as oxygen?**

Because the combination sets the preservation, and the ratio shows respiration.

From the lesson [Controlled-Atmosphere fruit storage explained](https://www.sarveshanalytics.com/learn/controlled-atmosphere-fruit-storage-explained.html)

**Where in a cement plant is oxygen measured?**

At the kiln inlet, the precalciner and preheater outlets, and at the stack.

From the lesson [Cement kiln gas analysis explained: O2, CO and safety](https://www.sarveshanalytics.com/learn/cement-kiln-gas-analysis-explained.html)

**Can your analyser trip an ESP on CO?**

We make no SIL claim and need your response and integrity requirements first.

From the lesson [Cement kiln gas analysis explained: O2, CO and safety](https://www.sarveshanalytics.com/learn/cement-kiln-gas-analysis-explained.html)

**Can the probe survive kiln dust?**

With the right probe material, filter and blowback, yes; tell us the dust load and temperature.

From the lesson [Cement kiln gas analysis explained: O2, CO and safety](https://www.sarveshanalytics.com/learn/cement-kiln-gas-analysis-explained.html)

**Do you supply cement plants?**

JK Cement is among our customers.

From the lesson [Cement kiln gas analysis explained: O2, CO and safety](https://www.sarveshanalytics.com/learn/cement-kiln-gas-analysis-explained.html)

**What is in blast furnace gas?**

Mostly nitrogen, CO and CO2 with a little hydrogen, and heavy dust.

From the lesson [Steel plant gases: coke oven, blast furnace, converter](https://www.sarveshanalytics.com/learn/steel-plant-gases-explained.html)

**Why is converter gas recovered only sometimes?**

Its composition changes during the blow; it is recovered when CO is high enough and oxygen is low.

From the lesson [Steel plant gases: coke oven, blast furnace, converter](https://www.sarveshanalytics.com/learn/steel-plant-gases-explained.html)

**Can you analyse coke oven gas?**

Send the composition; tar and ammonia need review and a suitable sample system.

From the lesson [Steel plant gases: coke oven, blast furnace, converter](https://www.sarveshanalytics.com/learn/steel-plant-gases-explained.html)

**Which analysers are used in steelworks?**

NDIR for CO and CO2, thermal conductivity for hydrogen in binary cases, and oxygen analysers, with probes for hot dusty gas.

From the lesson [Steel plant gases: coke oven, blast furnace, converter](https://www.sarveshanalytics.com/learn/steel-plant-gases-explained.html)

**What is a CEMS?**

A continuous emission monitoring system that measures stack gases and reports them corrected to regulatory conditions.

From the lesson [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html)

**Is SGI-320 approved by regulators?**

We make no statutory approval claim; send your regulatory requirement.

From the lesson [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html)

**What is the difference between hot-wet and cold-dry?**

Hot-wet keeps the sample hot and wet to the analyser; cold-dry dries it first.

From the lesson [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html)

**Do I need to measure oxygen?**

Yes, normally, for reference oxygen correction.

From the lesson [Emission monitoring (CEMS) basics: what a system does](https://www.sarveshanalytics.com/learn/emission-monitoring-cems-basics.html)

**What dew point is instrument air?**

Often class 2, minus 40 C pressure dew point or lower, depending on the application and climate.

From the lesson [Compressed air quality and dew point (ISO 8573-1)](https://www.sarveshanalytics.com/learn/compressed-air-quality-and-dew-point.html)

**What is the difference between refrigerant and desiccant dryers?**

Refrigerant dryers cool to about 3 C; desiccant dryers adsorb water to reach minus 40 C or lower.

From the lesson [Compressed air quality and dew point (ISO 8573-1)](https://www.sarveshanalytics.com/learn/compressed-air-quality-and-dew-point.html)

**Which analyser measures compressed air dew point?**

ACCURRA-S M015, minus 80 to plus 20 C.

From the lesson [Compressed air quality and dew point (ISO 8573-1)](https://www.sarveshanalytics.com/learn/compressed-air-quality-and-dew-point.html)

**Does ISO 8573-1 cover CO and CO2?**

Gaseous contaminants are handled in other parts of the standard; medical and breathing air have their own specifications.

