Gas analysis glossary
170 plain definitions of the terms used on analyser data sheets, in sample handling and in hazardous-area work, with a link to the lesson that explains each one.
4
- 4-20 mA
- An analogue signal standard in which the reading is carried as a current, 4 mA at the bottom of the range and 20 mA at the top. See Analyser outputs: 4-20 mA, relays and Modbus explained.
- Explained in: Analyser outputs: 4-20 mA, relays and Modbus explained, Wiring, grounding and electrical noise for gas analysers
A
- Absolute pressure
- Pressure measured from a perfect vacuum. Gauge pressure plus atmospheric pressure. Gas laws and partial pressure use absolute pressure.
- Explained in: How gases behave: pressure, temperature and partial pressure
- Accuracy
- How close a reading is to the true value. See Accuracy, precision, repeatability and drift in gas analysers.
- Explained in: Accuracy, precision, repeatability and drift in gas analysers, How to read a gas analyser datasheet, What accuracy do you really need from a gas analyser?
- Adsorption
- The sticking of gas molecules to a solid surface. Used in dryers and PSA beds, and a cause of slow moisture readings as water adsorbs on tube walls.
- Air ingress
- Air leaking into a process or a sample line. At ppm oxygen levels even a very small leak is significant: Trace oxygen measurement and air ingress: ppm done right.
- Alarm set point
- The value at which an alarm activates. Normally paired with a deadband to avoid chatter: Connecting a gas analyser to PLC, DCS and SCADA.
- Explained in: Connecting a gas analyser to PLC, DCS and SCADA
- Ambient temperature
- The temperature of the air around an instrument, which has limits in the datasheet.
- Explained in: Where to mount a gas analyser: panel, shelter or field
- Analyser (analyzer)
- An instrument that measures the concentration of a gas or other property of a process sample and outputs a signal such as 4-20 mA.
- Explained in: What is online gas analysis? A plain introduction, Online vs laboratory vs portable gas analysis
- Analyser house (shelter)
- A small building or cabin that holds analysers close to the process, protected from the weather and often heated and ventilated: Where to mount a gas analyser: panel, shelter or field.
- Explained in: Where to mount a gas analyser: panel, shelter or field
- Asphyxiant
- A gas, such as nitrogen or argon, that harms by displacing oxygen rather than by being toxic: Oxygen enrichment and oxygen deficiency hazards.
- ASU (air separation unit)
- A plant that separates air into oxygen, nitrogen and argon, usually by cryogenic distillation.
- Explained in: How an air separation plant works and what is measured
- ATEX
- European directives covering equipment for explosive atmospheres. Marking states group, category and temperature class. Other schemes, such as IECEx, are similar in intent.
- Explained in: Hazardous area classification: zones, gas groups, T-Classes
- Audit trail
- A record of actions, such as calibrations and setting changes, with date and who did them, used by quality systems: Records, audit trails and qualification for regulated plants.
- Explained in: Records, audit trails and qualification for regulated plants
- Auto-calibration
- A function that switches zero and span gas to the analyser and adjusts it automatically, started by a timer, keypad or Modbus command. Optional on most ACCURRA-S models.
- Explained in: Auto-Calibration and validation checks for gas analysers, Total cost of ownership of a gas analyser
- Autoignition temperature
- The temperature at which a flammable mixture ignites without a spark. Equipment surface temperature class must stay below it.
B
- Back pressure
- Pressure at the vent or outlet of a sample line, which changes the sample pressure at the sensor.
- Balance gas
- The main gas in a calibration mixture, such as nitrogen, that carries the small amounts of target gas: Calibration gases: choosing, handling and certificates.
- Explained in: Calibration gases: choosing, handling and certificates
- Barometric pressure
- The pressure of the atmosphere, which varies with weather and altitude; about 1013.25 hPa at sea level in a standard atmosphere.
- Explained in: How gases behave: pressure, temperature and partial pressure
- Binary mixture
- A gas mixture of two components. Thermal conductivity analysers measure a binary mixture directly.
- Explained in: How thermal conductivity gas analysers (TCD) work, Measuring hydrogen: purity, mixtures and safety limits, Steel plant gases: coke oven, blast furnace, converter
- Biomethane
- Biogas upgraded to a high methane content, usable like natural gas.
