ACCURRA-S is Sarvesh's safe-area online gas analyser for oxygen, CO, CO2, methane and moisture, from ppm to 100 %. It has a 4-20 mA output, failsafe relays and Modbus, and expects a dry, non-corrosive, non-toxic, non-flammable sample at 2 to 10 psig.
Why ACCURRA-S
Accurate from ppm to purity
Trace oxygen to 1 % of reading ±2 ppm; purity oxygen compensated for ambient pressure and temperature, which keeps error below 0.5 %.
Integrates without add-ons
Dual-range isolated 4–20 mA with range ID relay, six failsafe relays, Modbus RTU and output freeze during calibration come as standard.
Maintenance you can plan
Time-stamped calibration, alarm and fault history shows drift and residual sensor life, so consumable sensors are replaced on schedule.
Specifications
| Item | Specification | Notes |
|---|---|---|
| Display | Backlit LCD, 4 lines x 20 characters | |
| Analog output | 1 x 4-20 mA isolated, max load 500 ohm, dual range | Range freely selectable over the full measurement range |
| Relays | 6 x SPST rated 1 A 230 V AC, all failsafe | Alarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves |
| Communication | Modbus RTU over RS 485 | Read measured value, status, set-up and history; write auto-calibration |
| Enclosure | 19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mm | IP 20, safe area only; net about 6 kg, gross about 8 kg |
| Power | 100 to 240 V AC, 50/60 Hz, 45 W | |
| Inlet and outlet | 1/4 inch OD SS316 | |
| Ambient | 5 to 40 C; below 90 % RH non-condensing | Storage 0 to 50 C |
From the current Sarvesh product datasheets and company profile. Specifications may change without notice as products improve. The final responsibility to check that a product meets the process and analyser requirement remains with the customer.
Models and ranges
| Model | Measures | Range | Sensor | Notes |
|---|---|---|---|---|
| 3015 | Oxygen, trace | 0 to 1000 ppm (min 0 to 100 ppm) | Electrochemical | Sensor life above 18 months; sensor bypass valve supplied |
| 3030 | Oxygen, percentage | 0 to 25 % (min 0 to 5 %) | Electrochemical | Sensor life above 2 years |
| 3090 | Oxygen, purity | 0 to 100 % (min 0 to 5 %) | Electrochemical, ambient pressure and temperature compensated | Sensor life above 2 years |
| 3130 | Oxygen, percentage | 0 to 100 % (min 0 to 5 %) | Paramagnetic | No consumable sensor; no internal sample pump |
| 3190 | Oxygen, purity | 95 to 100 % | Paramagnetic, ambient pressure and temperature compensated | No consumable sensor; no internal sample pump |
| 2015 | Carbon monoxide | 0 to 1000 ppm (min 0 to 100 ppm) | Electrochemical cell | Sensor life above 2 years |
| 2030 | Carbon monoxide | 0 to 100 % (min 0 to 5 %) | NDIR | Sensor life above 5 years |
| 6015 | Carbon dioxide | 0 to 5000 ppm (min 0 to 100 ppm) | NDIR | Sensor life above 5 years |
| 6030 | Carbon dioxide | 0 to 20 % or 0 to 100 % (min 0 to 5 %) | NDIR | Sensor life above 5 years |
| 4030 | Methane | 0 to 100 % (min 0 to 5 %) | NDIR | Sensor life above 5 years |
| M015 | Moisture / dew point | -80 to +20 C dp (min -80 to 0 C dp) | Ceramic sensor | Sensor bypass valve supplied; field calibration not possible |
Performance by model
| Model | Range and resolution | Intrinsic error (accuracy) | Repeatability | Response T90 |
|---|---|---|---|---|
| 3015 | 0 to 1000 ppm, resolution 0.1 ppm | 1 % of reading +/- 2 ppm | Zero +/- 2 ppm; span 1 % of reading | Below 30 s |
