ACCURRA-TCD · Nitrous oxide

Online nitrous oxide (N2O) analyser based on thermal conductivity

ACCURRA-TCD lists nitrous oxide among the gases it measures in binary mixtures, with a platinum sensor and no reference gas.

Thermal conductivity cellA sample gas flows through a cell containing a heated platinum element. The rate at which the gas carries heat away is the measurement signal.Platinum element, held at constant temperatureSample inSample outHeat carried away by the gas sets the signal
  • ISO 9001:2015, certified by TÜV
  • Designed and built in Pune since 2002
  • Supplied to Reliance, Tata Steel, Indian Oil, Linde and ISRO
Quick answer

ACCURRA-TCD lists nitrous oxide among the gases it measures in binary mixtures, by thermal conductivity with no reference gas. N2O in helium or hydrogen works well, but N2O with carbon dioxide or argon is poor, so send the composition first.

Updated 3 November 2026 · Figures from Sarvesh datasheets

Specifications

ItemSpecificationNotes
GasNitrous oxide in a binary mixture
SensorPlatinum thermal conductivity
RepeatabilityBelow 1 % of full scaleLinearity below 1 % of range
Errors and driftBelow 1 % of full scale per week
Response timeT90 below 10 s at 0.6 LPM
Reference gasNot required

From the current Sarvesh product datasheets and company profile. Specifications may change without notice as products improve. The final responsibility to check that a product meets the process and analyser requirement remains with the customer.

Key features

  • Binary N2O measurement
  • Isolated 4-20 mA, programmable alarms, Modbus RTU
  • No reference gas

How it works

Nitrous oxide conducts heat about as poorly as carbon dioxide and argon, much less than hydrogen or helium and somewhat less than nitrogen or oxygen. Nitrous oxide in helium or hydrogen gives a large signal. Nitrous oxide in nitrogen or oxygen gives a smaller but usable one.

Because carbon dioxide and argon are similar, a mixture containing N2O with either of them is difficult for any thermal conductivity analyser. Tell us the full composition. Every application of this analyser needs a binary mixture of known background.

Requirements and limits

  • Binary mixtures only; N2O with CO2 or argon is difficult.

On site: utilities and calibration gases

What to have ready at your plant before installation and commissioning.

ItemRequirementNotes
Power at the analyser100 to 240 V AC, 50/60 Hz, 45 W
Zero gasZero-grade nitrogen, 99.999 %10 litre cylinder with regulator
Span gasThe measured gas at zero grade, 99.999 %: hydrogen, helium, argon or nitrous oxide10 litre cylinder with regulator
Sample at the analyser2 to 10 psig, 0.6 LPM, 5 to 40 °C, dust below 3 micronPressure and flow regulated by the user
Mounting19 inch rack or panel, in a safe area

Selection guidance

Choose ACCURRA-TCD for N2O as the variable component of a binary mixture. Please enquire first, because the range, background and required accuracy decide whether the method fits.

Applications

Application pages where this is part of the recommended solution:

Compatible products

Frequently asked questions

Does it need a reference gas?

No.

Can it measure N2O in CO2?

Their conductivities are close, so please enquire.

Which backgrounds are listed?

Nitrogen or oxygen, on 0 to 100, 0 to 5 or 96 to 100 percent ranges.

Who should I contact?

Send the composition, range and accuracy you need and we will review.

Talk to an applications engineer