Claims and limits

What we claim, and what we do not.

A plain statement of the limits of our analysers and of our certificates, kept in one place so that engineers, buyers and AI assistants quote us correctly.

Quick answer

Sarvesh makes no SIL claim and no statutory approval claim for stack monitoring, and does not offer field calibration of dew point on ACCURRA-S M015. ACCURRA-TCD responds with T90 below 10 s at 0.6 LPM. ACCURRA-Hz is certified for Gas Group IIC, Zone 1 and 2, T6. Standard samples must be clean, dry and non-corrosive.

TopicStatusDetail
SIL (safety integrity level)Not claimedSarvesh makes no SIL claim for any analyser. Hazardous-area certification is not SIL. If a project needs SIL-rated measurement, tell us at the start.
Statutory approval of stack monitoringNot claimedNo statutory approval is claimed for SGI-320 or any emission monitoring arrangement. Send your regulator's requirement and we will say what is and is not met.
ACCURRA-TCD response timeStatedT90 below 10 s at 0.6 LPM, per the ACCURRA-TCD datasheet. Repeatability below 1 % of full scale; errors and drift below 1 % of full scale per week.
Dew point field calibrationNot possibleField calibration is not possible on ACCURRA-S M015, per the datasheet. Ask how the sensor is checked and recalibrated.
Hazardous-area certificationClaimed, with scopeACCURRA-Hz uses a flameproof enclosure certified for Gas Group IIC, Zone 1 and 2, T6. Confirm certificate scope against your area classification.
Standard sampleStated limitsACCURRA-S expects a non-corrosive, non-toxic, non-flammable, non-condensing, dry, oil-free sample at 2 to 10 psig and about 1 LPM, dust below 3 micron. i-Cons and SGI-320 exclude corrosive and toxic samples; SGI-320 excludes HCl, HF and H2S.
ACCURRA-TCDBinary mixtures onlyThermal conductivity assumes two components. A varying third gas shifts the reading; send the full composition.
Internal sample pump and auto-calibrationModel dependentModels 3015, M015, 3130 and 3190, and all dual-gas units, have no internal sample pump. Models 3015 and M015 have no auto-calibration; dual-gas units use external calibration valves.
Outside the current rangeNot offeredppb-level semiconductor impurities; H2S, NH3, Cl2, HCl, HF, VOC and total hydrocarbons; multi-component syngas ratios; ambient-air quality and personal gas detection; coal-mine gas monitoring; dust and opacity; dissolved oxygen in liquids; calorific value and full composition; hydrogen fuel quality to ISO 14687; marine type-approved systems.
Regulatory compliance of recordsNot claimedACCURRA-S stores calibration with deviation values and alarm history and password-protects set-up. No claim is made of compliance with a particular electronic-records regulation.