| 3030 | 0 to 25 %, resolution 0.01 % | +/- 0.01 % of FS | +/- 0.04 % | Below 20 s |
| 3090 | 0 to 100 %, resolution 0.01 % | +/- 0.1 % of FS | +/- 0.5 % | Below 20 s |
| 3130 | 0 to 100 %, resolution 0.01 % | +/- 0.1 % of FS | +/- 0.1 % of FS | Below 5 s |
| 3190 | 95 to 100 %, resolution 0.01 % | +/- 0.5 % of FS | +/- 0.5 % of FS | Below 5 s |
| M015 | -80 to +20 C dp, resolution 1 C | +/- 2 C dp | 1 C dp | 2 min at 5 LPM, dry to wet |
| 2015 | 0 to 1000 ppm, resolution 0.1 ppm | 1 % of FS | Zero +/- 2 ppm; span +/- 10 ppm | Below 30 s |
| 2030 | 0 to 100 %, resolution 0.01 % | 2 % of FS | 2 % of reading | Below 30 s |
| 6015 | 0 to 5000 ppm, resolution 0.1 ppm | 1 % of FS | Zero +/- 10 ppm; span +/- 50 ppm | Below 60 s |
| 6030 | 0 to 20 % or 0 to 100 %, resolution 0.01 % | +/- 1 % of FS | +/- 0.05 % | Below 60 s |
| 4030 | 0 to 100 %, resolution 0.01 % | 2 % of FS | 2 % of reading | Below 45 s |
Key features
- Dual-channel: two different sensors in series, each with its own electronics, current output, relays and Modbus ID
- Two independent failsafe alarm relays, configured as Lo-LoLo, Lo-Hi or Hi-HiHi
- Dual-range 4-20 mA output with a range ID relay, useful when start-up readings are high
- Output freeze: outputs hold their last value during calibration check, calibration and auto-calibration
- Sensor bypass valve on trace oxygen and dew point models, to isolate the sensor during purging or upsets
- Optional internal solenoid valves for auto-calibration, started from the keypad, a cyclic timer or Modbus
- Calibration, alarm and fault history; last calibration with deviation values supports audits
- Password-protected calibration, set-up and diagnostics; built-in hardware diagnostic mode
How it works
Each ACCURRA-S sensor has its own electronics and microcontroller, which makes the hardware independent. A dual-channel unit houses two sensors in series, with common pneumatics, so you measure two gas components in one sample with separate outputs, relays and Modbus addresses. A fault in one channel does not stop the other.
The analyser is designed to be easy to integrate: an internal fine filter of glass micro fibre, a flow indicator and, on trace oxygen and dew point models, a sensor bypass valve. The sample handling between your process and the analyser decides years of trouble-free performance, so the catalogue states the sample conditions that ACCURRA-S expects.
Requirements and limits
- The sample gas should be non-corrosive, non-toxic, non-flammable, non-condensing and dry, and free from entrained oil.
- Sample pressure 2 to 10 psig maximum, regulated externally. Recommended flow 1 LPM.
- Dust or particulate below 3 micron; the built-in fine filter handles light loads, and heavier dust needs an upstream filter.
- Materials in contact with sample: SS316, Viton, PTFE, glass, aluminium and acrylic.
- Sample temperature 5 to 45 C, and a sample dew point at least 5 C below the lowest ambient temperature.
- Auto-calibration is not available on 3015 and M015, and on dual-gas units the calibration valves are external (relay outputs only). An internal sample pump is not available on 3015, M015, 3130, 3190 or any dual-gas unit. When the pump option is chosen, the range ID relay is unavailable.
- Field calibration is not possible on the dew point model M015.
- If your sample does not comply, we design a conditioning system around your application; flammable or toxic gas is reviewed case by case.