From the lesson [Compressed air quality and dew point (ISO 8573-1)](https://www.sarveshanalytics.com/learn/compressed-air-quality-and-dew-point.html)

**What oxygen level is safe for blanketing?**

Below the LOC of your material by the margin your safety assessment requires.

From the lesson [Inert gas blanketing in pharma and chemical plants](https://www.sarveshanalytics.com/learn/inert-gas-blanketing-in-pharma-and-chemical-plants.html)

**What gas is used for blanketing?**

Usually nitrogen; argon or CO2 in some cases.

From the lesson [Inert gas blanketing in pharma and chemical plants](https://www.sarveshanalytics.com/learn/inert-gas-blanketing-in-pharma-and-chemical-plants.html)

**Does blanketing need continuous oxygen monitoring?**

It is common and often required; follow your safety assessment and standards.

From the lesson [Inert gas blanketing in pharma and chemical plants](https://www.sarveshanalytics.com/learn/inert-gas-blanketing-in-pharma-and-chemical-plants.html)

**Can ACCURRA-S be used in a solvent area?**

Not in a classified area; use ACCURRA-Hz or INFINITY-Tx.

From the lesson [Inert gas blanketing in pharma and chemical plants](https://www.sarveshanalytics.com/learn/inert-gas-blanketing-in-pharma-and-chemical-plants.html)

**What oxygen level does a glove box hold?**

From a few ppm to below 1 ppm, depending on the box and use.

From the lesson [Glove boxes and Controlled-Atmosphere chambers explained](https://www.sarveshanalytics.com/learn/glove-boxes-and-controlled-atmosphere-chambers.html)

**Can ACCURRA-S 3015 measure 0.1 ppm?**

Its minimum range is 0 to 100 ppm; sub-ppm needs other sensors. Tell us your target.

From the lesson [Glove boxes and Controlled-Atmosphere chambers explained](https://www.sarveshanalytics.com/learn/glove-boxes-and-controlled-atmosphere-chambers.html)

**How do I avoid draining the box atmosphere?**

Return the sample or limit the flow.

From the lesson [Glove boxes and Controlled-Atmosphere chambers explained](https://www.sarveshanalytics.com/learn/glove-boxes-and-controlled-atmosphere-chambers.html)

**Can one analyser read oxygen and moisture?**

A dual-channel ACCURRA-S can read both from one sample.

From the lesson [Glove boxes and Controlled-Atmosphere chambers explained](https://www.sarveshanalytics.com/learn/glove-boxes-and-controlled-atmosphere-chambers.html)

## Choosing and buying

**What should a gas analyser specification include?**

Purpose, measurement, sample, tapping, site, utilities, outputs, calibration, standards and support.

From the lesson [How to specify a gas analyser: A practical checklist](https://www.sarveshanalytics.com/learn/how-to-specify-a-gas-analyser.html)

**Do I need to know the exact accuracy?**

State what the process needs; see the accuracy lesson.

From the lesson [How to specify a gas analyser: A practical checklist](https://www.sarveshanalytics.com/learn/how-to-specify-a-gas-analyser.html)

**Can I send a partial specification?**

Yes. Send what you have and we will ask for the rest.

From the lesson [How to specify a gas analyser: A practical checklist](https://www.sarveshanalytics.com/learn/how-to-specify-a-gas-analyser.html)

**What if the sample composition is unknown?**

Say so, send typical analyses, and we will propose how to confirm it.

From the lesson [How to specify a gas analyser: A practical checklist](https://www.sarveshanalytics.com/learn/how-to-specify-a-gas-analyser.html)

**What accuracy do I need for oxygen trim?**

Usually a few tenths of a percent of oxygen is adequate; control needs repeatability more than absolute accuracy.

From the lesson [What accuracy do you really need from a gas analyser?](https://www.sarveshanalytics.com/learn/what-accuracy-do-you-really-need.html)

**Is plus or minus 1 % of span good enough?**

It depends on the tolerance and where on the range the process runs.

From the lesson [What accuracy do you really need from a gas analyser?](https://www.sarveshanalytics.com/learn/what-accuracy-do-you-really-need.html)

**How does calibration gas uncertainty affect accuracy?**

It sets a floor: the reading cannot be known better than the reference.

From the lesson [What accuracy do you really need from a gas analyser?](https://www.sarveshanalytics.com/learn/what-accuracy-do-you-really-need.html)

**Should I pay for the best accuracy?**

Only if the decision needs it and the sample and calibration can support it.

