ACCURRA-Zr measures percent oxygen in flue gas in situ with a zirconia sensor, calibrated on plain air, with built-in automatic blowback. Options cover high dust loads and process temperatures up to 1400 C, for boilers, furnaces and stacks.
Why ACCURRA-Zr
A longer sensor life
Expected sensor life above 10 years on clean gas, above 5 years on low-sulphur natural gas and above 3 years on biomass or coal.
Calibration with air only
Single-point calibration uses air as the span gas, after an automatic blowback, so no special calibration gas is needed.
A clean probe and a steady reading
Programmable auto blowback with power-on and pre-calibration cycles; the output can hold its last value during blowback.
Specifications
| Item | Specification | Notes |
|---|---|---|
| Measurement | 0.1 to 21 % oxygen | Zirconia; no reference gas |
| Repeatability | Better than 1 % of FS | |
| Response time | T90 below 15 s | |
| Expected sensor life | Clean gas above 10 years; low-sulphur natural gas above 5 years; biomass or coal above 3 years | |
| Calibration | Single point with air | Blowback runs before every calibration |
| Probe | 300, 500 or 1000 mm; wetted parts SS316; 2 inch ANSI flange | Blowback connection 1/4 inch OD SS316 |
| Process | Gas 200 to 500 C; dust up to 10 g/Nm3; velocity 5 to 20 m/s | Probes for up to 1400 C and high dust load are optional |
| Fuels | Oil, gas and natural gas without blowback; coal and bagasse with blowback | |
| Electronics | Head mounted with display and keypad; backlit LCD 2 x 16 | Cast aluminium head, 285 x 165 x 230 mm, IP 66 |
| Outputs | Isolated 4-20 mA, dual range, max load 500 ohm; blowback relay; Modbus RTU over RS 485 | Read value, status and set-up; write blowback |
| Power and ambient | 100 to 240 V AC, 50/60 Hz, 45 W; ambient 10 to 50 C | Safe area |
From the current Sarvesh product datasheets and company profile. Specifications may change without notice as products improve. The final responsibility to check that a product meets the process and analyser requirement remains with the customer.
Key features
- Sensor located outside the stack or furnace, so it is protected from flow, dust and temperature variation; flue gas reaches it through a guide tube
- Single-point calibration with air, so no special calibration gas
- Built-in auto blowback with relay and solenoid valve, programmable from the keypad or Modbus; power-on, manual and pre-calibration blowback
- Dual-range 4-20 mA: a fixed 0 to 25 % range for start-up and a second user-set range
- Output freeze or follow during blowback and calibration
- Sensor and hardware diagnostics with sensor protection
- Rugged all-SS design; options for high dust load and up to 1400 C
How it works
A zirconia cell is a ceramic element that generates a voltage proportional to the difference in oxygen between the flue gas and a reference, usually air. In a conventional flue gas analyser the zirconia sensor, heater and thermocouple sit at the tip of the probe, exposed to furnace turbulence, temperature swings and abrasive dust, which causes reading errors and frequent sensor or heater failures.
ACCURRA-Zr places the sensor just outside the stack or furnace wall and draws the flue gas to it through a guide tube. The sensor stays at a constant temperature and away from the dust load, which gives a stable reading and a long sensor life. There is still no sample line to condense or block.
Oxygen in flue gas is the main measure of excess air. Controlling it down to the lowest safe level saves fuel. ACCURRA-Zr's air calibration and automatic blowback keep routine maintenance short in dusty ducts.
Requirements and limits
- For extractive measurement of other flue components, use SGI-320.
- Mount the probe horizontally. ACCURRA-Zr is not suitable where the flue gas flows from top to bottom.
On site: utilities and calibration gases
What to have ready at your plant before installation and commissioning.
| Item | Requirement | Notes |
|---|---|---|
| Power at the tapping point | 100 to 240 V AC, 50/60 Hz, 45 W | |
| Blowback air | Instrument air, up to 4 kg/cm² | Blowback solenoid valve supplied |
| Calibration | Air, single point | No calibration gas cylinder needed |
| Mounting | Horizontal, on a 2 inch ANSI flange with gasket, nuts and bolts | Not suitable where the flue gas flows from top to bottom |
| Process | Gas up to 500 °C; dust up to 10 g/Nm³; velocity 5 to 20 m/s |
Selection guidance
Choose ACCURRA-Zr for boilers, heaters, furnaces, HRSG and thermal oxidisers. For trace oxygen in process gas or protective atmospheres, choose ACCURRA-S, which uses an extractive sample.
Applications
Application pages where this is part of the recommended solution:
- Boiler and furnace oxygen (flue gas)
- Thermal oxidiser and scrubber efficiency
- Stack and emission monitoring
- Heat-treatment furnace atmosphere
Compatible products

SGI-300 / SGI-320
Peltier sample cooler with integrated pump, filter and condensate detector; SGI-320 for emission monitoring.
View SGI-300 / SGI-320i-Cons
Smart Peltier sample conditioner with diagnostics, auto blowback and Modbus, in a compact 19 inch cabinet.
View i-Cons
SP and SRP sample probes
Heated and non-heated probes with blowback, and retractable probes, in SS316, Inconel or ceramic.
View SP and SRP sample probes
ACCURRA-S
Online analyser for trace, percent and purity O2, CO, CO2, CH4 and dew point; can measure two gases from one sample.
View ACCURRA-SFrequently asked questions
How is ACCURRA-Zr calibrated?
With plain air at a single point, after an automatic blowback.
What gas temperature can it take?
The standard probe suits 200 to 500 C; probes for up to 1400 C are optional.
How long does the sensor last?
Above 10 years on clean gas, above 5 years on low-sulphur natural gas and above 3 years on biomass or coal.
Does it handle dust?
It has built-in automatic blowback and an option for high dust load samples.
Can it measure CO?
No. Use an extractive system for CO.
Related pages
- ACCURRA-S
- ACCURRA-S oxygen analysers
- ACCURRA-S carbon monoxide analysers
- ACCURRA-S carbon dioxide analysers
- ACCURRA-S methane analyser
- ACCURRA-S M015 moisture analyser
- ACCURRA-Hz
- Hazardous-area oxygen analyser
- INFINITY-Tx
- Hazardous-area hydrogen analyser
- Hazardous-area moisture analyser
- ACCURRA model numbers
- Browse all gas analysers