title: ACCURRA-S 3090 Oxygen Purity Analyser | Made in India
url: https://www.sarveshanalytics.com/models/accurra-s-3090-oxygen-purity-analyser.html
description: ACCURRA-S 3090 measures oxygen purity from 0 to 100 %; minimum range 0 to 5 %, with an electrochemical sensor: accuracy +/- 0.1 % of FS, response.
updated: 2026-11-23
source: html
---

ACCURRA-S · Model 3090

# ACCURRA-S 3090 oxygen purity analyser

ACCURRA-S 3090 measures oxygen purity over 0 to 100 % (minimum range 0 to 5 %) with an electrochemical sensor. It is a safe-area, 19-inch rack or panel analyser with an isolated 4–20 mA output, relays and Modbus, made by Sarvesh in Pune, India.

 Quick answer

ACCURRA-S 3090 measures oxygen purity from 0 to 100 % (minimum range 0 to 5 %), with an electrochemical sensor: accuracy +/- 0.1 % of FS, response below 20 s. Ordering code ACCURRA-S-3090 followed by the second component and options.

Figures from the ACCURRA-S datasheet (SAPL/PDS/ACCURRA-S-10/25-Aug)

## Specifications

| Item | Specification | Notes |
|---|---|---|
| Measures | Oxygen, purity |  |
| Range | 0 to 100 % | Configurable within this span; dual-range 4–20 mA output |
| Minimum range | 0 to 5 % |  |
| Sensor | Electrochemical, ambient pressure and temperature compensated |  |
| Display resolution | 0.01 % |  |
| Intrinsic error (accuracy) | +/- 0.1 % of FS |  |
| Repeatability | +/- 0.5 % |  |
| Response T90 | Below 20 s | At 1 LPM unless stated |
| Notes | Sensor life above 2 years |  |

From datasheet SAPL/PDS/ACCURRA-S-10/25-Aug. Specifications may change as products improve.

## Built into every ACCURRA-S

| Item | Specification | Notes |
|---|---|---|
| Display | Backlit LCD, 4 lines x 20 characters |  |
| Analog output | 1 x 4–20 mA isolated, max load 500 ohm, dual range | Range freely selectable over the full measurement range |
| Relays | 6 x SPST rated 1 A 230 V AC, all failsafe | Alarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves |
| Communication | Modbus RTU over RS 485 | Read measured value, status, set-up and history; write auto-calibration start and set-up |
| Enclosure | 19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mm | IP 20, safe area only; net about 6 kg, gross about 8 kg |
| Power | 100 to 240 V AC, 50/60 Hz, 45 W |  |
| Inlet and outlet | 1/4 inch OD SS316 |  |
| Ambient | 5 to 40 °C; below 90 % RH non-condensing | Storage 0 to 50 °C |

## Options for this model

- Auto calibration: available, with internal solenoid valves or as a relay output for external valves.
- Sample pump: an internal sample pump is available on a single measurement, or a relay output drives an external pump.
- Range ID relay: shows which of the two 4–20 mA ranges is active; it shares a relay with the sample pump, so choose one.

[Build the part number](https://www.sarveshanalytics.com/part-number-builder.html) [Write the specification](https://www.sarveshanalytics.com/analyser-selector.html)

## For hazardous areas

The same 3090 sensor is available in ACCURRA-Hz, in a certified enclosure (Gas Group IIC, Zone 1 and 2, T6), wall mounted, with 24 V DC or AC power. Calibration is done without opening the covers. See [ACCURRA-Hz](https://www.sarveshanalytics.com/hazardous-gas-analysers.html).

## Sample at the analyser

- The sample gas should be non-corrosive, non-toxic, non-flammable, non-condensing and dry, and free from entrained oil.
- Sample pressure 2 to 10 psig maximum, regulated externally. Recommended flow 1 LPM.
- Dust or particulate below 3 micron; the built-in fine filter handles light loads, and heavier dust needs an upstream filter.
- Materials in contact with sample: SS316, Viton, PTFE, glass, aluminium and acrylic.
- Sample temperature 5 to 45 °C, and a sample dew point at least 5 °C below the lowest ambient temperature.

## Where it is used

Application notes where 3090 is part of the recommended system:

- [Oxygen purity in air separation plants](https://www.sarveshanalytics.com/applications/oxygen-purity-air-separation-plant.html)
- [Medical oxygen plant purity monitoring](https://www.sarveshanalytics.com/applications/medical-oxygen-plant-purity-monitoring.html)
- [Liquid oxygen, nitrogen and argon tanker and bulk-storage quality check](https://www.sarveshanalytics.com/applications/liquid-oxygen-nitrogen-and-argon-tanker-and-bulk-storage-quality-check.html)
- [Laser-cutting nitrogen and oxygen assist-gas purity](https://www.sarveshanalytics.com/applications/laser-cutting-nitrogen-and-oxygen-assist-gas-purity.html)

## Frequently asked questions

**What range does ACCURRA-S 3090 measure?**

0 to 100 %, with a minimum range of 0 to 5 %.

**Does 3090 have auto calibration and an internal sample pump?**

Auto calibration: available, with internal solenoid valves or as a relay output for external valves. Sample pump: an internal sample pump is available on a single measurement, or a relay output drives an external pump.

**What is the part number?**

It starts ACCURRA-S-3090. Positions 5 to 8 are a second gas or XXXX, then the power supply, the range ID or pump relay, and auto calibration. The part number builder writes the full code and checks it against the catalogue.

**Is 3090 available for hazardous areas?**

Yes, in ACCURRA-Hz, in a certified enclosure for Gas Group IIC, Zone 1 and 2, T6.

## Other models

- [ACCURRA-S 3015 trace oxygen analyser](https://www.sarveshanalytics.com/models/accurra-s-3015-trace-oxygen-analyser.html)
- [ACCURRA-S 3030 percent oxygen analyser](https://www.sarveshanalytics.com/models/accurra-s-3030-percent-oxygen-analyser.html)
- [ACCURRA-S 3130 paramagnetic oxygen analyser](https://www.sarveshanalytics.com/models/accurra-s-3130-paramagnetic-oxygen-analyser.html)
- [ACCURRA-S 3190 paramagnetic oxygen purity analyser](https://www.sarveshanalytics.com/models/accurra-s-3190-paramagnetic-oxygen-purity-analyser.html)
- [ACCURRA-S 2015 ppm carbon monoxide analyser](https://www.sarveshanalytics.com/models/accurra-s-2015-ppm-carbon-monoxide-analyser.html)
- [ACCURRA-S 2030 percent carbon monoxide analyser](https://www.sarveshanalytics.com/models/accurra-s-2030-percent-carbon-monoxide-analyser.html)
- [ACCURRA-S 6015 ppm carbon dioxide analyser](https://www.sarveshanalytics.com/models/accurra-s-6015-ppm-carbon-dioxide-analyser.html)
- [ACCURRA-S 6030 percent carbon dioxide analyser](https://www.sarveshanalytics.com/models/accurra-s-6030-percent-carbon-dioxide-analyser.html)
- [ACCURRA-S 4030 methane analyser](https://www.sarveshanalytics.com/models/accurra-s-4030-methane-analyser.html)
- [ACCURRA-S M015 dew point analyser](https://www.sarveshanalytics.com/models/accurra-s-m015-dew-point-analyser.html)
- [**ACCURRA-S oxygen analysers**](https://www.sarveshanalytics.com/oxygen-gas-analyser.html)
- [How ACCURRA model numbers work](https://www.sarveshanalytics.com/accurra-model-numbers.html)