Design a sample system
Follow the sample from the tapping to the sensor and design each step on purpose.
What you will be able to do
- Choose a representative sample point
- Size tubing, flow and lag time
- Select filters, coolers and probes for the sample
- Keep condensate and dust out of the analyser
- Know which samples are out of scope
Your progress: 0 of 10 lessons read
Follow these lessons in order
- How to design a gas sample handling systemA good sample handling system delivers a representative, clean, dry sample at constant pressure and flow.The overview.Intermediate4 min
- Choosing the sample point for a gas analyserTake the sample where the gas is well mixed, at process conditions you understand, upstream of any air leak, away from dead zones, and with a tapping that cannot fill with liquid or dust.Where to tap.Intermediate4 min
- Sample lines: materials, length, diameter and lagUse stainless steel tube for trace oxygen, moisture and reactive gases, because plastic lets oxygen and water in and out.Tubing and time delay.Intermediate4 min
- Filters for gas analysers: dust, mist, membraneUse filtration in stages: a coarse filter or probe filter near the tapping for bulk dust, a finer filter after the cooler, a coalescing filter to remove liquid mist and oil aerosol, and a membrane filter as a final guard.Dust and mist.Intermediate3 min
- Condensate, moisture and sample cooling for gas analysisWet samples must be cooled to a fixed, low dew point so that water leaves as condensate at a controlled place, not in the analyser.Water in the sample.Intermediate3 min
- Sample pressure, flow and fast bypass loopsAn analyser needs a steady sample at its specified pressure and flow.Pressure and flow.Intermediate3 min
- Probes and blowback for hot, dusty and tarry gasHot, dusty or tarry gas needs a probe that survives the temperature, a filter at the tip that keeps dust out, and a way to clear the filter, usually automatic blowback with clean air or gas.Hard samples.Intermediate3 min
- Sampling cryogenic and liquefied gases: vaporisersA liquid sample must be vaporised completely before it reaches an analyser.Very cold samples.Intermediate3 min
- What samples an analyser cannot take: know the limitsStandard ACCURRA-S analysers expect a non-corrosive, non-toxic, non-flammable, non-condensing and dry sample, free of oil, at 2 to 10 psig and about 1 LPM, with dust below 3 micron.The limits.Intermediate3 min
- Sample system troubleshooting: symptoms to causesMost analyser faults are sample system faults.When it goes wrong.Intermediate3 min