- Blanketing
- Maintaining a slight positive pressure of inert gas above a liquid or powder so air cannot enter: Inert gas blanketing in pharma and chemical plants.
- Explained in: Inert gas blanketing in pharma and chemical plants
- Blast furnace gas
- A lean by-product fuel gas of iron making, mostly nitrogen, CO and CO2: Steel plant gases: coke oven, blast furnace, converter.
- Blowback
- A burst of clean air or gas sent back through a probe filter to clear dust, triggered on a timer or when sample flow fails.
- Explained in: How a zirconia oxygen analyser works, Filters for gas analysers: dust, mist, membrane, Probes and blowback for hot, dusty and tarry gas
- Bypass loop (fast loop)
- A line that circulates a large flow past the tapping, from which the analyser draws a small branch, to reduce lag: Sample pressure, flow and fast bypass loops.
- Explained in: Sample pressure, flow and fast bypass loops
C
- Calibration
- Adjusting an analyser against gas of known concentration so its reading is correct.
- Explained in: Why ambient pressure affects oxygen purity readings, What is online gas analysis? A plain introduction, Online vs laboratory vs portable gas analysis
- Calibration check
- Applying a known gas and recording the deviation without adjusting the analyser. Compare calibration, which adjusts: How gas analyser calibration works: zero and span.
- Explained in: How gas analyser calibration works: zero and span, How often should you calibrate a gas analyser?, Auto-Calibration and validation checks for gas analysers
- Calibration gas
- A certified gas mixture of known concentration used to calibrate or check an analyser: Calibration gases: choosing, handling and certificates.
- Explained in: Site utilities checklist for gas analysers, Calibration gases: choosing, handling and certificates, What accuracy do you really need from a gas analyser?
- Carbon molecular sieve
- An adsorbent used in nitrogen PSA generators that adsorbs oxygen faster than nitrogen.
- Carbon potential
- The carbon content steel would reach at equilibrium with a furnace atmosphere; the control variable in carburising: Heat treatment furnace atmospheres explained.
- Catchpot
- A vessel that removes liquid condensate from a sample gas before it reaches downstream components.
- CBG (compressed biogas)
- Biogas upgraded to a high methane content and compressed for use as fuel.
- Explained in: Measuring methane and biogas composition online, Biogas, landfill gas and digesters explained
- CEMS
- Continuous emission monitoring system: equipment that measures stack gases and reports them at regulatory conditions: Emission monitoring (CEMS) basics: what a system does.
- Explained in: Emission monitoring (CEMS) basics: what a system does
- Certificate of analysis
- The document that comes with a calibration gas cylinder, stating the concentration, its uncertainty, the date and the traceability.
- Chilled mirror hygrometer
- An instrument that cools a mirror until dew forms and detects it optically. A reference method for dew point: Moisture measurement methods compared for dry gas.
- Explained in: Moisture measurement methods compared for dry gas
- Clinker
- The hot, partly fused product of a cement kiln that is ground to make cement: Cement kiln gas analysis explained: O2, CO and safety.
- Coalescing filter
- A filter that merges fine liquid droplets into larger ones that drain away.
- Explained in: Filters for gas analysers: dust, mist, membrane
- Coke oven gas
- A by-product gas of coke making, rich in hydrogen and methane: Steel plant gases: coke oven, blast furnace, converter.
- Combustibles
- Unburnt gases such as CO and hydrogen in flue gas. They can burn at a hot zirconia cell and make the oxygen reading too low.
- Condensate
- Liquid that forms when a gas is cooled below its dew point.
- Explained in: How to design a gas sample handling system, Measuring carbon monoxide and carbon dioxide online, Measuring methane and biogas composition online
- Controlled atmosphere (CA)
- Storage with oxygen and CO2 held at set levels to slow ripening: Controlled-Atmosphere fruit storage explained.
- Explained in: Controlled-Atmosphere fruit storage explained
- Converter gas
- A CO-rich by-product gas of oxygen steelmaking, recovered when its composition allows: Steel plant gases: coke oven, blast furnace, converter.
- Cross-sensitivity
- A sensor's response to gases other than its target: Cross-Sensitivity and interference in gas analysers.
