# Sampling cryogenic and liquefied gases: why vaporise first

Analysers measure gas. A liquid draw from a tank or column has to become gas without changing composition on the way.

 Intermediate 3 min read Lesson 8 of 10[Sampling and conditioning](https://www.sarveshanalytics.com/learn/track-sampling.html)

Helpful first: [Sample pressure, flow and fast bypass loops](https://www.sarveshanalytics.com/learn/sample-pressure-flow-and-bypass-loops.html)

 Quick answer

A liquid sample must be vaporised completely before it reaches an analyser. If it boils off slowly, the lighter component leaves first and the sample no longer has the liquid's composition. A sample vaporiser heats the draw so that all of it becomes gas at once. For cryogenic oxygen, nitrogen and argon, use a vaporiser and keep the line short, dry and warm.

Updated 3 November 2026 · Figures from Sarvesh datasheets

**After this lesson you can**

- Explain fractionation when a liquid boils off partially
- Describe what a sample vaporiser does
- Plan a cryogenic sample line with a vaporiser and bypass

## The problem: boiling off in stages

A liquid mixture boils off its most volatile component first. If a liquid sample warms gradually in a line, the first vapour is rich in the light component and the residue is rich in the heavy one. An analyser fed this sample sees a composition that moves with time and does not equal the liquid. This is called fractionation.

## What a vaporiser does

A sample vaporiser heats the liquid so quickly and completely that every component turns to gas together. The gas then has the composition of the liquid. SV-200 is the cryogenic sample vaporiser in the Sarvesh range, with a bypass line for quick flushing; SV-400 serves wet samples and others. They are designed so that the sample is vaporised without loss: [sample-vaporisers](https://www.sarveshanalytics.com/sample-vaporisers.html).

## Building the sample path

1. Take the draw from the liquid line through a short insulated line to the vaporiser.
2. Vaporise at the vaporiser, then regulate pressure and flow of the gas.
3. Use the bypass to flush the vaporiser quickly at start-up and when changing sample.
4. Keep downstream tube warm and dry so the gas does not recondense or pick up moisture.
5. Vent safely: oxygen-enriched vent gas is a hazard: [Oxygen enrichment and oxygen deficiency hazards](https://www.sarveshanalytics.com/learn/oxygen-enrichment-and-oxygen-deficiency-hazards.html).

## Cryogenic safety

- Liquid oxygen and oxidising liquids need materials cleaned for oxygen service; hydrocarbon grease and oil are not allowed.
- Trapped liquid in a closed line can build dangerous pressure as it warms; include relief where the design requires.
- Cold lines and parts cause burns.

**Rule of thumb**

Treat any liquid draw as a safety design item as well as an analysis item. Ask an engineer to review the arrangement.

A vaporiser turns the whole liquid draw to gas at once, so the composition reaching the analyser is the composition of the liquid.

**In short**

- Liquid samples fractionate if they boil off slowly.
- A vaporiser turns the whole draw to gas at once.
- A bypass line flushes the vaporiser quickly.
- Cryogenic and oxygen service raise safety requirements.

## Check your understanding

Answer in your head or on paper first, then open the answer. Recalling an answer teaches more than rereading.

**What is fractionation?**

The lighter component of a liquid mixture boiling off first, so the vapour no longer has the liquid's composition.

**What is the bypass on a cryogenic vaporiser for?**

To flush the vaporiser and line quickly at start-up or when changing sample.

**Why must oxygen service use cleaned components?**

Oil and grease can ignite in oxygen-enriched conditions.

## Key terms

- [Vaporiser](https://www.sarveshanalytics.com/learn/glossary.html#t-vaporiser)
- [Cryogenic](https://www.sarveshanalytics.com/learn/glossary.html#t-cryogenic)
- [Fractionation](https://www.sarveshanalytics.com/learn/glossary.html#t-fractionation)

## Frequently asked questions

**Why can't I analyse liquid oxygen directly?**

Analysers measure gas. The liquid must be vaporised, and completely, to keep its composition.

**Which vaporiser does Sarvesh offer for cryogenic samples?**

SV-200, with a bypass line; SV-400 is also available.

**Can I analyse LPG with ACCURRA-S?**

Flammable samples are outside the standard ACCURRA-S sample specification; ask an engineer.

**How do I flush a vaporiser?**

Use the bypass line, per the product instructions.

## Related products

[

### SV-200 / SV-400 vaporisers

Sample vaporisers for cryogenic and wet samples with no loss of sample. View SV-200 / SV-400 vaporisers

](https://www.sarveshanalytics.com/sample-vaporisers.html)

## Where this applies

- [Oxygen purity in air separation plants](https://www.sarveshanalytics.com/applications/oxygen-purity-air-separation-plant.html)
- [Sample vaporisers SV-200 and SV-400](https://www.sarveshanalytics.com/sample-vaporisers.html)