A catch pot (add CP to the housing code) collects slugs of liquid before a filter; a bubbler (BB) washes a sample or adds a vapour; a drain vessel empties a filter under vacuum; the FD130 float drain empties one automatically between 1 and 15 bar; and the 910 to 920 aluminium inlet filters protect the open end of a sample line.
Catch pots
In a sample system with a lot of condensed liquid a filter can flood even when it is serviced often. A catch pot fitted just before the filter collects the liquid first. Many Headline housings can be supplied as catch pots; those with transparent bowls, such as the 137G, 751N or 780PC, show the liquid level at a glance.
To order: add CP to the housing code. 137G-CP (stainless steel, 7 bar, 100 °C) and 750N-CP (nylon) are in Headline's current catalogue.
Bubblers
A bubbler passes the gas sample through a liquid: to wash out an unwanted component, or to add a vapour, for example to keep a minimum relative humidity. Housings with transparent bowls, such as the 751N or 780PC, let you watch the liquid level.
To order: add BB to the housing code.
From Headline's instrumentation brochure; ask us for current availability.
Drain vessels DV127G, DV137G and DV147G
For filters under vacuum, where liquid cannot drain against the pressure. The vessel collects the liquid from the filter's drain port; valves at each end isolate it from the system so it can be emptied.
To order: DV127G, DV137G, DV147G.
| DV127G | DV137G | DV147G | |
|---|---|---|---|
| Port | 1/8" NPT | 1/4" NPT | 1/4" NPT |
| Volume | 36 cc | 105 cc | 255 cc |
| Rating | 7 bar, 100 °C | 7 bar, 100 °C | 7 bar, 100 °C |
| Materials | 316L, borosilicate glass bowl, Viton | 316L, borosilicate glass bowl, Viton | 316L, borosilicate glass bowl, Viton |
From Headline's instrumentation brochure; ask us for current availability.
Automatic liquid shut-off vessels 365N-V and 750N-V
A catch pot that also closes the line. Liquid collects in the clear nylon bowl and lifts a float; at the shut-off level the float's cone closes the outlet, so no more sample reaches the analyser until the cause has been found and the bowl emptied. It is fully mechanical and needs no power or control system.
To order: 365N-V (1/2" NPT) or 750N-V (1/4" NPT).
| 365N-V | 750N-V | |
|---|---|---|
| Ports | 1/2" NPT | 1/4" NPT |
| Head | Aluminium | Nylon |
| Bowl | Clear nylon | Clear nylon |
| Float and cone | PVC | PVC |
| Drain | Manual brass drain | None: depressurise and empty |
FD130 automatic float drain
Fits the drain port of a housing with a coalescing element, in a system under positive pressure. Liquid collects in the drain; as the level rises a float opens a valve and the liquid drains out. It works between 1 and 15 bar.
To order: FD130.
| FD130 | |
|---|---|
| Port | 1/4" NPT |
| Volume | 30 cc |
| Rating | 15 bar, 100 °C |
| Materials | 316L, Viton |
Aluminium inlet filters 910, 915 and 920
Single-port filters for the open end of a sample line, widely used in environmental monitoring. The open versions have a 6.3 mm spigot; the enclosed E versions have a cover and an NPT port, for places where the filter could be knocked. Rated to 150 °C. They take 12-32, 12-57 or 25-64 glass fibre elements, or stainless steel or PTFE elements of the same size.
To order: 910, 915, 920 (open); 910E, 915E, 920E (enclosed).
| Model | 910 / 910E | 915 / 915E | 920 / 920E |
|---|---|---|---|
| Port | ø6.3 mm / 1/8" NPT | ø6.3 mm / 1/4" NPT | ø6.3 mm / 1/2" NPT |
| Element | 12-32 | 12-57 | 25-64 |
From Headline's instrumentation brochure; ask us for current availability.
Selection checklist, port configurations and flow tables for every element size and grade.
The main Headline catalogue for sample systems: bonded microfibre, stainless steel and PTFE elements in industry-standard sizes and grades, and the stainless steel, PTFE, PVDF, plastic and aluminium housings that hold them, with dimensions and order codes.
Frequently asked questions
How do I order a catch pot or a bubbler?
Add CP (catch pot) or BB (bubbler) to the Headline housing code, for example 137G-CP. Housings with transparent bowls let you watch the liquid level.
Drain vessel or float drain?
A float drain (FD130) empties itself but needs a positive pressure of 1 to 15 bar. Under vacuum, liquid cannot drain against the pressure, so a drain vessel collects it and is emptied by hand between its isolating valves.
Can I buy these in India?
Sarvesh Analytics in Pune, India, is the exclusive representative of Headline Filters for the Indian subcontinent. Ask us for price and current availability.