ACCURRA-TCD measures refinery hydrogen purity by thermal conductivity with no reference gas: hydrogen in methane 0 to 100 %, 0 to 3 % or 97 to 100 %, for PSA product, make-up and recycle gas. Heavier hydrocarbons and H2S are reviewed first; ACCURRA-Hz covers Zone 1.
The measurement and why it matters
A hydroprocessing unit needs enough hydrogen at the reactor to keep the catalyst working and the coke down. The purity of the recycle gas, together with the system pressure, sets the hydrogen partial pressure, and the purity of the make-up gas sets how much purge is needed to hold it. Operators watch both.
Methane is the main impurity in most refinery hydrogen, and hydrogen in methane is a pair that thermal conductivity reads well: hydrogen conducts heat far better than methane. ACCURRA-TCD reads hydrogen in methane over 0 to 100 %, 0 to 3 % or 97 to 100 %. The 97 to 100 % range suits PSA product hydrogen, which is close to a clean pair.
Recycle gas is less tidy. It carries ethane, propane and heavier hydrocarbons, and H2S that the amine scrubber has not removed. Each of these moves a thermal conductivity reading by a different amount, so we work out the error from your composition before we quote, and say where the reading would not be good enough for its purpose.
Hydrogen sample points in a refinery are almost always in a classified area. ACCURRA-TCD stands in a safe-area analyser shelter or room with the sample piped to it; ACCURRA-Hz houses hydrogen by thermal conductivity in an enclosure certified for Gas Group IIC, Zone 1 and 2, T6.
Models and ranges
| Model | Measures | Range | Sensor and notes |
|---|---|---|---|
| ACCURRA-TCD | PSA product hydrogen | 97 to 100 % hydrogen in methane | Platinum thermal conductivity, no reference gas |
| ACCURRA-TCD | Make-up and recycle gas, by review | 0 to 100 % hydrogen in methane | Composition reviewed first |
| ACCURRA-Hz TCD housing | Hydrogen, hazardous area | Application specific | Gas Group IIC, Zone 1 and 2, T6 |
Process notes
PSA product hydrogen
The PSA unit's product is the cleanest hydrogen in the refinery, and a continuous purity reading shows when an adsorber bed is breaking through. The narrow 97 to 100 % range gives the best resolution where the operator looks.
ACCURRA-TCD has a T90 below 10 seconds at 0.6 LPM, repeatability below 1 % of full scale and drift below 1 % of full scale per week.
Recycle gas
On recycle gas, send the full composition with typical and extreme values, including C2 and heavier hydrocarbons and H2S. We say what the reading error will be and whether thermal conductivity suits the purpose.
Sampling and installation
- Provide about 0.6 LPM of sample at a regulated 2 to 10 psig. The sample should reach ACCURRA-TCD at 5 to 45 °C, with its dew point at least 5 °C below the lowest ambient.
- Keep the sample above its hydrocarbon dew point, or knock out liquid with a Headline coalescing filter and drain before the analyser.
- Tell us the H2S level: it decides the wetted materials and how the vent is handled.
- Vent the analyser outlet to the flare or a safe place agreed with the plant.
Send us your process data
The same pre-order design inputs our engineers design from, online. Fill in what you know, leave the rest. Your draft saves as you go, and you can print or send it when ready.
What matters most for this application
- The sample point: PSA product, make-up or recycle gas
- The full composition with typical and extreme values, including heavier hydrocarbons and H2S
- Pressure and temperature at the tapping point, and the area classification where the analyser will stand
Recommended Sarvesh products

ACCURRA-TCD
Hydrogen in methane, with no reference gas.
View ACCURRA-TCD
ACCURRA-Hz
Hydrogen measurement in Zone 1 or 2.
View ACCURRA-HzEngineered analyser panels
Analyser panels with knock-out, filtration and a safe vent.
View Engineered analyser panels
Coalescing filters for hydrocarbon liquids.
View the Headline rangeAsk for a price
Leave your details and an applications engineer in Pune, India, replies with questions or a proposal.
Frequently asked questions
Can thermal conductivity measure refinery hydrogen purity?
Yes, where the gas is close to a two-gas mixture such as hydrogen in methane. For recycle gas with heavier hydrocarbons we work out the error first.
What range suits PSA product hydrogen?
97 to 100 % hydrogen in methane.
Does the analyser need a reference gas?
No. ACCURRA-TCD uses a platinum thermal conductivity sensor with no reference gas.
Can it stand in Zone 1?
ACCURRA-TCD is a safe-area instrument; ACCURRA-Hz is certified for Gas Group IIC, Zone 1 and 2, T6.