- Explained in: What is an NDIR gas analyser?, How thermal conductivity gas analysers (TCD) work, Cross-Sensitivity and interference in gas analysers
- Cryogenic
- Relating to very low temperatures, such as liquid oxygen, nitrogen and argon, which boil at about minus 183, minus 196 and minus 186 C.
- Explained in: Sampling cryogenic and liquefied gases: vaporisers, How an air separation plant works and what is measured
D
- DCS
- Distributed control system: a plant control system made of networked controllers and operator stations.
- Explained in: Connecting a gas analyser to PLC, DCS and SCADA
- Dead leg
- A branch of tubing with no flow. It holds old gas and adds lag and contamination.
- Explained in: Response time, T90 and sample lag in gas analysis, Trace oxygen measurement and air ingress: ppm done right, Choosing the sample point for a gas analyser
- Dead volume
- Volume in a sample system that is not swept by the flow, such as dead legs and oversized fittings.
- Detection limit
- The smallest concentration an analyser can distinguish from zero.
- Explained in: Accuracy, precision, repeatability and drift in gas analysers
- Dew point
- The temperature at which water vapour in a gas begins to condense.
- Explained in: What is dew point and how is it measured?, ppm, percent and ppb explained for gas analysis, Dry basis vs wet basis in gas measurement
- Digester
- A sealed vessel in which organic matter is broken down without air to produce biogas: Biogas, landfill gas and digesters explained.
- Drain
- A route for condensate to leave a sample system, which must be continuous and safe: Condensate, moisture and sample cooling for gas analysis.
- Explained in: Site utilities checklist for gas analysers
- Drift
- A slow change in an analyser's reading over time with the same input gas.
- Explained in: Online vs laboratory vs portable gas analysis, Accuracy, precision, repeatability and drift in gas analysers, How often should you calibrate a gas analyser?
- Dry basis
- A gas value stated as if all water vapour were removed. Contrast wet basis: Dry basis vs wet basis in gas measurement.
- Explained in: Dry basis vs wet basis in gas measurement, Combustion, excess air and flue gas oxygen explained, Emission monitoring (CEMS) basics: what a system does
- Dual-channel analyser
- An analyser that measures two gases from one sample in one enclosure, such as ACCURRA-S oxygen with CO2 or oxygen with dew point.
- Explained in: Measuring methane and biogas composition online, Glove boxes and Controlled-Atmosphere chambers explained
E
- Electrochemical cell
- A sensor that generates a current from a chemical reaction with the target gas, used for trace oxygen and CO. The cell is consumed over time.
- Explained in: Electrochemical vs paramagnetic oxygen sensors, Gas analysis technologies compared: which sensor for which gas, Cross-Sensitivity and interference in gas analysers
- Electrolyser
- A device that splits water into hydrogen and oxygen using electricity: Hydrogen electrolysers and gas analysis: what to measure.
- Enclosure rating (IP)
- The ingress protection rating that states how well an enclosure keeps out dust and water, for example IP65.
- Explained in: Where to mount a gas analyser: panel, shelter or field
- Endothermic gas
- A furnace atmosphere made by partly burning natural gas over a catalyst, mostly CO, hydrogen and nitrogen: Heat treatment furnace atmospheres explained.
- Explained in: Heat treatment furnace atmospheres explained
- ESP (electrostatic precipitator)
- A dust collector that charges particles with high voltage. It can ignite gas if CO and unburnt fuel are present: Cement kiln gas analysis explained: O2, CO and safety.
- Excess air
- Air supplied to a burner beyond the amount needed for exact combustion, estimated from flue gas oxygen: Excess air from flue gas oxygen calculator.
- Explained in: Combustion, excess air and flue gas oxygen explained
- Explosive atmosphere
- A mixture of air with flammable gas, vapour or dust, within its explosive limits, that can burn or explode if ignited.
- Extractive analysis
- Measuring a gas by drawing a sample from the process to an analyser, usually through a conditioning system.
- Explained in: Extractive vs In-Situ gas analysis, What is online gas analysis? A plain introduction
F
- Failsafe relay
- A relay that moves to its alarm state if power or the instrument fails.
