ACCURRA-S · Model 6015

ACCURRA-S 6015 ppm carbon dioxide analyser

ACCURRA-S 6015 measures carbon dioxide in ppm over 0 to 5000 ppm (minimum range 0 to 100 ppm) with a ndir sensor. It is a safe-area, 19-inch rack or panel analyser with an isolated 4–20 mA output, relays and Modbus, made by Sarvesh in Pune, India.

Quick answer

ACCURRA-S 6015 measures carbon dioxide in ppm from 0 to 5000 ppm (minimum range 0 to 100 ppm), with a ndir sensor: accuracy 1 % of FS, response below 60 s. Ordering code ACCURRA-S-6015 followed by the second component and options.

Figures from the ACCURRA-S datasheet (SAPL/PDS/ACCURRA-S-10/25-Aug)

Specifications

ItemSpecificationNotes
MeasuresCarbon dioxide
Range0 to 5000 ppmConfigurable within this span; dual-range 4–20 mA output
Minimum range0 to 100 ppm
SensorNDIR
Display resolution0.1 ppm
Intrinsic error (accuracy)1 % of FS
RepeatabilityZero +/- 10 ppm; span +/- 50 ppm
Response T90Below 60 sAt 1 LPM unless stated
NotesSensor life above 5 years

From datasheet SAPL/PDS/ACCURRA-S-10/25-Aug. Specifications may change as products improve.

Built into every ACCURRA-S

ItemSpecificationNotes
DisplayBacklit LCD, 4 lines x 20 characters
Analog output1 x 4–20 mA isolated, max load 500 ohm, dual rangeRange freely selectable over the full measurement range
Relays6 x SPST rated 1 A 230 V AC, all failsafeAlarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves
CommunicationModbus RTU over RS 485Read measured value, status, set-up and history; write auto-calibration start and set-up
Enclosure19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mmIP 20, safe area only; net about 6 kg, gross about 8 kg
Power100 to 240 V AC, 50/60 Hz, 45 W
Inlet and outlet1/4 inch OD SS316
Ambient5 to 40 °C; below 90 % RH non-condensingStorage 0 to 50 °C

Options for this model

  • Auto calibration: available, with internal solenoid valves or as a relay output for external valves.
  • Sample pump: an internal sample pump is available on a single measurement, or a relay output drives an external pump.
  • Range ID relay: shows which of the two 4–20 mA ranges is active; it shares a relay with the sample pump, so choose one.

Build the part number Write the specification

Dual measurement

The 6015 can share an ACCURRA-S enclosure with a second sensor in series, each with its own electronics, 4–20 mA output, relays and Modbus address. The catalogue lists it with 2015 (carbon monoxide) and 3015 (oxygen, trace). Other pairs are possible on request.

Sample at the analyser

  • The sample gas should be non-corrosive, non-toxic, non-flammable, non-condensing and dry, and free from entrained oil.
  • Sample pressure 2 to 10 psig maximum, regulated externally. Recommended flow 1 LPM.
  • Dust or particulate below 3 micron; the built-in fine filter handles light loads, and heavier dust needs an upstream filter.
  • Materials in contact with sample: SS316, Viton, PTFE, glass, aluminium and acrylic.
  • Sample temperature 5 to 45 °C, and a sample dew point at least 5 °C below the lowest ambient temperature.

Where it is used

Application notes where 6015 is part of the recommended system:

Frequently asked questions

What range does ACCURRA-S 6015 measure?

0 to 5000 ppm, with a minimum range of 0 to 100 ppm.

Does 6015 have auto calibration and an internal sample pump?

Auto calibration: available, with internal solenoid valves or as a relay output for external valves. Sample pump: an internal sample pump is available on a single measurement, or a relay output drives an external pump.

What is the part number?

It starts ACCURRA-S-6015. Positions 5 to 8 are a second gas or XXXX, then the power supply, the range ID or pump relay, and auto calibration. The part number builder writes the full code and checks it against the catalogue.