On site: utilities and calibration gases
What to have ready at your plant before installation and commissioning.
| Item | Requirement | Notes |
|---|---|---|
| Power at the analyser | 100 to 240 V AC, 50/60 Hz, 45 W | |
| Zero gas | Zero-grade nitrogen, 99.9999 % | 10 litre cylinder with regulator |
| Span gas, trace oxygen | Oxygen in nitrogen near the top of the range, for example 1000 ppm | 10 litre cylinder with regulator |
| Span gas, 3090 | Zero-grade oxygen; zero on zero-grade nitrogen | |
| Span gas, 3190 | Zero-grade oxygen; zero on 95 % oxygen in nitrogen | |
| Span gas, percent CO, CO₂ or CH₄ | Zero-grade CO, CO₂ or CH₄, 99.999 % | |
| Instrument air and service nitrogen | 4 kg/cm² minimum | |
| Sample at the analyser | 2 to 10 psig, about 1 LPM, 5 to 45 °C, dust below 3 micron | Pressure and flow regulated by the user |
| Mounting | 19 inch rack or panel, in a safe area |
Options and ordering
- Ordering code: first and second component gas (model numbers above); auto-calibration as none, relay output only, or internal solenoid valves; relay 4 as range ID, internal sample pump or relay for an external pump; AC power
- Sample conditioning: SGI-300 cooler, i-Cons complete sampling system, SP-300 heated probe, filters, scrubbers and catchpots
Selection guidance
Choose ACCURRA-S for purity measurement, gas certification, trace impurities, process monitoring in oxygen, nitrogen, argon and CO2 plants, welding gases, fermenter off-gas and furnace atmospheres.
For binary gases such as hydrogen, helium and argon, see the ACCURRA-TCD pages. For hazardous areas see ACCURRA-Hz. For percent oxygen in a flue duct see ACCURRA-Zr.
Applications
Application pages where this is part of the recommended solution:
- Oxygen purity in air separation plants
- Nitrogen purity and trace oxygen
- Gas purity certification and cylinder filling
- Dew point and trace moisture in process gases
- Fermenter and bioreactor off-gas
- Nitrogen blanketing and inerting
- Heat-treatment furnace atmosphere
- Semiconductor and electronics gas purity
Compatible products
i-Cons
Smart Peltier sample conditioner with diagnostics, auto blowback and Modbus, in a compact 19 inch cabinet.
View i-Cons
SGI-300 / SGI-320
Peltier sample cooler with integrated pump, filter and condensate detector; SGI-320 for emission monitoring.
View SGI-300 / SGI-320
SP and SRP sample probes
Heated and non-heated probes with blowback, and retractable probes, in SS316, Inconel or ceramic.
View SP and SRP sample probesHeadline filters
Particulate, coalescing and PTFE membrane filters from Headline Filters, UK, for analyser protection.
View Headline filtersACCURRA-TCD
Thermal conductivity analyser for binary mixtures of H2, He, Ar, N2 and others, with no reference gas.
View ACCURRA-TCDFrequently asked questions
Which gases can ACCURRA-S measure?
Oxygen, carbon monoxide, carbon dioxide, methane and moisture or dew point, in ranges from ppm to 100 percent depending on the sensor model.
Can one analyser measure two gases?
Yes. The dual-channel version houses two sensors with common pneumatics and independent outputs. See the dual-gas analyser page for the combinations.
What must the sample be like?
Non-corrosive, non-toxic, non-flammable, dry and free of oil, at 2 to 10 psig and about 1 LPM.
Is ACCURRA-S suitable for hazardous areas?
No. It is a safe-area analyser. ACCURRA-Hz is the hazardous-area version.
How does it connect to my control system?
Through an isolated 4-20 mA output with dual range, failsafe relays and Modbus RTU over RS 485.
Related pages
- ACCURRA-S oxygen analysers
- ACCURRA-S carbon monoxide analysers
- ACCURRA-S carbon dioxide analysers
- ACCURRA-S methane analyser
- ACCURRA-S M015 moisture analyser
- ACCURRA-Hz
- Hazardous-area oxygen analyser
- INFINITY-Tx
- Hazardous-area hydrogen analyser
- Hazardous-area moisture analyser
- ACCURRA-Zr
- ACCURRA model numbers
- Browse all gas analysers