- Explained in: What is limiting oxygen concentration (LOC)?, How to read a gas analyser datasheet, Analyser outputs: 4-20 mA, relays and Modbus explained
- Fine filter
- A final-stage filter before the analyser; in ACCURRA-S a glass micro fibre filter is built in.
- Explained in: How to design a gas sample handling system, Filters for gas analysers: dust, mist, membrane, Maintenance schedule and spares for gas analysers
- Flame arrestor
- A device that stops a flame travelling along a pipe by absorbing its heat in narrow passages.
- Flameproof enclosure (Ex d)
- An enclosure that withstands an internal explosion and prevents it igniting the surrounding atmosphere.
- Explained in: Hazardous area gas analysers explained, Protection concepts: flameproof, intrinsic safety, purging, Inert gas blanketing in pharma and chemical plants
- Flue gas
- The gas leaving a furnace, boiler or kiln after combustion.
- Fractionation
- The separation of a liquid mixture by boiling, the lighter component leaving first. It is why liquid samples must be vaporised completely: Sampling cryogenic and liquefied gases: vaporisers.
- Explained in: Sampling cryogenic and liquefied gases: vaporisers
- Frost point
- The temperature at which moisture deposits as ice. Dew points below 0 C in dry gas are usually frost points.
- Explained in: Moisture measurement methods compared for dry gas
G
- Galvanic isolation
- Electrical separation between circuits so no direct current path exists, used to break ground loops and protect electronics.
- Explained in: Wiring, grounding and electrical noise for gas analysers
- Gas chromatograph
- An instrument that separates the components of a mixture in a column and measures each. Used for multi-component analysis; not in the Sarvesh range.
- Explained in: Gas analysis technologies compared: which sensor for which gas
- Gas group IIC
- The most demanding hazardous-area gas group, covering hydrogen and acetylene.
- Explained in: Hazardous area gas analysers explained, Measuring hydrogen: purity, mixtures and safety limits, Flammability, explosive limits and gas safety basics
- Gauge pressure
- Pressure measured relative to the atmosphere. Add atmospheric pressure to get absolute pressure.
- Ground loop
- A circuit formed by two ground connections at different potentials, which drives current through a signal path and adds noise: Wiring, grounding and electrical noise for gas analysers.
- Explained in: Wiring, grounding and electrical noise for gas analysers
H
- H2S
- Hydrogen sulphide: a toxic, corrosive, flammable gas found in biogas and some process streams. It is outside the standard ACCURRA-S sample limits.
- Hazardous area
- A place where an explosive atmosphere may occur. Classified into zones: Hazardous area classification: zones, gas groups, T-Classes.
- Heat tracing
- Electrical or steam heating of a line to keep it above the dew point of the gas inside.
- Explained in: Sample lines: materials, length, diameter and lag, Condensate, moisture and sample cooling for gas analysis, Probes and blowback for hot, dusty and tarry gas
- Hot-wet extraction
- Keeping the sample hot and wet from the probe to a heated analyser so that soluble gases are not lost: Emission monitoring (CEMS) basics: what a system does.
- HRSG
- Heat recovery steam generator, which makes steam from the hot exhaust of a gas turbine.
- Hygroscopic
- Readily absorbing moisture from the surroundings.
- Explained in: Moisture measurement methods compared for dry gas
- Hysteresis
- A deadband that makes an alarm switch on at one value and off at a different value, to prevent chatter.
- Explained in: Connecting a gas analyser to PLC, DCS and SCADA
I
- IECEx
- An international scheme for certifying equipment for explosive atmospheres.
- Explained in: Hazardous area classification: zones, gas groups, T-Classes
- In-situ analysis
- Measuring the gas where it flows, with the sensor in the duct or pipe and no sample line.
- Explained in: How a zirconia oxygen analyser works, Extractive vs In-Situ gas analysis, What is online gas analysis? A plain introduction
- Inerting
- Replacing air in a vessel with an inert gas such as nitrogen to keep oxygen below the level that supports combustion.
- Explained in: What is limiting oxygen concentration (LOC)?, Flammability, explosive limits and gas safety basics, Inerting and purging: how they work and how to prove them
- Interference
- An error in a reading caused by another gas or condition in the sample.
- Explained in: Cross-Sensitivity and interference in gas analysers
- Interlock
- A control or safety logic that stops or prevents an action when a condition, such as high oxygen, is met.
