ACCURRA-S-TCD 7230 measures argon by thermal conductivity in nitrogen, air or oxygen, carbon dioxide and xenon; repeatability below 1 % of full scale, T90 below 10 s at 0.6 LPM, no reference gas. Ordering code ACCURRA-S-TCD-7230.
Specifications
| Item | Specification | Notes |
|---|---|---|
| Measures | Argon in a binary gas mixture | Thermal conductivity |
| In nitrogen, air or oxygen | 0 to 100 %, 0 to 5 % or 96 to 100 % | Range options |
| In carbon dioxide | 40 to 100 % or 80 to 100 % | Range options |
| In xenon | 0 to 100 %, 0 to 5 % or 97 to 100 % | Range options |
| Sensor | Platinum thermal conductivity; no sealed reference and no reference gas flow | |
| Response time | T90 below 10 s at 0.6 LPM | |
| Repeatability | Below 1 % of full scale | |
| Linearity | Below 1 % of range | |
| Errors and drift | Below 1 % of full scale per week | |
| Display resolution | 0.01 % | |
| Sample flow | 0.6 LPM recommended | Regulated externally; built-in flow indicator |
From datasheet SAPL/PDS/ACCURRA-TCD-08/25-Aug. Specifications may change as products improve.
Built into every ACCURRA-S-TCD
| Item | Specification | Notes |
|---|---|---|
| Display | Backlit LCD, 4 lines x 20 characters | |
| Analog output | 1 x 4–20 mA isolated, max load 500 ohm, dual range | Range freely selectable over the full measurement range |
| Relays | 6 x SPST rated 1 A 230 V AC, all failsafe | Alarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves |
| Communication | Modbus RTU over RS 485 | Read measured value, status, set-up and history; write auto-calibration start and set-up |
| Enclosure | 19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mm | IP 20, safe area only; net about 6 kg, gross about 8 kg |
| Power | 100 to 240 V AC, 50/60 Hz, 45 W | |
| Inlet and outlet | 1/4 inch OD SS316 | |
| Ambient | 5 to 40 °C; below 90 % RH non-condensing | Storage 0 to 50 °C |
Options for this model
- Auto calibration: available, with internal solenoid valves or as a relay output for external valves; a dual measurement has relay outputs only.
- Sample pump: an internal sample pump is available on a single measurement; not on a dual measurement.
- Second measurement: trace or percent oxygen, ppm or percent carbon dioxide, or dew point, in the same enclosure.
Build the part number Write the specification
Where it is used
Application notes where 7230 is part of the recommended system:
Frequently asked questions
Which background gases does 7230 work in?
Nitrogen, air or oxygen: 0 to 100 %, 0 to 5 % or 96 to 100 %; Carbon dioxide: 40 to 100 % or 80 to 100 %; Xenon: 0 to 100 %, 0 to 5 % or 97 to 100 %. Ask us for other mixtures.
Does it need a reference gas?
No. The platinum thermal conductivity sensor needs no sealed reference and no reference gas flow.
What is the part number?
It starts ACCURRA-S-TCD-7230, then a second component or XXXX, the power supply (A), the range ID or pump relay, and auto calibration. The part number builder writes the full code.