The complete measurement path.

Analysers, sample conditioning, probes and filtration, designed to work together and supported by the people who built them.

ACCURRA-S analyser
Safe area · Online

ACCURRA-S

The analyser we introduced in 2007 and keep improving. Trace, percentage and purity measurement of O₂, CO, CO₂ and CH₄, with moisture and dew point options. The dual-gas version measures two gases from one sample.

GasRangeSensor
Oxygen0–100 / 1000 ppm
0–25 / 0–100 %
95–100 %
Electrochemical, paramagnetic
Carbon dioxide0–5000 ppm
0–20 / 0–100 %
NDIR
Carbon monoxide0–1000 ppm
0–100 %
Electrochemical, NDIR
Methane0–100 %NDIR
Moisture−80 to +20 °C dpCeramic
  • Isolated dual-range 4–20 mA output with range ID relay, user-configurable over the full range
  • Modbus RTU over RS 485; alarm set points as Lo–LoLo, Lo–Hi or Hi–HiHi
  • Calibration check, calibration and alarm history, output hold during calibration
  • Dual-level password protection; optional cyclic and remote auto-calibration; optional internal sample pump
Full specifications, applications and FAQs →
Thermal conductivity cellA sample gas flows through a cell containing a heated platinum element. The rate at which the gas carries heat away is the measurement signal.Platinum element, held at constant temperatureSample inSample outHeat carried away by the gas sets the signal
Safe area · Binary gas

ACCURRA-TCD

A thermal conductivity analyser for binary gas mixtures, with a proprietary platinum sensor that needs no sealed reference and no reference gas flow.

MeasuresH₂, He, Ar, N₂, N₂O, Ne, Kr
Response T90< 10 s at 0.6 LPM
Repeatability, linearity< 1 %
Error and drift< 1 % of FS per week
Full specifications, applications and FAQs →
ACCURRA-S 3090 oxygen purity analyser
Safe area · Purity

ACCURRA-S 3090

Oxygen purity analyser using electrochemical or paramagnetic sensing, with ambient pressure and temperature compensation.

Repeatability< 0.5 %
Response T90< 20 s at 1 LPM
Error and drift< 0.5 %
Full specifications, applications and FAQs →
ACCURRA-Zr flue gas analyser
Safe area · Flue gas

ACCURRA-Zr

Zirconia-based oxygen measurement, 0.1 to 21 %, for furnaces and boilers. The sensor sits outside the duct, protected from dust, flow and temperature swings.

  • Response T90 below 15 s; repeatability better than 1 % of FS
  • Single-point calibration with plain air
  • Built-in automatic blowback
  • Option for high dust load samples
  • Option for temperatures up to 1400 °C
Full specifications, applications and FAQs →
ACCURRA-Hz flameproof analyser
Hazardous area · Online

ACCURRA-Hz

Process gas analysis in classified areas.

  • O₂ from ppm to purity; CO from ppm to percent; CO₂, CH₄ and H₂
  • Calibration without opening the covers
  • Purity O₂ with ambient pressure and temperature compensation
  • Enclosure certified for Gas Group IIC, Zone 1 and 2, T6, to IS/IEC 60079-1:2014; IP 66, wall mounted
  • Optional ACCURRA-CONNECT software operates the analyser remotely over Modbus
Full specifications, applications and FAQs →
INFINITY-Tx gas transmitter
Hazardous area · Transmitter

INFINITY-Tx

A loop-powered, two-wire gas transmitter for direct field installation.

  • Percentage oxygen and trace CO
  • Inbuilt filter and scrubber
  • Works with very low flow and low pressure samples
Full specifications, applications and FAQs →
i-Cons smart sample conditioner
Smart sample conditioner

i-Cons

Your sampling panel compressed into a compact 19-inch cabinet. A Peltier-based extractive system with the essential components built in and a controller that watches the whole sample path.

