Safe area analyser · Made in Pune since 2007

ACCURRA-S online gas analyser for O2, CO, CO2, CH4 and moisture

Trace, percent and purity measurement on one 19-inch platform, with each gas matched to the sensor that suits its range. Two gases can share one enclosure, each with its own electronics, output and Modbus ID.

ACCURRA-S
  • ISO 9001:2015, certified by TÜV
  • Designed and built in Pune since 2002
  • Supplied to Reliance, Tata Steel, Indian Oil, Linde and ISRO
Quick answer

ACCURRA-S is Sarvesh's safe-area online gas analyser for oxygen, CO, CO2, methane and moisture, from ppm to 100 %. It has a 4-20 mA output, failsafe relays and Modbus, and expects a dry, non-corrosive, non-toxic, non-flammable sample at 2 to 10 psig.

Updated 3 November 2026 · Figures from Sarvesh datasheets

Why ACCURRA-S

Accurate from ppm to purity

Trace oxygen to 1 % of reading ±2 ppm; purity oxygen compensated for ambient pressure and temperature, which keeps error below 0.5 %.

Integrates without add-ons

Dual-range isolated 4–20 mA with range ID relay, six failsafe relays, Modbus RTU and output freeze during calibration come as standard.

Maintenance you can plan

Time-stamped calibration, alarm and fault history shows drift and residual sensor life, so consumable sensors are replaced on schedule.

Specifications

ItemSpecificationNotes
DisplayBacklit LCD, 4 lines x 20 characters
Analog output1 x 4-20 mA isolated, max load 500 ohm, dual rangeRange freely selectable over the full measurement range
Relays6 x SPST rated 1 A 230 V AC, all failsafeAlarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves
CommunicationModbus RTU over RS 485Read measured value, status, set-up and history; write auto-calibration
Enclosure19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mmIP 20, safe area only; net about 6 kg, gross about 8 kg
Power100 to 240 V AC, 50/60 Hz, 45 W
Inlet and outlet1/4 inch OD SS316
Ambient5 to 40 C; below 90 % RH non-condensingStorage 0 to 50 C

From the current Sarvesh product datasheets and company profile. Specifications may change without notice as products improve. The final responsibility to check that a product meets the process and analyser requirement remains with the customer.

Models and ranges

ModelMeasuresRangeSensorNotes
3015Oxygen, trace0 to 1000 ppm (min 0 to 100 ppm)ElectrochemicalSensor life above 18 months; sensor bypass valve supplied
3030Oxygen, percentage0 to 25 % (min 0 to 5 %)ElectrochemicalSensor life above 2 years
3090Oxygen, purity0 to 100 % (min 0 to 5 %)Electrochemical, ambient pressure and temperature compensatedSensor life above 2 years
3130Oxygen, percentage0 to 100 % (min 0 to 5 %)ParamagneticNo consumable sensor; no internal sample pump
3190Oxygen, purity95 to 100 %Paramagnetic, ambient pressure and temperature compensatedNo consumable sensor; no internal sample pump
2015Carbon monoxide0 to 1000 ppm (min 0 to 100 ppm)Electrochemical cellSensor life above 2 years
2030Carbon monoxide0 to 100 % (min 0 to 5 %)NDIRSensor life above 5 years
6015Carbon dioxide0 to 5000 ppm (min 0 to 100 ppm)NDIRSensor life above 5 years
6030Carbon dioxide0 to 20 % or 0 to 100 % (min 0 to 5 %)NDIRSensor life above 5 years
4030Methane0 to 100 % (min 0 to 5 %)NDIRSensor life above 5 years
M015Moisture / dew point-80 to +20 C dp (min -80 to 0 C dp)Ceramic sensorSensor bypass valve supplied; field calibration not possible

Performance by model

ModelRange and resolutionIntrinsic error (accuracy)RepeatabilityResponse T90
30150 to 1000 ppm, resolution 0.1 ppm1 % of reading +/- 2 ppmZero +/- 2 ppm; span 1 % of readingBelow 30 s
30300 to 25 %, resolution 0.01 %+/- 0.01 % of FS+/- 0.04 %Below 20 s
30900 to 100 %, resolution 0.01 %+/- 0.1 % of FS+/- 0.5 %Below 20 s
31300 to 100 %, resolution 0.01 %+/- 0.1 % of FS+/- 0.1 % of FSBelow 5 s
319095 to 100 %, resolution 0.01 %+/- 0.5 % of FS+/- 0.5 % of FSBelow 5 s
M015-80 to +20 C dp, resolution 1 C+/- 2 C dp1 C dp2 min at 5 LPM, dry to wet
20150 to 1000 ppm, resolution 0.1 ppm1 % of FSZero +/- 2 ppm; span +/- 10 ppmBelow 30 s
20300 to 100 %, resolution 0.01 %2 % of FS2 % of readingBelow 30 s
60150 to 5000 ppm, resolution 0.1 ppm1 % of FSZero +/- 10 ppm; span +/- 50 ppmBelow 60 s
60300 to 20 % or 0 to 100 %, resolution 0.01 %+/- 1 % of FS+/- 0.05 %Below 60 s
40300 to 100 %, resolution 0.01 %2 % of FS2 % of readingBelow 45 s

