ACCURRA-S · Model 3015

ACCURRA-S 3015 trace oxygen analyser

ACCURRA-S 3015 measures trace oxygen over 0 to 1000 ppm (minimum range 0 to 100 ppm) with an electrochemical sensor. It is a safe-area, 19-inch rack or panel analyser with an isolated 4–20 mA output, relays and Modbus, made by Sarvesh in Pune, India.

Quick answer

ACCURRA-S 3015 measures trace oxygen from 0 to 1000 ppm (minimum range 0 to 100 ppm), with an electrochemical sensor: accuracy 1 % of reading +/- 2 ppm, response below 30 s. Ordering code ACCURRA-S-3015 followed by the second component and options.

Figures from the ACCURRA-S datasheet (SAPL/PDS/ACCURRA-S-10/25-Aug)

Specifications

ItemSpecificationNotes
MeasuresOxygen, trace
Range0 to 1000 ppmConfigurable within this span; dual-range 4–20 mA output
Minimum range0 to 100 ppm
SensorElectrochemical
Display resolution0.1 ppm
Intrinsic error (accuracy)1 % of reading +/- 2 ppm
RepeatabilityZero +/- 2 ppm; span 1 % of reading
Response T90Below 30 sAt 1 LPM unless stated
NotesSensor life above 18 months; sensor bypass valve supplied

From datasheet SAPL/PDS/ACCURRA-S-10/25-Aug. Specifications may change as products improve.

Built into every ACCURRA-S

ItemSpecificationNotes
DisplayBacklit LCD, 4 lines x 20 characters
Analog output1 x 4–20 mA isolated, max load 500 ohm, dual rangeRange freely selectable over the full measurement range
Relays6 x SPST rated 1 A 230 V AC, all failsafeAlarm 1, Alarm 2, fault, sample pump or range ID, two for auto-calibration valves
CommunicationModbus RTU over RS 485Read measured value, status, set-up and history; write auto-calibration start and set-up
Enclosure19 inch rack or panel; 134 H x 483 W x 180 (short) or 280 (long) D mmIP 20, safe area only; net about 6 kg, gross about 8 kg
Power100 to 240 V AC, 50/60 Hz, 45 W
Inlet and outlet1/4 inch OD SS316
Ambient5 to 40 °C; below 90 % RH non-condensingStorage 0 to 50 °C

Options for this model

  • Auto calibration: not available on this model.
  • Sample pump: no internal sample pump on this model; a relay output can drive an external pump.
  • Range ID relay: shows which of the two 4–20 mA ranges is active; it shares a relay with the sample pump, so choose one.
  • A sensor bypass valve is supplied, to isolate the sensor during purging or upsets.

Build the part number Write the specification

Dual measurement

The 3015 can share an ACCURRA-S enclosure with a second sensor in series, each with its own electronics, 4–20 mA output, relays and Modbus address. The catalogue lists it with 2015 (carbon monoxide), 6015 (carbon dioxide) and M015 (moisture / dew point). Other pairs are possible on request.

For hazardous areas

The same 3015 sensor is available in ACCURRA-Hz, in a certified enclosure (Gas Group IIC, Zone 1 and 2, T6), wall mounted, with 24 V DC or AC power. Calibration is done without opening the covers. See ACCURRA-Hz.

Sample at the analyser

  • The sample gas should be non-corrosive, non-toxic, non-flammable, non-condensing and dry, and free from entrained oil.
  • Sample pressure 2 to 10 psig maximum, regulated externally. Recommended flow 1 LPM.
  • Dust or particulate below 3 micron; the built-in fine filter handles light loads, and heavier dust needs an upstream filter.
  • Materials in contact with sample: SS316, Viton, PTFE, glass, aluminium and acrylic.
  • Sample temperature 5 to 45 °C, and a sample dew point at least 5 °C below the lowest ambient temperature.

Where it is used

Application notes where 3015 is part of the recommended system:

Frequently asked questions

What range does ACCURRA-S 3015 measure?

0 to 1000 ppm, with a minimum range of 0 to 100 ppm.

Does 3015 have auto calibration and an internal sample pump?

Auto calibration: not available on this model. Sample pump: no internal sample pump on this model; a relay output can drive an external pump.

What is the part number?

It starts ACCURRA-S-3015. Positions 5 to 8 are a second gas or XXXX, then the power supply, the range ID or pump relay, and auto calibration. The part number builder writes the full code and checks it against the catalogue.

Is 3015 available for hazardous areas?

Yes, in ACCURRA-Hz, in a certified enclosure for Gas Group IIC, Zone 1 and 2, T6.