- Intrinsic safety (Ex i)
- A protection concept that limits electrical energy so that no spark or heat can ignite a flammable atmosphere: Protection concepts: flameproof, intrinsic safety, purging.
- Explained in: Protection concepts: flameproof, intrinsic safety, purging
L
- Lambda
- The ratio of actual air to the air needed for exact combustion. Lambda 1 is stoichiometric: Excess air from flue gas oxygen calculator.
- Explained in: Combustion, excess air and flue gas oxygen explained
- Landfill gas
- A methane-rich gas produced by waste in landfills, drawn from wells and used or flared.
- Leak test
- A test, with pressure and a detection method, that shows whether a sample system or enclosure leaks.
- LEL
- Lower explosive limit: the lowest concentration of a flammable gas in air that can burn or explode. About 4 % for hydrogen and 5 % for methane: Flammability, explosive limits and gas safety basics.
- Explained in: Measuring hydrogen: purity, mixtures and safety limits, Flammability, explosive limits and gas safety basics, Hydrogen electrolysers and gas analysis: what to measure
- Limiting oxygen concentration (LOC)
- The oxygen level below which a fuel and inert gas mixture cannot sustain a flame.
- Explained in: What is limiting oxygen concentration (LOC)?, Inerting and purging: how they work and how to prove them, Inert gas blanketing in pharma and chemical plants
- Linearity
- How closely the response of an analyser follows a straight line over its range.
- Explained in: Accuracy, precision, repeatability and drift in gas analysers
- Live zero
- The use of 4 mA, not 0 mA, for the bottom of a 4-20 mA range, so a broken wire can be recognised.
- Explained in: Analyser outputs: 4-20 mA, relays and Modbus explained
M
- Membrane filter
- A thin porous membrane, often PTFE, that removes fine particles and aerosols and repels water: Filters for gas analysers: dust, mist, membrane.
- Explained in: Filters for gas analysers: dust, mist, membrane
- Modbus RTU
- A serial communication protocol, here over RS 485, used to read measurements, status and history from an analyser.
- Explained in: How to read a gas analyser datasheet, Analyser outputs: 4-20 mA, relays and Modbus explained, Connecting a gas analyser to PLC, DCS and SCADA
- Molar volume
- The volume of one mole of gas: 22.414 litres at 0 C and 1013.25 hPa for an ideal gas.
- Explained in: ppm to mg/Nm3: volume vs mass concentration
- Mole fraction
- The share of molecules in a mixture that are a given gas. Percent and ppm by volume are mole fractions.
- Explained in: How gases behave: pressure, temperature and partial pressure
- Molecular sieve
- A porous solid that adsorbs molecules by size, used to dry gas and remove CO2 and in PSA beds.
N
- NDIR
- Non-dispersive infrared: a method that measures gases such as CO, CO2 and methane by their absorption of infrared light.
- Explained in: What is an NDIR gas analyser?, Gas analysis technologies compared: which sensor for which gas, Cross-Sensitivity and interference in gas analysers
- Needle valve
- A fine adjustment valve used to set sample flow.
- Normal cubic metre (Nm3)
- A cubic metre of gas at normal conditions, usually 0 C and 1013.25 hPa: ppm to mg/Nm3: volume vs mass concentration.
- Explained in: ppm to mg/Nm3: volume vs mass concentration
O
- Oxygen deficiency
- An atmosphere with less oxygen than normal, commonly below about 19.5 %, which endangers life: Oxygen enrichment and oxygen deficiency hazards.
- Explained in: Oxygen enrichment and oxygen deficiency hazards, Controlled-Atmosphere fruit storage explained
- Oxygen enrichment
- An atmosphere with more oxygen than normal, commonly above about 23.5 %, in which materials ignite easily.
- Explained in: Oxygen enrichment and oxygen deficiency hazards
P
- Paramagnetic sensor
- An oxygen sensor that uses oxygen's attraction to a magnetic field. It has no consumed element.
- Explained in: Electrochemical vs paramagnetic oxygen sensors, Gas analysis technologies compared: which sensor for which gas
- Partial pressure
- The pressure a gas in a mixture would exert alone: its mole fraction times the total pressure: How gases behave: pressure, temperature and partial pressure.
