ACCURRA-S-TCD · Model 7330

ACCURRA-S-TCD 7330 nitrogen analyser

ACCURRA-S-TCD 7330 measures nitrogen in a binary gas mixture by thermal conductivity, with a platinum sensor that needs no sealed reference and no reference gas. Made by Sarvesh in Pune, India.

Quick answer

ACCURRA-S-TCD 7330 measures nitrogen by thermal conductivity in argon, carbon dioxide and hydrogen; repeatability below 1 % of full scale, T90 below 10 s at 0.6 LPM, no reference gas. Ordering code ACCURRA-S-TCD-7330.

Figures from the ACCURRA-S-TCD datasheet (SAPL/PDS/ACCURRA-TCD-08/25-Aug)

Specifications

ItemSpecificationNotes
MeasuresNitrogen in a binary gas mixtureThermal conductivity
In argon0 to 100 %, 0 to 5 % or 97 to 100 %Range options
In carbon dioxide0 to 100 %, 0 to 5 % or 96 to 100 %Range options
In hydrogen0 to 100 %, 0 to 5 % or 97 to 100 %Range options
SensorPlatinum thermal conductivity; no sealed reference and no reference gas flow
Response timeT90 below 10 s at 0.6 LPM
RepeatabilityBelow 1 % of full scale
LinearityBelow 1 % of range
Errors and driftBelow 1 % of full scale per week
Display resolution0.01 %
Sample flow0.6 LPM recommendedRegulated externally; built-in flow indicator

From datasheet SAPL/PDS/ACCURRA-TCD-08/25-Aug. Specifications may change as products improve.

Built into every ACCURRA-S-TCD

ItemSpecificationNotes
DisplayBacklit LCD, 4 lines x 20 characters
Analog output1 x 4–20 mA isolated, max load 500 ohm, dual rangeRange freely selectable over the full measurement range
Relays6 x SPST rated 1 A 230 V AC, all failsafeAlarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves
CommunicationModbus RTU over RS 485Read measured value, status, set-up and history; write auto-calibration start and set-up
Enclosure19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mmIP 20, safe area only; net about 6 kg, gross about 8 kg
Power100 to 240 V AC, 50/60 Hz, 45 W
Inlet and outlet1/4 inch OD SS316
Ambient5 to 40 °C; below 90 % RH non-condensingStorage 0 to 50 °C

Options for this model

  • Auto calibration: available, with internal solenoid valves or as a relay output for external valves; a dual measurement has relay outputs only.
  • Sample pump: an internal sample pump is available on a single measurement; not on a dual measurement.
  • Second measurement: trace or percent oxygen, ppm or percent carbon dioxide, or dew point, in the same enclosure.

Build the part number Write the specification

Dual measurement

The catalogue lists nitrogen with 3015 (oxygen, trace) in one ACCURRA-S-TCD, each with its own electronics, 4–20 mA output, relays and Modbus address.

Where it is used

Application notes where 7330 is part of the recommended system:

Frequently asked questions

Which background gases does 7330 work in?

Argon: 0 to 100 %, 0 to 5 % or 97 to 100 %; Carbon dioxide: 0 to 100 %, 0 to 5 % or 96 to 100 %; Hydrogen: 0 to 100 %, 0 to 5 % or 97 to 100 %. Ask us for other mixtures.

Does it need a reference gas?

No. The platinum thermal conductivity sensor needs no sealed reference and no reference gas flow.

What is the part number?

It starts ACCURRA-S-TCD-7330, then a second component or XXXX, the power supply (A), the range ID or pump relay, and auto calibration. The part number builder writes the full code.