  • Peltier cooler with PTFE heat exchanger; peristaltic pump with variable rate
  • Built-in sample pump, condensate sensor, flow detector, auto-cal valves and fine filter
  • Controls probe blowback, sample valves and heated hose; auto blowback on flow failure with pulsating blowback for effective cleaning
  • Auto start-up sequence and auto shutdown on hardware failure
  • Fail-safe fault relay; alarm and fault history; maintenance reminders for pump tube, filters and pump kit
  • Modbus RTU over RS 485 for remote monitoring and remote calibration check
Full specifications, applications and FAQs →
SGI-300 sample cooler
Peltier sample cooler

SGI-300 and SGI-320

A compact Peltier cooler with solid heat exchanger in PTFE, an integrated peristaltic pump, filter and condensate detector. SGI-320 packages the approach as an emission monitoring system.

  • Relay output controls the sample pump from condensate and cooler temperature
  • Less tubing and wiring, faster integration, smaller panel footprint
Full specifications, applications and FAQs →
Analyser panel with i-Cons
Complete systems

Engineered analyser panels

The probe, heated line, i-Cons conditioner and ACCURRA analyser, built into one free-standing IP 54 panel and handed over with two sets of as-built drawings and manuals.

  • Bulkheads for sample, calibration, blowback, vent and drain; bottom cable entry
  • 1/4 inch PTFE internal tubing with SS316 fittings; provision for one 19 inch analyser
  • Exhaust fan or panel air conditioner, LED light, toughened-glass door
Full specifications, applications and FAQs →
Sample probe
Sample extraction

Sample probes SP-100 to SRP-400

ModelType
SP-100Simple probe, with or without in-situ filter at the tip
SP-200Probe with blowback, non-heated
SP-300Probe with heated filter
SRP-400Retractable probe
Probe materialSuitable up to
SS 316550 °C
Inconel 6251100 °C
Ceramic1400 °C
PTFE180 °C
  • Separate sample and blowback lines; 1/2 inch blowback line for efficient cleaning
  • Non-return valve in the sample line removes the need for a heated solenoid or air-operated valve
  • SRP-400: no welding on the sample path, separate sample and bleed ports, pressure indication and orifice-controlled safe bleed
  • Filter elements in SS 316, Hastelloy, Monel or PTFE; SP-300 on a 2 or 4 inch ANSI flange with a 10 micron sintered SS316 heated filter
  • SRP-400 rated to 100 kg/cm2, with a non fly-off design and NACE compliance available
Full specifications, applications and FAQs →
Sample vaporiserA liquid or cold sample passes through a heated block and leaves as gas with the same composition.Liquid sample inGas out to analyserHeatedblock
Sample vaporisers

Vaporisers for cryogenic and wet samples

SV-400 vaporises condensed sample on the go for accurate wet analysis with no loss of sample. SV-200 suits cryogenic gas measurement and avoids condensation in the sample line.

FeatureSV-200SV-400
Heater200 W, 230 V AC100 W, 110 or 230 V AC
FlowMax 10 LPMMax 6 LPM
ControlThermostat; Pt100 at inlet and outletBuilt-in controller, 100–200 °C

Catchpots, scrubbers, bubblers, safe drains and wash-water separators are also available. Ask us for details.

Full specifications, applications and FAQs →
Filter housing with elementSample enters the housing, passes through the filter element and leaves clean, while captured liquid drains from the bottom.Sample inClean sample outDrainHousing and element
Exclusive representative for India

Headline Filters, UK

Headline is the largest European maker of bonded microfibre filter elements and housings. We supply the full range for the Indian subcontinent.

  • Particulate, coalescing and PTFE membrane filtration to protect analysers, valves and regulators
  • Disposable inline filters and absorbers, emission filters, compressed air, gas, analyser, fast loop, high pressure, natural gas, adsorption and dehydration filters, Guardian membrane separators
  • Housings in SS316, Monel, Hastelloy, PTFE, PVDF, polypropylene, nylon and aluminium
  • Elements in SS316, PTFE, glass microfibre, Monel and Hastelloy; scrubber elements available
  • Up to 700 bar and 500 °C, depending on model
Full specifications, applications and FAQs →

The final responsibility to confirm that a product meets the process and analyser requirement rests with the customer. Specifications may change without notice as products improve.

Not sure which one fits?

Send us the gas, range, pressure, temperature and area classification.

Request a quote