- Explained in: How gases behave: pressure, temperature and partial pressure
- Peltier cooler
- A solid-state cooler with no refrigerant or compressor, used to cool and dry sample gas.
- Explained in: How to design a gas sample handling system, Dry basis vs wet basis in gas measurement, Condensate, moisture and sample cooling for gas analysis
- Permeation
- The slow passage of gas through the wall of a material such as plastic tubing, a cause of oxygen and moisture error at trace levels.
- Explained in: Sample lines: materials, length, diameter and lag
- PLC
- Programmable logic controller: a small industrial computer that runs control logic.
- Explained in: Connecting a gas analyser to PLC, DCS and SCADA
- Poisoning (sensor)
- Loss of sensor performance caused by a contaminant, such as sulphur, chlorine or silicone: Sensor life, poisoning and replacement in gas analysers.
- Explained in: Sensor life, poisoning and replacement in gas analysers
- ppb
- Parts per billion by volume: one part in a billion. 1000 ppb is 1 ppm: ppm, percent and ppb explained for gas analysis.
- Explained in: ppm, percent and ppb explained for gas analysis
- ppm
- Parts per million by volume. 10,000 ppm is 1 %.
- Explained in: What is dew point and how is it measured?, How gases behave: pressure, temperature and partial pressure, ppm, percent and ppb explained for gas analysis
- Precision
- How closely repeated readings of the same gas agree with each other.
- Explained in: Accuracy, precision, repeatability and drift in gas analysers
- Preheater
- A tower of cyclones that heats raw meal with kiln gas before the kiln in a cement plant.
- Pressure dew point
- The dew point of a gas at its line pressure, usually much higher than the dew point after expansion to atmosphere: Compressed air quality and dew point (ISO 8573-1).
- Explained in: Compressed air quality and dew point (ISO 8573-1)
- Pressure regulator
- A device that reduces and holds a steady downstream pressure.
- Explained in: Sample pressure, flow and fast bypass loops
- Probe
- The first part of a sampling system, inserted in the process to draw a sample through a filter.
- Explained in: Extractive vs In-Situ gas analysis, Choosing the sample point for a gas analyser, Probes and blowback for hot, dusty and tarry gas
- Proof test
- A periodic test that reveals hidden failures in a safety function, such as applying test gas to an analyser.
- Explained in: What SIL means for gas analysers, and what it does not
- PSA / VPSA
- Pressure swing adsorption and vacuum pressure swing adsorption, processes that make nitrogen or oxygen on site using adsorbent beds.
- Explained in: PSA, VPSA and membrane gas generators explained
- Purge
- To replace the gas in a vessel or line with another gas, usually inert, before use or opening: Inerting and purging: how they work and how to prove them.
- Explained in: Trace oxygen measurement and air ingress: ppm done right, Commissioning an online gas analyser: A Step-by-Step guide, Inerting and purging: how they work and how to prove them
- Purging (Ex p)
- A protection concept in which an enclosure is kept at a slight positive pressure of clean air or inert gas to keep hazardous gas out.
- Explained in: Protection concepts: flameproof, intrinsic safety, purging
- Purity analyser
- An analyser that measures a high concentration of a gas, near 100 %, such as oxygen purity from 95 to 100 %.
- Explained in: Why ambient pressure affects oxygen purity readings, Choosing an oxygen analyser: trace, percent or purity, How an air separation plant works and what is measured
R
- Range
- The span of concentration an analyser measures, for example 0 to 1000 ppm.
- Explained in: Why ambient pressure affects oxygen purity readings, Electrochemical vs paramagnetic oxygen sensors, Choosing an oxygen analyser: trace, percent or purity
- Reference conditions
- The temperature and pressure, dry or wet basis and reference oxygen to which a concentration is stated.
- Explained in: ppm to mg/Nm3: volume vs mass concentration, Emission monitoring (CEMS) basics: what a system does
- Reference gas
- A second gas supplied to some analysers for comparison. ACCURRA-TCD needs none.
- Explained in: How a zirconia oxygen analyser works, How thermal conductivity gas analysers (TCD) work
- Repeatability
- The agreement between repeated readings of the same gas under the same conditions.
- Explained in: Accuracy, precision, repeatability and drift in gas analysers
- Resolution
- The smallest change an instrument can display or output.