Key features

  • Dual-channel: two different sensors in series, each with its own electronics, current output, relays and Modbus ID
  • Two independent failsafe alarm relays, configured as Lo-LoLo, Lo-Hi or Hi-HiHi
  • Dual-range 4-20 mA output with a range ID relay, useful when start-up readings are high
  • Output freeze: outputs hold their last value during calibration check, calibration and auto-calibration
  • Sensor bypass valve on trace oxygen and dew point models, to isolate the sensor during purging or upsets
  • Optional internal solenoid valves for auto-calibration, started from the keypad, a cyclic timer or Modbus
  • Calibration, alarm and fault history; last calibration with deviation values supports audits
  • Password-protected calibration, set-up and diagnostics; built-in hardware diagnostic mode

How it works

Each ACCURRA-S sensor has its own electronics and microcontroller, which makes the hardware independent. A dual-channel unit houses two sensors in series, with common pneumatics, so you measure two gas components in one sample with separate outputs, relays and Modbus addresses. A fault in one channel does not stop the other.

The analyser is designed to be easy to integrate: an internal fine filter of glass micro fibre, a flow indicator and, on trace oxygen and dew point models, a sensor bypass valve. The sample handling between your process and the analyser decides years of trouble-free performance, so the catalogue states the sample conditions that ACCURRA-S expects.

Requirements and limits

  • The sample gas should be non-corrosive, non-toxic, non-flammable, non-condensing and dry, and free from entrained oil.
  • Sample pressure 2 to 10 psig maximum, regulated externally. Recommended flow 1 LPM.
  • Dust or particulate below 3 micron; the built-in fine filter handles light loads, and heavier dust needs an upstream filter.
  • Materials in contact with sample: SS316, Viton, PTFE, glass, aluminium and acrylic.
  • Sample temperature 5 to 45 C, and a sample dew point at least 5 C below the lowest ambient temperature.
  • Auto-calibration is not available on 3015 and M015, and on dual-gas units the calibration valves are external (relay outputs only). An internal sample pump is not available on 3015, M015, 3130, 3190 or any dual-gas unit. When the pump option is chosen, the range ID relay is unavailable.
  • Field calibration is not possible on the dew point model M015.
  • If your sample does not comply, we design a conditioning system around your application; flammable or toxic gas is reviewed case by case.

On site: utilities and calibration gases

What to have ready at your plant before installation and commissioning.

ItemRequirementNotes
Power at the analyser100 to 240 V AC, 50/60 Hz, 45 W
Zero gasZero-grade nitrogen, 99.9999 %10 litre cylinder with regulator
Span gas, trace oxygenOxygen in nitrogen near the top of the range, for example 1000 ppm10 litre cylinder with regulator
Span gas, 3090Zero-grade oxygen; zero on zero-grade nitrogen
Span gas, 3190Zero-grade oxygen; zero on 95 % oxygen in nitrogen
Span gas, percent CO, CO₂ or CH₄Zero-grade CO, CO₂ or CH₄, 99.999 %
Instrument air and service nitrogen4 kg/cm² minimum
Sample at the analyser2 to 10 psig, about 1 LPM, 5 to 45 °C, dust below 3 micronPressure and flow regulated by the user
Mounting19 inch rack or panel, in a safe area

Options and ordering

  • Ordering code: first and second component gas (model numbers above); auto-calibration as none, relay output only, or internal solenoid valves; relay 4 as range ID, internal sample pump or relay for an external pump; AC power
  • Sample conditioning: SGI-300 cooler, i-Cons complete sampling system, SP-300 heated probe, filters, scrubbers and catchpots

Selection guidance

Choose ACCURRA-S for purity measurement, gas certification, trace impurities, process monitoring in oxygen, nitrogen, argon and CO2 plants, welding gases, fermenter off-gas and furnace atmospheres.

For binary gases such as hydrogen, helium and argon, see the ACCURRA-TCD pages. For hazardous areas see ACCURRA-Hz. For percent oxygen in a flue duct see ACCURRA-Zr.

Applications

Application pages where this is part of the recommended solution:

Compatible products

Frequently asked questions

Which gases can ACCURRA-S measure?

Oxygen, carbon monoxide, carbon dioxide, methane and moisture or dew point, in ranges from ppm to 100 percent depending on the sensor model.

Can one analyser measure two gases?

Yes. The dual-channel version houses two sensors with common pneumatics and independent outputs. See the dual-gas analyser page for the combinations.

What must the sample be like?

Non-corrosive, non-toxic, non-flammable, dry and free of oil, at 2 to 10 psig and about 1 LPM.

Is ACCURRA-S suitable for hazardous areas?

No. It is a safe-area analyser. ACCURRA-Hz is the hazardous-area version.

How does it connect to my control system?

Through an isolated 4-20 mA output with dual range, failsafe relays and Modbus RTU over RS 485.

Talk to an applications engineer

Send the gas, range, pressure, temperature and area classification.

Request a quoteDownload the datasheetGas analysers datasheets, PDF 4.8 MB+91 94 2300 4179Sales; support +91 94 2300 4178