ACCURRA-Hz · Hydrogen

Hazardous-area hydrogen analyser for purge and reaction safety

ACCURRA-Hz model 1030 measures hydrogen from 0 to 100 percent in nitrogen by thermal conductivity, in a flameproof enclosure for classified areas. Other background gases are available on request.

Hazardous-area hydrogen analyserHazardous-area hydrogen analyser
  • ISO 9001:2015, certified by TÜV
  • Designed and built in Pune since 2002
  • Supplied to Reliance, Tata Steel, Indian Oil, Linde and ISRO
Quick answer

ACCURRA-Hz measures percent or purity hydrogen in a thermal conductivity housing for classified areas, in a known background of nitrogen, air or oxygen. The enclosure is certified for Gas Group IIC, Zone 1 and 2, T6. No SIL claim is made.

Updated 3 November 2026 · Figures from Sarvesh datasheets

Specifications

ItemSpecificationNotes
MeasurementHydrogen 0 to 100 % in N2 (min 0 to 5 %)Thermal conductivity, model 1030
AccuracyBelow 1 % of FSRepeatability below 1 % of FS; T90 below 10 s at 0.6 LPM
BackgroundsNitrogen as standardOther backgrounds and ranges on request
EnclosureGas Group IIC, Zone 1 and 2, T6As per IS/IEC 60079-1:2014
CalibrationWithout opening covers

From the current Sarvesh product datasheets and company profile. Specifications may change without notice as products improve. The final responsibility to check that a product meets the process and analyser requirement remains with the customer.

Key features

  • Thermal conductivity hydrogen measurement in a certified enclosure
  • Alarm relays for purge permissives
  • Modbus and 4-20 mA outputs

How it works

Hydrogen is lighter than air and flammable over a wide range, so hydrogen areas are classified and every measurement in them must use certified equipment. Typical measurements are hydrogen in nitrogen at purge, hydrogen in oxygen in electrolysers, and hydrogen purity in reaction loops.

A thermal conductivity reading assumes a known binary mixture. Tell us the background and any variable third gas; we will choose the sensor and conditioning.

Requirements and limits

  • Confirm certificate scope for your classification; no SIL claim is made.

On site: utilities and calibration gases

What to have ready at your plant before installation and commissioning.

ItemRequirementNotes
Power at the analyser24 V DC, 30 W, or 230 V AC
Zero gasZero-grade nitrogen, 99.9999 %10 litre cylinder with regulator
Span gas, trace oxygen1000 ppm oxygen in nitrogen10 litre cylinder with regulator
Span gas, 3090Zero-grade oxygen; zero on zero-grade nitrogen
Span gas, 3190Zero-grade oxygen; zero on 95 % oxygen in nitrogen
Instrument air and service nitrogen4 kg/cm² minimum
MountingWall mounted in the classified areaCalibration without opening the covers

Selection guidance

Choose ACCURRA-Hz where hydrogen is measured inside the classified area. For safe-area hydrogen analysis see ACCURRA-TCD.

Applications

Application pages where this is part of the recommended solution:

Compatible products

Frequently asked questions

What backgrounds?

Nitrogen as standard; ask us for other background gases.

Does it need a reference gas?

No, it uses thermal conductivity.

Can I calibrate without opening the cover?

Yes.

Does it have a SIL rating?

We make no SIL claim.