- Explained in: Accuracy, precision, repeatability and drift in gas analysers
- Respiratory quotient
- The ratio of CO2 produced to oxygen consumed by respiring fruit or cells. Used in some storage and fermentation controls.
- Explained in: Controlled-Atmosphere fruit storage explained
- Response time (T90)
- The time for a reading to reach 90 % of a step change in concentration.
- Explained in: Response time, T90 and sample lag in gas analysis, How to read a gas analyser datasheet
- Rotameter
- A flow meter with a float in a tapered tube, used to see and set sample flow.
- Explained in: Sample pressure, flow and fast bypass loops, Controlled-Atmosphere fruit storage explained
- RS-485
- A two-wire serial communication standard used by Modbus RTU, allowing many devices on one cable over long distances.
- Explained in: Analyser outputs: 4-20 mA, relays and Modbus explained
S
- Safe area
- A place classified as free of explosive atmospheres, where ordinary equipment may be used.
- Safety instrumented function
- A function that detects a dangerous condition and brings the process to a safe state, with a defined safety integrity level: What SIL means for gas analysers, and what it does not.
- Explained in: What SIL means for gas analysers, and what it does not
- Sample conditioning
- Treating a sample, by filtering, cooling, drying and regulating, so that it suits the analyser.
- Sample handling system
- The probe, lines, cooler, pump and filters that deliver a conditioned sample to an analyser.
- Explained in: Extractive vs In-Situ gas analysis, How to design a gas sample handling system, What is online gas analysis? A plain introduction
- Sample lag
- The time for a sample to travel from the tapping to the analyser, equal to line volume divided by flow: Gas sample line lag time calculator.
- Explained in: Response time, T90 and sample lag in gas analysis, Sample lines: materials, length, diameter and lag
- Sample point (tapping)
- The connection on the process from which gas is taken for analysis: Choosing the sample point for a gas analyser.
- Explained in: Choosing the sample point for a gas analyser
- SCADA
- Supervisory control and data acquisition: software that collects plant data, shows it to operators and stores it.
- Explained in: Connecting a gas analyser to PLC, DCS and SCADA
- Scrubber
- A filter or chemical bed that removes an interfering or harmful gas from a sample before the sensor.
- Explained in: Cross-Sensitivity and interference in gas analysers, What samples an analyser cannot take: know the limits
- Selectivity
- The ability of a sensor to respond only to its target gas.
- Sensor bypass valve
- A valve that isolates trace oxygen or dew point sensors during purging or upsets so they recover quickly.
- Explained in: What is dew point and how is it measured?, Response time, T90 and sample lag in gas analysis, Moisture measurement methods compared for dry gas
- Sensor life
- The time over which a sensor performs within specification, which depends on the type and the sample: Sensor life, poisoning and replacement in gas analysers.
- Explained in: Maintenance schedule and spares for gas analysers, Sensor life, poisoning and replacement in gas analysers, Total cost of ownership of a gas analyser
- Shielded cable
- A cable with a conductive screen that protects the signal from electrical noise.
- Explained in: Wiring, grounding and electrical noise for gas analysers
- SIL
- Safety integrity level, a measure of the reliability of a safety function under functional safety standards. Sarvesh makes no SIL claim for its analysers.
- Explained in: Hazardous area gas analysers explained, What SIL means for gas analysers, and what it does not, Hydrogen electrolysers and gas analysis: what to measure
- Siloxane
- A silicon-containing organic compound found in landfill gas and some biogas that can damage sensors and engines.
- Span
- The difference between the top and bottom of an analyser's range. A span error scales readings; compare zero error.
- Explained in: Accuracy, precision, repeatability and drift in gas analysers, How gas analyser calibration works: zero and span
- Span drift
- A change over time in the scale of an analyser's response.
- Span gas
- Gas of known concentration near the top of the range, used to set the analyser's gain during calibration.
- Explained in: What is an NDIR gas analyser?, Online vs laboratory vs portable gas analysis, Sample system troubleshooting: symptoms to causes
- Stoichiometric
- Exactly the proportions of fuel and air needed for complete combustion.
