title: ACCURRA-S 2030 Percent Carbon Monoxide Analyser | Sarvesh
url: https://www.sarveshanalytics.com/models/accurra-s-2030-percent-carbon-monoxide-analyser.html
description: ACCURRA-S 2030 measures carbon monoxide in percent from 0 to 100 %; minimum range 0 to 5 %, with a ndir sensor: accuracy 2 % of FS, response below 30.
updated: 2026-11-23
source: html
---

ACCURRA-S · Model 2030

# ACCURRA-S 2030 percent carbon monoxide analyser

ACCURRA-S 2030 measures carbon monoxide in percent over 0 to 100 % (minimum range 0 to 5 %) with a ndir sensor. It is a safe-area, 19-inch rack or panel analyser with an isolated 4–20 mA output, relays and Modbus, made by Sarvesh in Pune, India.

 Quick answer

ACCURRA-S 2030 measures carbon monoxide in percent from 0 to 100 % (minimum range 0 to 5 %), with a ndir sensor: accuracy 2 % of FS, response below 30 s. Ordering code ACCURRA-S-2030 followed by the second component and options.

Figures from the ACCURRA-S datasheet (SAPL/PDS/ACCURRA-S-10/25-Aug)

## Specifications

| Item | Specification | Notes |
|---|---|---|
| Measures | Carbon monoxide |  |
| Range | 0 to 100 % | Configurable within this span; dual-range 4–20 mA output |
| Minimum range | 0 to 5 % |  |
| Sensor | NDIR |  |
| Display resolution | 0.01 % |  |
| Intrinsic error (accuracy) | 2 % of FS |  |
| Repeatability | 2 % of reading |  |
| Response T90 | Below 30 s | At 1 LPM unless stated |
| Notes | Sensor life above 5 years |  |

From datasheet SAPL/PDS/ACCURRA-S-10/25-Aug. Specifications may change as products improve.

## Built into every ACCURRA-S

| Item | Specification | Notes |
|---|---|---|
| Display | Backlit LCD, 4 lines x 20 characters |  |
| Analog output | 1 x 4–20 mA isolated, max load 500 ohm, dual range | Range freely selectable over the full measurement range |
| Relays | 6 x SPST rated 1 A 230 V AC, all failsafe | Alarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves |
| Communication | Modbus RTU over RS 485 | Read measured value, status, set-up and history; write auto-calibration start and set-up |
| Enclosure | 19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mm | IP 20, safe area only; net about 6 kg, gross about 8 kg |
| Power | 100 to 240 V AC, 50/60 Hz, 45 W |  |
| Inlet and outlet | 1/4 inch OD SS316 |  |
| Ambient | 5 to 40 °C; below 90 % RH non-condensing | Storage 0 to 50 °C |

## Options for this model

- Auto calibration: available, with internal solenoid valves or as a relay output for external valves.
- Sample pump: an internal sample pump is available on a single measurement, or a relay output drives an external pump.
- Range ID relay: shows which of the two 4–20 mA ranges is active; it shares a relay with the sample pump, so choose one.

[Build the part number](https://www.sarveshanalytics.com/part-number-builder.html) [Write the specification](https://www.sarveshanalytics.com/analyser-selector.html)

## Dual measurement

The 2030 can share an ACCURRA-S enclosure with a second sensor in series, each with its own electronics, 4–20 mA output, relays and Modbus address. The catalogue lists it with 3030 (oxygen, percentage), 3130 (oxygen, percentage) and 6030 (carbon dioxide). Other pairs are possible on request.

## For hazardous areas

The same 2030 sensor is available in ACCURRA-Hz, in a certified enclosure (Gas Group IIC, Zone 1 and 2, T6), wall mounted, with 24 V DC or AC power. Calibration is done without opening the covers. See [ACCURRA-Hz](https://www.sarveshanalytics.com/hazardous-gas-analysers.html).

## Sample at the analyser

- The sample gas should be non-corrosive, non-toxic, non-flammable, non-condensing and dry, and free from entrained oil.
- Sample pressure 2 to 10 psig maximum, regulated externally. Recommended flow 1 LPM.
- Dust or particulate below 3 micron; the built-in fine filter handles light loads, and heavier dust needs an upstream filter.
- Materials in contact with sample: SS316, Viton, PTFE, glass, aluminium and acrylic.
- Sample temperature 5 to 45 °C, and a sample dew point at least 5 °C below the lowest ambient temperature.

