Sample conditioning · Safe area

i-Cons smart sample conditioner and SGI-320 emission system

A complete extractive sampling system in one 19-inch enclosure: Peltier cooler, sample and peristaltic pumps, fine filter, condensate and flow sensors, calibration valves and the controller that runs them. It warns before consumables run out and shuts down safely on a fault.

i-Cons and SGI-320i-Cons and SGI-320
  • ISO 9001:2015, certified by TÜV
  • Designed and built in Pune since 2002
  • Supplied to Reliance, Tata Steel, Indian Oil, Linde and ISRO
Quick answer

i-Cons is a Peltier sample conditioner in a 19-inch cabinet with pump, filter, condensate and flow sensors, blowback and heated-hose control, and Modbus. It suits samples at 120 to 550 C and up to 60 C dew point, excluding HCl, HF and H2S. SGI-320 adds the analysers.

Updated 2 November 2026 · Figures from Sarvesh datasheets

Why i-Cons and SGI-320

Less to design and wire

The components of a sampling panel come built in and pre-wired, which saves panel space and integration time.

Faults stopped before the analyser

The sample pump trips on condensate, blowback starts automatically on flow failure, and the system shuts down on a hardware fault.

Maintenance on schedule

Remaining life of filters, pump tubes and pump kits is tracked and reported over Modbus, with a reminder before replacement is due.

Specifications

ItemSpecificationNotes
CoolerPeltier, one PTFE heat exchangerSample flow up to 240 LPH (4 LPM); outlet 5 C
Sample into the coolerUp to 160 C; dew point up to 65 C; dust below 3 micron
Fine filter0.1 micron coalescingLong-life SS condensate sensor
PumpFree flow 5 LPM with fast loop valve; optional 8 LPM8 LPM for sample lines of 60 to 90 m
Heated hose30, 45, 60, 75 or 90 m; 40 W/m, 230 V AC; 1/4 inch OD PTFE tubeTemperature set by user, Pt100 feedback
InputsPt100 from the heated hose; digital inputs for calibration in progress, zero, span 1 and span 2Potential-free contacts from the analyser
OutputsTriac for the heated hose SSR or contactor; relay 1 alarm or fault; relay 2 probe blowback valveRelays SPST, 230 V AC 1 A
CommunicationModbus RTU over RS 485Read values, status, set-up and maintenance history; write set-up and operate calibration valves
DisplayBacklit LCD, 4 x 20, with 7 keys
Enclosure19 inch rack or panel, 310 x 483 x 285 mm, about 16 kgSafe area, IP 20; 230 V AC 50 Hz, 600 VA excluding heat tracer; ambient 10 to 40 C
Wetted materialsSS316, PTFE, Viton, silicone, engineering plastic
SGI-320Probe, heated line, conditioner and analysers as one systemNO, SO2, CO, CO2, O2; optional NOx converter and remote auto-calibration

From the current Sarvesh product datasheets and company profile. Specifications may change without notice as products improve. The final responsibility to check that a product meets the process and analyser requirement remains with the customer.

Key features

  • Plug-and-play complete sample handling system with menu-driven controller
  • Built-in sample pump, peristaltic pump with auto-drain, condensate sensor, flow detector, auto-cal valves and fine filter
  • Controls probe blowback, sample valve and heated hose; pulsating blowback; automatic blowback on flow failure
  • Pump trips on condensate detection, during calibration and on hardware fault
  • Fail-safe fault relay, alarm and fault history, maintenance reminders for filters, auto-drain tubes and pump
  • Digital input from the analyser stops the pump during calibration

How it works

Most extractive analyser faults begin in the sample path. i-Cons treats the path as one system. It runs the blowback, keeps the heated hose at temperature, cools and dries the gas with a Peltier cooler, and trips the pump the moment it sees condensate or a flow failure. It records usage of consumables and prompts for replacement before they fail.

SGI-320 is the same engineering delivered as a ready-to-use analyser system: a sample probe with auto-blowback, a heated line, the conditioning and the analysers, with optional NOx converter and remote auto-calibration. It is described in detail on the stack and emission monitoring page.

Requirements and limits

  • Not suitable for samples containing HCl, HF, H2S or other highly corrosive, toxic or flammable gases.
  • Suitable for samples where SO2, NOx and CO are below 1000 ppm.
  • Safe area only.

On site: utilities and calibration gases

What to have ready at your plant before installation and commissioning.

ItemRequirementNotes
Power230 V AC, 50 Hz, 600 VA for i-Cons, plus the analyserHeat tracing excluded
Heated hose and probeA separate power supply
Instrument air1/2 inch connection at 6 bar near the tapping pointFor probe blowback
LocationAn air-conditioned, dust-free room for the analyser and i-Cons
Calibration gasesZero and span gases for the analysers, in 10 litre cylinders with regulators

Options and ordering

  • Probe: SP-200 unheated for low dew point samples, SP-300 heated for high dew point samples, or SP-300 with blowback tank; insertion length 500, 1000 or 1500 mm
  • Prefabricated heated hose in 30, 45, 60, 75 or 90 m lengths
  • Internal auto-calibration solenoid valves, 24 V DC: zero and span, zero and two spans, or zero and three spans
  • Remote calibration check: Modbus commands switch the zero, span 1 and span 2 valves, as CPCB guidelines ask
  • Catchpot or water separator with an external peristaltic pump controlled by i-Cons; high-flow sample pump for 60 to 90 m lines

Selection guidance

Choose i-Cons for combustion, flue gas, process and emission analysis where the sample is hot, wet or dusty and you want fewer components and diagnostics. Choose SGI-300 for a cooler only. Choose SGI-320 for a complete multi-gas emission system.

Applications

Application pages where this is part of the recommended solution:

Compatible products

Frequently asked questions

What does i-Cons replace?

The usual sampling panel: cooler, pump, filter, condensate detector, flow detector and the control wiring between them.

How far can the sample line be?

Up to 60 m with the standard pump; the high-flow pump, 8 LPM free flow, supports 60 to 90 m.

Where can I not use it?

On samples containing HCl, HF or H2S, or other highly corrosive, toxic or flammable samples.

Is SGI-320 the same as i-Cons?

SGI-320 applies the i-Cons design to a complete emission monitoring system that includes the analysers.

Talk to an applications engineer

Send the gas, range, pressure, temperature and area classification.