ACCURRA-S-TCD · Model 7130

ACCURRA-S-TCD 7130 helium analyser

ACCURRA-S-TCD 7130 measures helium in a binary gas mixture by thermal conductivity, with a platinum sensor that needs no sealed reference and no reference gas. Made by Sarvesh in Pune, India.

Quick answer

ACCURRA-S-TCD 7130 measures helium by thermal conductivity in nitrogen or air and argon; repeatability below 1 % of full scale, T90 below 10 s at 0.6 LPM, no reference gas. Ordering code ACCURRA-S-TCD-7130.

Figures from the ACCURRA-S-TCD datasheet (SAPL/PDS/ACCURRA-TCD-08/25-Aug)

Specifications

ItemSpecificationNotes
MeasuresHelium in a binary gas mixtureThermal conductivity
In nitrogen or air0 to 100 %, 0 to 3 % or 97 to 100 %Range options
In argon0 to 100 %, 0 to 3 % or 97 to 100 %Range options
SensorPlatinum thermal conductivity; no sealed reference and no reference gas flow
Response timeT90 below 10 s at 0.6 LPM
RepeatabilityBelow 1 % of full scale
LinearityBelow 1 % of range
Errors and driftBelow 1 % of full scale per week
Display resolution0.01 %
Sample flow0.6 LPM recommendedRegulated externally; built-in flow indicator

From datasheet SAPL/PDS/ACCURRA-TCD-08/25-Aug. Specifications may change as products improve.

Built into every ACCURRA-S-TCD

ItemSpecificationNotes
DisplayBacklit LCD, 4 lines x 20 characters
Analog output1 x 4–20 mA isolated, max load 500 ohm, dual rangeRange freely selectable over the full measurement range
Relays6 x SPST rated 1 A 230 V AC, all failsafeAlarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves
CommunicationModbus RTU over RS 485Read measured value, status, set-up and history; write auto-calibration start and set-up
Enclosure19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mmIP 20, safe area only; net about 6 kg, gross about 8 kg
Power100 to 240 V AC, 50/60 Hz, 45 W
Inlet and outlet1/4 inch OD SS316
Ambient5 to 40 °C; below 90 % RH non-condensingStorage 0 to 50 °C

Options for this model

  • Auto calibration: available, with internal solenoid valves or as a relay output for external valves; a dual measurement has relay outputs only.
  • Sample pump: an internal sample pump is available on a single measurement; not on a dual measurement.
  • Second measurement: trace or percent oxygen, ppm or percent carbon dioxide, or dew point, in the same enclosure.

Build the part number Write the specification

Where it is used

Application notes where 7130 is part of the recommended system:

Frequently asked questions

Which background gases does 7130 work in?

Nitrogen or air: 0 to 100 %, 0 to 3 % or 97 to 100 %; Argon: 0 to 100 %, 0 to 3 % or 97 to 100 %. Ask us for other mixtures.

Does it need a reference gas?

No. The platinum thermal conductivity sensor needs no sealed reference and no reference gas flow.

What is the part number?

It starts ACCURRA-S-TCD-7130, then a second component or XXXX, the power supply (A), the range ID or pump relay, and auto calibration. The part number builder writes the full code.