## Where it is used

Application notes where 2030 is part of the recommended system:

- [Gas purity certification and cylinder filling](https://www.sarveshanalytics.com/applications/gas-purity-certification-and-cylinder-filling.html)
- [Heat-treatment furnace atmosphere](https://www.sarveshanalytics.com/applications/heat-treatment-furnace-atmosphere.html)
- [Steel plant process gases](https://www.sarveshanalytics.com/applications/steel-plant-process-gas-analysis.html)
- [Pyrolysis, gasification and syngas](https://www.sarveshanalytics.com/applications/pyrolysis-gasification-and-syngas.html)
- [Optical fibre and cable manufacturing](https://www.sarveshanalytics.com/applications/optical-fibre-and-cable-manufacturing.html)
- [Carburising and endothermic gas carbon potential](https://www.sarveshanalytics.com/applications/carburising-and-endothermic-gas-carbon-potential.html)
- [Foundry cupola and melting furnace gas](https://www.sarveshanalytics.com/applications/foundry-cupola-and-melting-furnace-gas.html)
- [Cement kiln and precalciner O2 and CO](https://www.sarveshanalytics.com/applications/cement-kiln-and-precalciner-o2-and-co.html)
- [Coal mill inerting and fire protection](https://www.sarveshanalytics.com/applications/coal-mill-inerting-and-fire-protection.html)
- [Boiler and burner factory test benches](https://www.sarveshanalytics.com/applications/boiler-and-burner-factory-test-benches.html)
- [Shift, CO2 removal and methanator: CO and CO2](https://www.sarveshanalytics.com/applications/ammonia-plant-shift-co2-removal-and-methanator-co-and-co2.html)

## Frequently asked questions

**What range does ACCURRA-S 2030 measure?**

0 to 100 %, with a minimum range of 0 to 5 %.

**Does 2030 have auto calibration and an internal sample pump?**

Auto calibration: available, with internal solenoid valves or as a relay output for external valves. Sample pump: an internal sample pump is available on a single measurement, or a relay output drives an external pump.

**What is the part number?**

It starts ACCURRA-S-2030. Positions 5 to 8 are a second gas or XXXX, then the power supply, the range ID or pump relay, and auto calibration. The part number builder writes the full code and checks it against the catalogue.

**Is 2030 available for hazardous areas?**

Yes, in ACCURRA-Hz, in a certified enclosure for Gas Group IIC, Zone 1 and 2, T6.

## Other models

- [ACCURRA-S 3015 trace oxygen analyser](https://www.sarveshanalytics.com/models/accurra-s-3015-trace-oxygen-analyser.html)
- [ACCURRA-S 3030 percent oxygen analyser](https://www.sarveshanalytics.com/models/accurra-s-3030-percent-oxygen-analyser.html)
- [ACCURRA-S 3090 oxygen purity analyser](https://www.sarveshanalytics.com/models/accurra-s-3090-oxygen-purity-analyser.html)
- [ACCURRA-S 3130 paramagnetic oxygen analyser](https://www.sarveshanalytics.com/models/accurra-s-3130-paramagnetic-oxygen-analyser.html)
- [ACCURRA-S 3190 paramagnetic oxygen purity analyser](https://www.sarveshanalytics.com/models/accurra-s-3190-paramagnetic-oxygen-purity-analyser.html)
- [ACCURRA-S 2015 ppm carbon monoxide analyser](https://www.sarveshanalytics.com/models/accurra-s-2015-ppm-carbon-monoxide-analyser.html)
- [ACCURRA-S 6015 ppm carbon dioxide analyser](https://www.sarveshanalytics.com/models/accurra-s-6015-ppm-carbon-dioxide-analyser.html)
- [ACCURRA-S 6030 percent carbon dioxide analyser](https://www.sarveshanalytics.com/models/accurra-s-6030-percent-carbon-dioxide-analyser.html)
- [ACCURRA-S 4030 methane analyser](https://www.sarveshanalytics.com/models/accurra-s-4030-methane-analyser.html)
- [ACCURRA-S M015 dew point analyser](https://www.sarveshanalytics.com/models/accurra-s-m015-dew-point-analyser.html)
- [**ACCURRA-S carbon monoxide analysers**](https://www.sarveshanalytics.com/carbon-monoxide-gas-analyser.html)
- [How ACCURRA model numbers work](https://www.sarveshanalytics.com/accurra-model-numbers.html)