ACCURRA-S M015 measures dew point from -80 to +20 °C dp (minimum range -80 to 0 °C dp), with a ceramic sensor: accuracy +/- 2 °C dp, response 2 min at 5 LPM, dry to wet. Ordering code ACCURRA-S-M015 followed by the second component and options.
Specifications
| Item | Specification | Notes |
|---|---|---|
| Measures | Moisture / dew point | |
| Range | -80 to +20 °C dp | Configurable within this span; dual-range 4–20 mA output |
| Minimum range | -80 to 0 °C dp | |
| Sensor | Ceramic sensor | |
| Display resolution | 1 °C | |
| Intrinsic error (accuracy) | +/- 2 °C dp | |
| Repeatability | 1 °C dp | |
| Response T90 | 2 min at 5 LPM, dry to wet | At 1 LPM unless stated |
| Notes | Sensor bypass valve supplied; field calibration not possible |
From datasheet SAPL/PDS/ACCURRA-S-10/25-Aug. Specifications may change as products improve.
Built into every ACCURRA-S
| Item | Specification | Notes |
|---|---|---|
| Display | Backlit LCD, 4 lines x 20 characters | |
| Analog output | 1 x 4–20 mA isolated, max load 500 ohm, dual range | Range freely selectable over the full measurement range |
| Relays | 6 x SPST rated 1 A 230 V AC, all failsafe | Alarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves |
| Communication | Modbus RTU over RS 485 | Read measured value, status, set-up and history; write auto-calibration start and set-up |
| Enclosure | 19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mm | IP 20, safe area only; net about 6 kg, gross about 8 kg |
| Power | 100 to 240 V AC, 50/60 Hz, 45 W | |
| Inlet and outlet | 1/4 inch OD SS316 | |
| Ambient | 5 to 40 °C; below 90 % RH non-condensing | Storage 0 to 50 °C |
Options for this model
- Auto calibration: not available on this model.
- Sample pump: no internal sample pump on this model; a relay output can drive an external pump.
- Range ID relay: shows which of the two 4–20 mA ranges is active; it shares a relay with the sample pump, so choose one.
- A sensor bypass valve is supplied, to isolate the sensor during purging or upsets.
- Field calibration is not possible on M015; ask us how the sensor is checked and recalibrated.
Build the part number Write the specification
Dual measurement
The M015 can share an ACCURRA-S enclosure with a second sensor in series, each with its own electronics, 4–20 mA output, relays and Modbus address. The catalogue lists it with 3015 (oxygen, trace). Other pairs are possible on request.
Sample at the analyser
- The sample gas should be non-corrosive, non-toxic, non-flammable, non-condensing and dry, and free from entrained oil.
- Sample pressure 2 to 10 psig maximum, regulated externally. Recommended flow 1 LPM.
- Dust or particulate below 3 micron; the built-in fine filter handles light loads, and heavier dust needs an upstream filter.
- Materials in contact with sample: SS316, Viton, PTFE, glass, aluminium and acrylic.
- Sample temperature 5 to 45 °C, and a sample dew point at least 5 °C below the lowest ambient temperature.
Where it is used
Application notes where M015 is part of the recommended system:
- Oxygen purity in air separation plants
- Nitrogen purity and trace oxygen
- Dew point and trace moisture in process gases
- Gas purity certification and cylinder filling
- Electrolyser safety and quality
- Heat-treatment furnace atmosphere
- Semiconductor and electronics gas purity
- Cold rolling mill furnaces
- Compressed-air and instrument-air dew point (ISO 8573)
- Medical-air and breathing-air quality
- Glove boxes and isolators (O2 and moisture)
- Lithium-ion battery dry rooms and electrode drying
Frequently asked questions
What range does ACCURRA-S M015 measure?
-80 to +20 °C dp, with a minimum range of -80 to 0 °C dp.
Does M015 have auto calibration and an internal sample pump?
Auto calibration: not available on this model. Sample pump: no internal sample pump on this model; a relay output can drive an external pump.
What is the part number?
It starts ACCURRA-S-M015. Positions 5 to 8 are a second gas or XXXX, then the power supply, the range ID or pump relay, and auto calibration. The part number builder writes the full code and checks it against the catalogue.
Other models
- ACCURRA-S 3015 trace oxygen analyser
- ACCURRA-S 3030 percent oxygen analyser
- ACCURRA-S 3090 oxygen purity analyser
- ACCURRA-S 3130 paramagnetic oxygen analyser
- ACCURRA-S 3190 paramagnetic oxygen purity analyser
- ACCURRA-S 2015 ppm carbon monoxide analyser
- ACCURRA-S 2030 percent carbon monoxide analyser
- ACCURRA-S 6015 ppm carbon dioxide analyser
- ACCURRA-S 6030 percent carbon dioxide analyser
- ACCURRA-S 4030 methane analyser
- ACCURRA-S M015 moisture analyser
- How ACCURRA model numbers work