ACCURRA-S-TCD 1030 measures hydrogen by thermal conductivity in nitrogen, air, CO2 or methane, argon, helium and oxygen; repeatability below 1 % of full scale, T90 below 10 s at 0.6 LPM, no reference gas. Ordering code ACCURRA-S-TCD-1030.
Specifications
| Item | Specification | Notes |
|---|---|---|
| Measures | Hydrogen in a binary gas mixture | Thermal conductivity |
| In nitrogen, air, CO2 or methane | 0 to 100 %, 0 to 3 % or 97 to 100 % | Range options |
| In argon | 0 to 100 %, 0 to 3 % or 97 to 100 % | Range options |
| In helium | 20 to 100 %, 20 to 40 % or 85 to 100 % | Range options |
| In oxygen | 0 to 100 %, 0 to 4 % or 97 to 100 % | Range options |
| Sensor | Platinum thermal conductivity; no sealed reference and no reference gas flow | |
| Response time | T90 below 10 s at 0.6 LPM | |
| Repeatability | Below 1 % of full scale | |
| Linearity | Below 1 % of range | |
| Errors and drift | Below 1 % of full scale per week | |
| Display resolution | 0.01 % | |
| Sample flow | 0.6 LPM recommended | Regulated externally; built-in flow indicator |
From datasheet SAPL/PDS/ACCURRA-TCD-08/25-Aug. Specifications may change as products improve.
Built into every ACCURRA-S-TCD
| Item | Specification | Notes |
|---|---|---|
| Display | Backlit LCD, 4 lines x 20 characters | |
| Analog output | 1 x 4–20 mA isolated, max load 500 ohm, dual range | Range freely selectable over the full measurement range |
| Relays | 6 x SPST rated 1 A 230 V AC, all failsafe | Alarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves |
| Communication | Modbus RTU over RS 485 | Read measured value, status, set-up and history; write auto-calibration start and set-up |
| Enclosure | 19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mm | IP 20, safe area only; net about 6 kg, gross about 8 kg |
| Power | 100 to 240 V AC, 50/60 Hz, 45 W | |
| Inlet and outlet | 1/4 inch OD SS316 | |
| Ambient | 5 to 40 °C; below 90 % RH non-condensing | Storage 0 to 50 °C |
Options for this model
- Auto calibration: available, with internal solenoid valves or as a relay output for external valves; a dual measurement has relay outputs only.
- Sample pump: an internal sample pump is available on a single measurement; not on a dual measurement.
- Second measurement: trace or percent oxygen, ppm or percent carbon dioxide, or dew point, in the same enclosure.
Build the part number Write the specification
Dual measurement
The catalogue lists hydrogen with 3015 (oxygen, trace) in one ACCURRA-S-TCD, each with its own electronics, 4–20 mA output, relays and Modbus address.
For hazardous areas
Hydrogen by thermal conductivity is also available in ACCURRA-Hz (model 1030, 0 to 100 % in nitrogen), in a certified enclosure for Gas Group IIC, Zone 1 and 2, T6. See hazardous-area hydrogen.
Where it is used
Application notes where 1030 is part of the recommended system:
- Hydrogen purity
- Electrolyser safety and quality
- Heat-treatment furnace atmosphere
- Steel plant process gases
- Float glass bath atmosphere
- Reactors, hydrogenation and polymerisation
- Pyrolysis, gasification and syngas
- Chlor-alkali hydrogen stream: oxygen in hydrogen
- Hydrogen-cooled generator purity and purge
- Ammonia synthesis loop and purge gas hydrogen
- Urea plant CO2 feed: purity and passivation oxygen
- Refinery hydrogen purity: make-up, recycle and PSA
Frequently asked questions
Which background gases does 1030 work in?
Nitrogen, air, CO2 or methane: 0 to 100 %, 0 to 3 % or 97 to 100 %; Argon: 0 to 100 %, 0 to 3 % or 97 to 100 %; Helium: 20 to 100 %, 20 to 40 % or 85 to 100 %; Oxygen: 0 to 100 %, 0 to 4 % or 97 to 100 %. Ask us for other mixtures.
Does it need a reference gas?
No. The platinum thermal conductivity sensor needs no sealed reference and no reference gas flow.
What is the part number?
It starts ACCURRA-S-TCD-1030, then a second component or XXXX, the power supply (A), the range ID or pump relay, and auto calibration. The part number builder writes the full code.