From the lesson [What accuracy do you really need from a gas analyser?](https://www.sarveshanalytics.com/learn/what-accuracy-do-you-really-need.html)

**How much does an online gas analyser cost to run?**

It depends on the sample and the analyser; count calibration gas, filters, sensors and labour.

From the lesson [Total cost of ownership of a gas analyser](https://www.sarveshanalytics.com/learn/total-cost-of-ownership-of-a-gas-analyser.html)

**Is the cheapest quote the best?**

Rarely. Compare scope and lifetime cost.

From the lesson [Total cost of ownership of a gas analyser](https://www.sarveshanalytics.com/learn/total-cost-of-ownership-of-a-gas-analyser.html)

**How do I lower running costs?**

Good sampling, a routine, auto-calibration and spares.

From the lesson [Total cost of ownership of a gas analyser](https://www.sarveshanalytics.com/learn/total-cost-of-ownership-of-a-gas-analyser.html)

**Should I include downtime in the comparison?**

Yes, for critical measurements.

From the lesson [Total cost of ownership of a gas analyser](https://www.sarveshanalytics.com/learn/total-cost-of-ownership-of-a-gas-analyser.html)

**How do I compare gas analyser quotes?**

On the same scope and basis, with exclusions and running costs listed.

From the lesson [How to compare gas analyser quotes like for like](https://www.sarveshanalytics.com/learn/comparing-gas-analyser-quotes.html)

**What should every quote include?**

Scope, performance, certificates, documents, calibration, commissioning and support.

From the lesson [How to compare gas analyser quotes like for like](https://www.sarveshanalytics.com/learn/comparing-gas-analyser-quotes.html)

**Is a lower price a warning sign?**

Check what is missing.

From the lesson [How to compare gas analyser quotes like for like](https://www.sarveshanalytics.com/learn/comparing-gas-analyser-quotes.html)

**Can you quote against my specification?**

Yes. Send it with the application data sheet.

From the lesson [How to compare gas analyser quotes like for like](https://www.sarveshanalytics.com/learn/comparing-gas-analyser-quotes.html)

**What should I ask an analyser supplier?**

About experience on similar streams, sample limits, certification, calibration, spares and service.

From the lesson [Questions to ask a gas analyser supplier before you buy](https://www.sarveshanalytics.com/learn/questions-to-ask-a-gas-analyser-supplier.html)

**How do I judge whether a supplier's claims hold?**

Ask what they cannot do, and check that the answer is specific.

From the lesson [Questions to ask a gas analyser supplier before you buy](https://www.sarveshanalytics.com/learn/questions-to-ask-a-gas-analyser-supplier.html)

**Do you provide training?**

Ask us about training and service for your site.

From the lesson [Questions to ask a gas analyser supplier before you buy](https://www.sarveshanalytics.com/learn/questions-to-ask-a-gas-analyser-supplier.html)

**Who do I call for support?**

Sales +91 94 2300 4179 and support +91 94 2300 4178.

From the lesson [Questions to ask a gas analyser supplier before you buy](https://www.sarveshanalytics.com/learn/questions-to-ask-a-gas-analyser-supplier.html)

**What is the application data sheet?**

An online form for the pre-order design inputs we need to recommend an analyser.

From the lesson [The application data sheet: what to send Sarvesh](https://www.sarveshanalytics.com/learn/the-application-data-sheet-and-what-to-send-us.html)

**Is it free?**

Yes.

From the lesson [The application data sheet: what to send Sarvesh](https://www.sarveshanalytics.com/learn/the-application-data-sheet-and-what-to-send-us.html)

**Can I attach files?**

Yes, up to a total size noted on the form.

From the lesson [The application data sheet: what to send Sarvesh](https://www.sarveshanalytics.com/learn/the-application-data-sheet-and-what-to-send-us.html)

**How soon will I hear back?**

An applications engineer will reply by email or phone; for urgent cases call sales on +91 94 2300 4179.

From the lesson [The application data sheet: what to send Sarvesh](https://www.sarveshanalytics.com/learn/the-application-data-sheet-and-what-to-send-us.html)

Your question is not here? [Ask an engineer](https://www.sarveshanalytics.com/contact.html#enquiry); we add the good ones.