Refinery and petrochemicals

Hydrogen purity analysers for refinery PSA product, make-up and recycle gas

Hydrotreaters and hydrocrackers live on hydrogen partial pressure, and the hydrogen comes from a reformer and PSA unit or from another unit's off-gas. ACCURRA-TCD measures hydrogen purity by thermal conductivity with no reference gas, on PSA product, make-up gas and recycle gas.

Quick answer

ACCURRA-TCD measures refinery hydrogen purity by thermal conductivity with no reference gas: hydrogen in methane 0 to 100 %, 0 to 3 % or 97 to 100 %, for PSA product, make-up and recycle gas. Heavier hydrocarbons and H2S are reviewed first; ACCURRA-Hz covers Zone 1.

Figures from Sarvesh datasheets

The measurement and why it matters

A hydroprocessing unit needs enough hydrogen at the reactor to keep the catalyst working and the coke down. The purity of the recycle gas, together with the system pressure, sets the hydrogen partial pressure, and the purity of the make-up gas sets how much purge is needed to hold it. Operators watch both.

Methane is the main impurity in most refinery hydrogen, and hydrogen in methane is a pair that thermal conductivity reads well: hydrogen conducts heat far better than methane. ACCURRA-TCD reads hydrogen in methane over 0 to 100 %, 0 to 3 % or 97 to 100 %. The 97 to 100 % range suits PSA product hydrogen, which is close to a clean pair.

Recycle gas is less tidy. It carries ethane, propane and heavier hydrocarbons, and H2S that the amine scrubber has not removed. Each of these moves a thermal conductivity reading by a different amount, so we work out the error from your composition before we quote, and say where the reading would not be good enough for its purpose.

Hydrogen sample points in a refinery are almost always in a classified area. ACCURRA-TCD stands in a safe-area analyser shelter or room with the sample piped to it; ACCURRA-Hz houses hydrogen by thermal conductivity in an enclosure certified for Gas Group IIC, Zone 1 and 2, T6.

Models and ranges

ModelMeasuresRangeSensor and notes
ACCURRA-TCDPSA product hydrogen97 to 100 % hydrogen in methanePlatinum thermal conductivity, no reference gas
ACCURRA-TCDMake-up and recycle gas, by review0 to 100 % hydrogen in methaneComposition reviewed first
ACCURRA-Hz TCD housingHydrogen, hazardous areaApplication specificGas Group IIC, Zone 1 and 2, T6

Process notes

PSA product hydrogen

The PSA unit's product is the cleanest hydrogen in the refinery, and a continuous purity reading shows when an adsorber bed is breaking through. The narrow 97 to 100 % range gives the best resolution where the operator looks.

ACCURRA-TCD has a T90 below 10 seconds at 0.6 LPM, repeatability below 1 % of full scale and drift below 1 % of full scale per week.

Recycle gas

On recycle gas, send the full composition with typical and extreme values, including C2 and heavier hydrocarbons and H2S. We say what the reading error will be and whether thermal conductivity suits the purpose.

Sampling and installation

  • Provide about 0.6 LPM of sample at a regulated 2 to 10 psig. The sample should reach ACCURRA-TCD at 5 to 45 °C, with its dew point at least 5 °C below the lowest ambient.
  • Keep the sample above its hydrocarbon dew point, or knock out liquid with a Headline coalescing filter and drain before the analyser.
  • Tell us the H2S level: it decides the wetted materials and how the vent is handled.
  • Vent the analyser outlet to the flare or a safe place agreed with the plant.

Send us your process data

Application data sheet

The same pre-order design inputs our engineers design from, online. Fill in what you know, leave the rest. Your draft saves as you go, and you can print or send it when ready.

Start the data sheet

What matters most for this application

  • The sample point: PSA product, make-up or recycle gas
  • The full composition with typical and extreme values, including heavier hydrocarbons and H2S
  • Pressure and temperature at the tapping point, and the area classification where the analyser will stand

Ask for a price

Leave your details and an applications engineer in Pune, India, replies with questions or a proposal.

We use these details only to reply to your enquiry. Privacy notice.

Frequently asked questions

Can thermal conductivity measure refinery hydrogen purity?

Yes, where the gas is close to a two-gas mixture such as hydrogen in methane. For recycle gas with heavier hydrocarbons we work out the error first.

What range suits PSA product hydrogen?

97 to 100 % hydrogen in methane.

Does the analyser need a reference gas?

No. ACCURRA-TCD uses a platinum thermal conductivity sensor with no reference gas.

Can it stand in Zone 1?

ACCURRA-TCD is a safe-area instrument; ACCURRA-Hz is certified for Gas Group IIC, Zone 1 and 2, T6.