ACCURRA-TCD · Argon

Online argon gas analyser based on thermal conductivity

ACCURRA-TCD measures argon in a binary mixture by thermal conductivity, with a platinum sensor and no reference gas. Use it for argon purity and for argon in nitrogen, helium or hydrogen.

Thermal conductivity cellA sample gas flows through a cell containing a heated platinum element. The rate at which the gas carries heat away is the measurement signal.Platinum element, held at constant temperatureSample inSample outHeat carried away by the gas sets the signal
  • ISO 9001:2015, certified by TÜV
  • Designed and built in Pune since 2002
  • Supplied to Reliance, Tata Steel, Indian Oil, Linde and ISRO
Quick answer

ACCURRA-TCD measures argon in a binary mixture by thermal conductivity with no reference gas. It works very well against helium or hydrogen and moderately against nitrogen. For oxygen or CO2 impurity in argon, use ACCURRA-S instead.

Updated 3 November 2026 · Figures from Sarvesh datasheets

Specifications

ItemSpecificationNotes
GasArgon in a binary mixture
SensorPlatinum thermal conductivity
RepeatabilityBelow 1 % of full scaleLinearity below 1 % of range
Errors and driftBelow 1 % of full scale per week
Response timeT90 below 10 s at 0.6 LPM
Reference gasNot required

From the current Sarvesh product datasheets and company profile. Specifications may change without notice as products improve. The final responsibility to check that a product meets the process and analyser requirement remains with the customer.

Key features

  • Binary argon measurement
  • Isolated 4-20 mA, programmable alarms, Modbus RTU
  • No reference gas

How it works

Argon conducts heat roughly a third less well than nitrogen, so an argon and nitrogen mixture gives a clear but modest signal, and argon in helium or hydrogen gives a very strong one. The analyser's range is chosen to suit.

Typical uses: argon purity at air separation and filling plants, argon in furnace and welding atmospheres, argon-nitrogen blends. Argon and oxygen have similar thermal conductivity, so oxygen in argon is better measured with an oxygen sensor; use ACCURRA-S for oxygen impurity in argon.

Requirements and limits

  • Binary mixtures only.

On site: utilities and calibration gases

What to have ready at your plant before installation and commissioning.

ItemRequirementNotes
Power at the analyser100 to 240 V AC, 50/60 Hz, 45 W
Zero gasZero-grade nitrogen, 99.999 %10 litre cylinder with regulator
Span gasThe measured gas at zero grade, 99.999 %: hydrogen, helium, argon or nitrous oxide10 litre cylinder with regulator
Sample at the analyser2 to 10 psig, 0.6 LPM, 5 to 40 °C, dust below 3 micronPressure and flow regulated by the user
Mounting19 inch rack or panel, in a safe area

Selection guidance

Choose ACCURRA-TCD for argon as the main component of a binary mixture, and ACCURRA-S for trace impurities in argon such as oxygen, CO2 and moisture.

Applications

Application pages where this is part of the recommended solution:

Compatible products

Frequently asked questions

Can it measure oxygen in argon?

Argon and oxygen have similar conductivity; use an ACCURRA-S oxygen model for oxygen impurity.

Does it need a reference gas?

No.

What about argon-CO2 mixes?

Argon in CO2 is listed on wide ranges only, 40 to 100 or 80 to 100 percent, because the two conduct heat similarly. Low levels of CO2 in argon are better read by NDIR on ACCURRA-S 6030.

Which backgrounds?

Nitrogen, air, oxygen, carbon dioxide or xenon, one at a time.

Talk to an applications engineer

Send the gas, range, pressure, temperature and area classification.