- Explained in: Combustion, excess air and flue gas oxygen explained
- Stratification
- Layering of a gas stream so that composition or temperature differs across a duct: Choosing the sample point for a gas analyser.
- Explained in: Choosing the sample point for a gas analyser
- Syngas
- Synthesis gas, a mixture mainly of carbon monoxide and hydrogen made by gasification.
- Explained in: Steel plant gases: coke oven, blast furnace, converter
T
- TCD (thermal conductivity detector)
- A sensor that measures a gas by how well it carries heat away from a heated element.
- Explained in: Gas analysis technologies compared: which sensor for which gas, How thermal conductivity gas analysers (TCD) work, Measuring hydrogen: purity, mixtures and safety limits
- TDLAS
- Tunable diode laser absorption spectroscopy: measuring a gas by laser light absorbed on one narrow line. Not in the Sarvesh range.
- Explained in: Gas analysis technologies compared: which sensor for which gas
- Temperature class
- A marking, T1 to T6, giving the maximum surface temperature of hazardous-area equipment, from 450 C to 85 C.
- Explained in: Hazardous area classification: zones, gas groups, T-Classes
- Trace measurement
- Measurement at low concentration, typically ppm.
- Explained in: What is dew point and how is it measured?, Choosing an oxygen analyser: trace, percent or purity, ppm, percent and ppb explained for gas analysis
- Traceability
- The unbroken chain of comparisons linking a measurement to a national or international standard.
- Explained in: Calibration gases: choosing, handling and certificates, How often should you calibrate a gas analyser?, Records, audit trails and qualification for regulated plants
U
- UEL
- Upper explosive limit: the highest concentration of a flammable gas in air that can still burn. About 75 % for hydrogen.
- Explained in: Flammability, explosive limits and gas safety basics
- ULO (ultra-low oxygen)
- A controlled-atmosphere regime that holds oxygen at around 1 to 2 % or lower.
- Uncertainty
- A stated range within which the true value is expected to lie, given on calibration certificates.
V
- Validation
- Checking, with a known gas, that an analyser reads correctly, without necessarily adjusting it: Auto-Calibration and validation checks for gas analysers.
- Explained in: Records, audit trails and qualification for regulated plants
- Vaporiser
- A heated device that turns a liquid or cold sample into gas of the same composition.
- Explained in: Sampling cryogenic and liquefied gases: vaporisers, How an air separation plant works and what is measured
- Vent
- The outlet where a sample leaves an analyser, which must be safe and have steady low back pressure.
- Explained in: Sample pressure, flow and fast bypass loops, Site utilities checklist for gas analysers, Inerting and purging: how they work and how to prove them
- Volume fraction
- The share of a mixture's volume taken by a component. For gases, equal to mole fraction.
- Explained in: ppm, percent and ppb explained for gas analysis
W
- Warm-up time
- The time after power-up for a sensor or heater to stabilise before readings are reliable.
- Wet basis
- A gas value stated with water vapour counted as part of the gas: Dry basis vs wet basis in gas measurement.
- Explained in: Dry basis vs wet basis in gas measurement
Z
- Zeolite
- A porous mineral used as an adsorbent, for example in oxygen PSA generators, where it adsorbs nitrogen.
- Zero
- The analyser reading with no target gas present. A zero error shifts every reading equally.
- Explained in: Accuracy, precision, repeatability and drift in gas analysers, How gas analyser calibration works: zero and span
- Zero drift
- A change over time in the analyser's reading at zero gas.
- Zero gas
- Gas free of the target component, used to set the analyser's zero during calibration.
- Explained in: What is an NDIR gas analyser?, Sample system troubleshooting: symptoms to causes, Commissioning an online gas analyser: A Step-by-Step guide
- Zirconia sensor
- A heated ceramic sensor that measures oxygen from the voltage generated between two sides of a zirconia element.
- Explained in: How a zirconia oxygen analyser works, Electrochemical vs paramagnetic oxygen sensors, Gas analysis technologies compared: which sensor for which gas
- Zone 1 / Zone 2
- Hazardous-area zones where flammable gas may occur occasionally (Zone 1) or is not likely in normal operation (Zone 2).
- Explained in: Hazardous area gas analysers explained, Where to mount a gas analyser: panel, shelter or field, Hazardous area classification: zones, gas groups, T-Classes