ACCURRA-S · Model 3190

ACCURRA-S 3190 paramagnetic oxygen purity analyser

ACCURRA-S 3190 measures oxygen purity over 95 to 100 % with a paramagnetic sensor. It is a safe-area, 19-inch rack or panel analyser with an isolated 4–20 mA output, relays and Modbus, made by Sarvesh in Pune, India.

Quick answer

ACCURRA-S 3190 measures oxygen purity from 95 to 100 %, with a paramagnetic sensor: accuracy +/- 0.5 % of FS, response below 5 s. Ordering code ACCURRA-S-3190 followed by the second component and options.

Figures from the ACCURRA-S datasheet (SAPL/PDS/ACCURRA-S-10/25-Aug)

Specifications

ItemSpecificationNotes
MeasuresOxygen, purity
Range95 to 100 %Configurable within this span; dual-range 4–20 mA output
Minimum rangeNot applicable
SensorParamagnetic, ambient pressure and temperature compensated
Display resolution0.01 %
Intrinsic error (accuracy)+/- 0.5 % of FS
Repeatability+/- 0.5 % of FS
Response T90Below 5 sAt 1 LPM unless stated
NotesNo consumable sensor; no internal sample pump

From datasheet SAPL/PDS/ACCURRA-S-10/25-Aug. Specifications may change as products improve.

Built into every ACCURRA-S

ItemSpecificationNotes
DisplayBacklit LCD, 4 lines x 20 characters
Analog output1 x 4–20 mA isolated, max load 500 ohm, dual rangeRange freely selectable over the full measurement range
Relays6 x SPST rated 1 A 230 V AC, all failsafeAlarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves
CommunicationModbus RTU over RS 485Read measured value, status, set-up and history; write auto-calibration start and set-up
Enclosure19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mmIP 20, safe area only; net about 6 kg, gross about 8 kg
Power100 to 240 V AC, 50/60 Hz, 45 W
Inlet and outlet1/4 inch OD SS316
Ambient5 to 40 °C; below 90 % RH non-condensingStorage 0 to 50 °C

Options for this model

  • Auto calibration: available, with internal solenoid valves or as a relay output for external valves.
  • Sample pump: no internal sample pump on this model; a relay output can drive an external pump.
  • Range ID relay: shows which of the two 4–20 mA ranges is active; it shares a relay with the sample pump, so choose one.

Build the part number Write the specification

For hazardous areas

The same 3190 sensor is available in ACCURRA-Hz, in a certified enclosure (Gas Group IIC, Zone 1 and 2, T6), wall mounted, with 24 V DC or AC power. Calibration is done without opening the covers. See ACCURRA-Hz.

Sample at the analyser

  • The sample gas should be non-corrosive, non-toxic, non-flammable, non-condensing and dry, and free from entrained oil.
  • Sample pressure 2 to 10 psig maximum, regulated externally. Recommended flow 1 LPM.
  • Dust or particulate below 3 micron; the built-in fine filter handles light loads, and heavier dust needs an upstream filter.
  • Materials in contact with sample: SS316, Viton, PTFE, glass, aluminium and acrylic.
  • Sample temperature 5 to 45 °C, and a sample dew point at least 5 °C below the lowest ambient temperature.

Where it is used

Application notes where 3190 is part of the recommended system:

Frequently asked questions

What range does ACCURRA-S 3190 measure?

95 to 100 %.

Does 3190 have auto calibration and an internal sample pump?

Auto calibration: available, with internal solenoid valves or as a relay output for external valves. Sample pump: no internal sample pump on this model; a relay output can drive an external pump.

What is the part number?

It starts ACCURRA-S-3190. Positions 5 to 8 are a second gas or XXXX, then the power supply, the range ID or pump relay, and auto calibration. The part number builder writes the full code and checks it against the catalogue.

Is 3190 available for hazardous areas?

Yes, in ACCURRA-Hz, in a certified enclosure for Gas Group IIC, Zone 1 and 2, T6.